CC BY-NC-ND 4.0 · Endoscopy 2023; 55(S 01): E1-E2
DOI: 10.1055/a-1899-8441
E-Videos

A case of early esophageal cancer with pharyngeal stenosis treated by endoscopic submucosal multi-tunnel dissection using an ultra-thin endoscope

Kohei Ono
Department of Gastrointestinal Endoscopy, NTT Medical Center Tokyo, Tokyo, Japan
,
Ken Ohata
Department of Gastrointestinal Endoscopy, NTT Medical Center Tokyo, Tokyo, Japan
,
Yoshiaki Kimoto
Department of Gastrointestinal Endoscopy, NTT Medical Center Tokyo, Tokyo, Japan
,
Rindo Ishii
Department of Gastrointestinal Endoscopy, NTT Medical Center Tokyo, Tokyo, Japan
,
Ryoju Negishi
Department of Gastrointestinal Endoscopy, NTT Medical Center Tokyo, Tokyo, Japan
,
Yohei Minato
Department of Gastrointestinal Endoscopy, NTT Medical Center Tokyo, Tokyo, Japan
,
Takashi Muramoto
Department of Gastrointestinal Endoscopy, NTT Medical Center Tokyo, Tokyo, Japan
› Author Affiliations

A 70-year-old man presented with early-stage esophageal cancer located in a semicircle in the upper esophagus. The hypopharynx was getting narrower owing to postoperative chemoradiotherapy for past pharyngeal cancer, and it was impossible to pass the pyriform sinus using a conventional scope. Therefore, we performed endoscopic submucosal multi-tunnel dissection, which has been shown to be effective for large esophageal cancers

We used an ultra-thin endoscope (EG-L580NW7; Fujifilm, Tokyo, Japan) that could pass through the stenosis. The diameter of the endoscope was 5.8 mm and that of the working channel diameter 2.4 mm. We further used small-caliber endoscopic submucosal dissection (ESD) devices, the SOUTEN endoknife (Kaneka Medix, Tokyo, Japan) ([Fig. 1]) and the RAICHO2 hemostatic forceps (Kaneka Medix) ([Fig. 2]). The distal attachment placed on the ultra-thin endoscope tip was created by hand using transparent tape [1]. A local injection (25 gauge needle, 3 mm; TOP, Tokyo, Japan) was performed using glycerol. The electrosurgical unit (VIO 300 D; ERBE, Tübingen, Germany) was set to Endocut Mode I (effect 3, duration 2, interval 2) for mucosal incision and forced coagulation mode (effect 2, 45 W) for submucosal dissection. We created two tunnels and then resected the submucosa between the tunnels ([Fig. 3]) ([Video 1]).

Zoom Image
Fig. 1 The SOUTEN 1.5-mm needle-knife with a knob-shaped tip attached to the top of the snare loop. The diameter of the outer sheath is 2.35 mm.
Zoom Image
Fig. 2 The RAICHO2 rotatable monopolar-type hemostatic device. The diameter of the outer sheath is 2.3 mm.
Zoom Image
Fig. 3 Remaining parts between the two tunnels.

Video 1 A case of early esophageal cancer with pharyngeal stenosis treated by endoscopic submucosal multi-tunnel dissection using an ultra-thin endoscope.


Quality:

The ultra-thin endoscope made it possible to resect the esophageal lesions with pharyngeal stenosis [2]. But its thinness sometimes resulted in unsteady movement. Endoscopic submucosal multi-tunnel dissection allowed the endoscope itself to be fastened by a tunnel, making it possible to stabilize the maneuverability and the operative field even with an ultra-thin endoscope, allowing ESD to be completed ([Fig. 4]). There is no previous report of endoscopic submucosal multi-tunnel dissection using an ultra-thin endoscope. Using this technique, ESD can be performed on large esophageal lesions even in situations where only an ultra-thin endoscope is available.

Zoom Image
Fig. 4 a Resected specimen: en bloc resection was achieved. b Pathological examination confirmed the diagnosis of squamous cell carcinoma, 22 × 19 mm, depth LPM, no vascular or lymphovascular invasion.

Endoscopy_UCTN_Code_CCL_1AB_2AC_3AB

Endoscopy E-Videos
https://eref.thieme.de/e-videos

Endoscopy E-Videos is an open access online section, reporting on interesting cases and new techniques in gastroenterological endoscopy. All papers include a high quality video and all contributions are freely accessible online. Processing charges apply (currently EUR 375), discounts and wavers acc. to HINARI are available.

This section has its own submission website at https://mc.manuscriptcentral.com/e-videos



Publication History

Article published online:
01 September 2022

© 2022. The Author(s). This is an open access article published by Thieme under the terms of the Creative Commons Attribution-NonDerivative-NonCommercial License, permitting copying and reproduction so long as the original work is given appropriate credit. Contents may not be used for commercial purposes, or adapted, remixed, transformed or built upon. (https://creativecommons.org/licenses/by-nc-nd/4.0/)

Georg Thieme Verlag KG
Rüdigerstraße 14, 70469 Stuttgart, Germany

 
  • References

  • 1 Kurebayashi M, Sakai E, Ohata K. Usefulness of a handmade distal endoscope attachment with a transparent tape. Video GIE 2020; 5: 226-228
  • 2 Muramoto T, Sakai E, Ohata K. Thin-endoscope endoscopic submucosal dissection for early esophageal cancer with postoperative stricture. Digestive Endoscopy 2020; 32: e11-e12