Optimization of spatial resolution and detection efficiency for photon/electron/neutron/ion counting detectors with Microchannel Plates and Quad Timepix readout

, and

Published 7 November 2018 © 2018 IOP Publishing Ltd and Sissa Medialab
, , 20th International Workshop On Radiation Imaging Detectors Citation A.S. Tremsin et al 2018 JINST 13 C11005 DOI 10.1088/1748-0221/13/11/C11005

1748-0221/13/11/C11005

Abstract

Although event counting detectors with Microchannel Plates (MCPs) were originally developed primarily for applications with low input fluxes, recent progress in readout electronics substantially extended their counting rate capabilities, while still maintaining very low dark count rates and no readout noise. Detection of many simultaneous events and operation at rates exceeding 100 MHz has been demonstrated with combination of MCPs and a Timepix readout. The spatial resolution in such devices can be an order of magnitude better than the 55 μm pixel size of Timepix chips when event centroiding is implemented. However, a compromise between spatial resolution and the number of accepted events has to be found for such detectors operating at high rates. Here we present an MCP/Timepix detector capable of simultaneous detection of an image with the highest spatial resolution of ∼6.5 μm, but reduced number of counts and an image with all registered photons but reduced 55 μm spatial resolution. A compromise between detector resolution and the acquisition duty cycle is also discussed. It was confirmed that high spatial resolution can still be maintained through event centroiding even with 60% of pixels hit by more than one photon per acquisition frame.

Export citation and abstract BibTeX RIS

10.1088/1748-0221/13/11/C11005