On isocentre adjustment and quality control in linear accelerator based radiosurgery with circular collimators and room lasers

, , , , , and

Published under licence by IOP Publishing Ltd
, , Citation H Treuer et al 2000 Phys. Med. Biol. 45 2331 DOI 10.1088/0031-9155/45/8/318

0031-9155/45/8/2331

Abstract

We have developed a densitometric method for measuring the isocentric accuracy and the accuracy of marking the isocentre position for linear accelerator based radiosurgery with circular collimators and room lasers. Isocentric shots are used to determine the accuracy of marking the isocentre position with room lasers and star shots are used to determine the wobble of the gantry and table rotation movement, the effect of gantry sag, the stereotactic collimator alignment, and the minimal distance between gantry and table rotation axes. Since the method is based on densitometric measurements, beam spot stability is implicitly tested. The method developed is also suitable for quality assurance and has proved to be useful in optimizing isocentric accuracy. The method is simple to perform and only requires a film box and film scanner for instrumentation. Thus, the method has the potential to become widely available and may therefore be useful in standardizing the description of linear accelerator based radiosurgical systems.