Performance of an inductively coupled plasma source ion trap mass spectrometer

Abstract

A unique system incorporating features of inductively coupled plasma (ICP) and ion trap mass spectrometry (ITMS) is described. The combined ICP-ITMS system is characterized instrumentally and the initial performance features are presented. Bare atomic ions are observed for all elements, including previously problematic high oxygen bond strength elements that produced high levels of MOn+ and M(OH)n+ ions in early ICP-ITMS experiments. Approximate, unoptimized detection limits of 10–500 pg ml–1 are reported. Other advantages of the system include the complete destruction of typical ICP-MS polyatomic ions and the near complete neutralization of plasma Ar+, both of which contribute to much simpler spectra below 80 m/z.

Article information

Article type
Paper

J. Anal. At. Spectrom., 1994,9, 1053-1058

Performance of an inductively coupled plasma source ion trap mass spectrometer

D. W. Koppenaal, C. J. Barinaga and M. R. Smith, J. Anal. At. Spectrom., 1994, 9, 1053 DOI: 10.1039/JA9940901053

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements