A cantilever with an integrated ZnO piezoelectric actuator in feedback with a piezoresistive sensor is utilized in an atomic force microscope (AFM) to achieve a new high speed imaging technique. The imaging bandwidth is increased from 0.6 to 6 kHz by bending the cantilever over sample topography with the actuator rather than moving the sample with a 2 in. piezotube. Images taken in the constant force mode with a 3 mm/s tip velocity of a sample containing 2 μm vertical steps are presented. The effects of electrical coupling from the actuator were eliminated by measuring the piezoresistor sensor with a lock‐in amplifier.
REFERENCES
1.
S. C.
Minne,
Ph.
Flueckinger,
H. T.
Soh
, and C. F.
Quate,
J. Vac. Sci. Technol. B
13
, 1380
(1995
).2.
S. C.
Minne,
H. T.
Soh,
Ph.
Flueckinger
, and C. F.
Quate,
Appl. Phys. Lett.
66
, 703
(1995
).3.
4.
5.
6.
7.
S.
Akamine,
T. R.
Albrecht,
M. J.
Zdeblick
, and C. F.
Quate,
IEEE Electron Device Lett.
10
, 490
(1989
).8.
T. Itoh and T. Suga, in The Proceedings of Transducers ’95, Stockholm (Royal Swedish Academy of Engineering Science, 1995), Vol. IV A, p. 632.
9.
T.
Fujii,
S.
Watanabe,
M.
Suzuki
, and T.
Fujiu,
J. Vac. Sci. Technol. B
13
, 1119
(1995
).10.
M. Tortonese, H. Yamada, R. C. Barrett, and C. F. Quate, in The Proceedings of Transducers ’91 (IEEE, New York, 1991), IEEE Publication No. 91CH2817-5, p. 448.
11.
EG&G 5302, Princeton Applied Research, Princeton, NJ 08543.
12.
Arlunya TF5000, Dindima Group, Victoria 3133, Australia.
This content is only available via PDF.
© 1996 American Institute of Physics.
1996
American Institute of Physics
You do not currently have access to this content.