Trace element analysis of biological material following pressure digestion with nitric acid-hydrogen peroxide and microwave heating

Abstract

A commercial microwave acid-digestion system employing three types of closed vessels, a pressure-relief PFA-Teflon bomb (A), a pressure-relief Berghof all PTEE bomb (B) and a completely closed Parr microwave acid digestion bomb (C), was evaluated for sample digestion prior to the determination of trace elements in a biological tissue. The digestion procedure was based on sample dissolution with an HNO3-H2O2 mixture using microwave heating. The resulting solutions, analysed by flame and graphite furnace atomic absorption spectrometry, showed good agreement for all three closed-vessel techniques. Sample preparation time was approximately 10 min for vessels B and C, and 15 min for vessel A (including subsequent cooling time and preparation of the final solution).

Article information

Article type
Paper

J. Anal. At. Spectrom., 1989,4, 323-327

Trace element analysis of biological material following pressure digestion with nitric acid-hydrogen peroxide and microwave heating

H. Matusiewicz, R. E. Sturgeon and S. S. Berman, J. Anal. At. Spectrom., 1989, 4, 323 DOI: 10.1039/JA9890400323

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements