Probe area effect on performance of phoswich detectors for simultaneous alpha and beta detection

, , , and

Published 3 October 2019 © 2019 IOP Publishing Ltd and Sissa Medialab
, , 21st International Workshop On Radiation Imaging Detectors Citation W. Kim et al 2019 JINST 14 C10005 DOI 10.1088/1748-0221/14/10/C10005

1748-0221/14/10/C10005

Abstract

According to the Multi-Agency Radiation Survey and Site Investigation Manual, a site characterization and a final status survey are required to determine whether a potentially contaminated site satisfies the release criteria. Before conducting sampling at a site, direct measurements with a portable instrument to identify hotspot areas can provide an outline of the sampling plan to facilitate accurate analysis. As the instrument, a phosphor sandwich (phoswich) detector has the advantage of being able to detect alpha and beta rays simultaneously. In this study, large-area phoswich detectors with various probe sizes were fabricated, optimized, and calibrated to investigate variations in the minimum detectable concentration (MDC). To determine the optimal thickness of the light guide, commercially available optical design software LightTools was used to simulate light transport. Each phoswich detector was coupled to an optimal light guide to achieve the highest instrument efficiency. The proposed instrument was calibrated using 241Am, 90Sr/Y, 99Tc, and 14C radioactive sources, and experimental data were employed in the MDC calculation. Although it is well known that a large probe size improves the overall performance, the results indicated that probe sizes greater than 20 × 20 cm do not enhance instrument performance.