Development of a simultaneous imaging system to measure the optical and gamma ray images of Ir-192 source for high-dose-rate brachytherapy

, , , , , and

Published 9 December 2021 © 2021 IOP Publishing Ltd and Sissa Medialab
, , Citation J. Nagata et al 2021 JINST 16 T12005 DOI 10.1088/1748-0221/16/12/T12005

This article is corrected by 2022 JINST 17 E01001

1748-0221/16/12/T12005

Abstract

In high-dose-rate (HDR) brachytherapy, verification of the Ir-192 source's position during treatment is needed because such a source is extremely radioactive. One of the methods used to measure the source position is based on imaging the gamma rays from the source, but the absolute position in a patient cannot be confirmed. To confirm the absolute position, it is necessary to acquire an optical image in addition to the gamma ray image at the same time as well as the same position. To simultaneously image the gamma ray and optical images, we developed an imaging system composed of a low-sensitivity, high-resolution gamma camera integrated with a CMOS camera. The gamma camera has a 1-mm-thick cerium-doped yttrium aluminum perovskite (YAIO3: YAP(Ce)) scintillator plate optically coupled to a position-sensitive photomultiplier (PSPMT), and a 0.1-mm-diameter pinhole collimator was mounted in front of the camera to improve spatial resolution and reduce sensitivity. We employed the concept of a periscope by placing two mirrors tilted at 45 degrees facing each other in front of the gamma camera to image the same field of view (FOV) for the gamma camera and the CMOS camera. The spatial resolution of the imaging system without the mirrors at 100 mm from the Ir-192 source was 3.2 mm FWHM, and the sensitivity was 0.283 cps/MBq. There was almost no performance degradation observed when the mirrors were positioned in front of the gamma camera. The developed system could measure the Ir-192 source positions in optical and gamma ray images. We conclude that the developed imaging system has the potential to measure the absolute position of an Ir-192 source in real-time clinical measurements.

Export citation and abstract BibTeX RIS

10.1088/1748-0221/16/12/T12005