Linearisation of the response from a platinum resistance thermometer

Published under licence by IOP Publishing Ltd
, , Citation J H Nobbs 1982 J. Phys. E: Sci. Instrum. 15 716 DOI 10.1088/0022-3735/15/7/006

0022-3735/15/7/716

Abstract

Two simple circuits of low construction cost have been developed in which the current through a platinum thermometer element (DIN 43760) depends on its resistance in such a way as to make the voltage generated across it vary linearly with temperature. The linearity over the temperature range -150 degrees C to greater than 300 degrees C is to within 0.2 degrees C.

Export citation and abstract BibTeX RIS