Bilateral key comparison of INMETRO and INTI humidity standards in the dew/frost-point temperature range from -30 °C to +60 °C (SIM.T-K6.7)

and

©2020 BIPM & IOP Publishing Ltd
, , Citation Julio D Brionizio and Javier G Skabar 2020 Metrologia 57 03002 DOI 10.1088/0026-1394/57/1A/03002

0026-1394/57/1A/03002

Abstract

The aim of this paper is to describe a bilateral comparison carried out by the hygrometry laboratories of the National Metrological Institutes of Brazil and Argentina, INMETRO and INTI, respectively. The comparison, registered as a bilateral key comparison of the Inter American Metrology System (SIM) as SIM.T-K6.7 - INMETRO/INTI, was performed in the range from -30 °C to 60 °C of dew/frost-point temperatures. This paper presents the facilities of the laboratories, the measurement procedures, the uncertainty analysis, the compatibility of the measurement results (by means of the normalised error and by the degree of equivalence) and the degree of equivalence for the dew-point temperatures 1 °C and 20 °C between INTI and the CCT-K6 key comparison reference value by means of a previous comparison between INMETRO and the National Institute of Standards and Technology (NIST).

Main text

To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/.

The final report has been peer-reviewed and approved for publication by the CCT, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).