Characteristic and quality control test in sector collimator gamma knife perfexion at Siloam hospital

, and

Published under licence by IOP Publishing Ltd
, , Citation Jhon Wesly Manik et al 2017 J. Phys.: Conf. Ser. 795 012010 DOI 10.1088/1742-6596/795/1/012010

1742-6596/795/1/012010

Abstract

In this study conducted to evaluate the collimator 4 mm, 8 mm, 16 mm to determine the level of precision Gamma Knife Perfexion there three years and eight months has not carried out measurements after the first year and verify the alignment of sector 1-8. Measurement with three axes (x-axis, y-axis, z-axis) using the film Gafcromic EBT-3, which will be signaled to a central point in the film to be measured and given a dose of 5 Gy for 10 minutes and then scanned and analyzed using ImageJ softwere 1.50 and dose rate in film comparison with measurent dose rate using electrometer. Alignment of Patient-Positioning System (PPS) with Radiation Focal Point (RFP) determination of quality control testing each collimator 4 mm in the standard test with < 0.1 and > 0.4 mm passive voice AAPM 54. The results of the standard are still in conformity item 0.21 mm for 4 mm and corelasi dose rate film and measurent using electrometer collimator 4 mm = 0.965, 8 mm = 0.964, 16 mm = 0.959.

Export citation and abstract BibTeX RIS

Content from this work may be used under the terms of the Creative Commons Attribution 3.0 licence. Any further distribution of this work must maintain attribution to the author(s) and the title of the work, journal citation and DOI.