Determination of the wavelength dependence of the differential pathlength factor from near-infrared pulse signals

, , , , , and

Published under licence by IOP Publishing Ltd
, , Citation Matthias Kohl et al 1998 Phys. Med. Biol. 43 1771 DOI 10.1088/0031-9155/43/6/028

0031-9155/43/6/1771

Abstract

For the calculation of changes in oxyhaemoglobin, deoxyhaemoglobin and the redox state of cytochrome-c-oxidase from attenuation data via a modified Beer-Lambert equation the wavelength dependence of the differential pathlength factor (DPF) has to be taken into account. The DPF, i.e. the ratio of the mean optical pathlength and the physical light source-detector separation at each wavelength, determines the crosstalk between the different concentrations and is therefore essential for a sensitive detection of chromophore changes. Here a simple method is suggested to estimate the wavelength dependence of the DPF from pulse-induced attenuation changes measured on the head of adult humans. The essence is that the DPF is the ratio of the attenuation changes over absorption coefficient changes, and that the spectral form of the pulse correlated absorption coefficient change can be assumed to be proportional to the extinction coefficient of blood. Indicators for the validity of the DPF derived for wavelengths between 700 and 970 nm are the stability of the calculated haemoglobin and cytochrome signals with variations of the wavelength range included for their calculation and its overall agreement with the data available from the literature.

Export citation and abstract BibTeX RIS