Neutron dosimetry in high energy X-ray beams of medical accelerators

and

Published under licence by IOP Publishing Ltd
, , Citation M Sohrabi and K Z Morgan 1979 Phys. Med. Biol. 24 756 DOI 10.1088/0031-9155/24/4/007

0031-9155/24/4/756

Abstract

High energy X-ray beams from medical accelerators are used in cancer therapy. In such beams, neutrons are also produced due to photon-neutron interactions in the target materials, collimator and beam flattening filter. The feasibility of employing fast-neutron-induced recoil particle tracks in polycarbonate foils developed by the electrochemical etching (ECE) method for the dosimetry of such neutrons was investigated. The experiments were carried out in, and out of, the beam of a Varian Clinac-18, an Allis-Chalmers 25 MeV betatron and a Brown Boveri 45 MeV betatron. Neutron dose equivalents were found both as a function of the photon dose and of the distance away from the centre of the beam. The neutron dose equivalent was found to be 0.6 and 1.3% of the X-ray dose at the centre position of the beam for a TSD=100 cm and a 10*10 cm2 field for the 25 and 45 MeV betatrons respectively.

Export citation and abstract BibTeX RIS

10.1088/0031-9155/24/4/007