Issue 16, 2014

Delayed voltammetric with respect to amperometric electrochemical detection of concentration changes in microchannels

Abstract

The time response of an electrode incorporated into a fluidic channel to variations in analyte concentration of the outer-sphere redox probe ferrocenemethanol was investigated both for amperometry (AMP) and cyclic voltammetry (CV). The experimental data show that the temporal resolution of CV is not as good as that of AMP, as CV cannot properly detect fast concentration transients. The delayed response of CV was previously reported, for neurotransmitters, and mostly attributed to the adsorption of the analyte on the electrode surface. By using an outer-sphere redox couple, we show that mass transport also significantly delays the response of CV. The experimental delay time in CV was understood from mass transfer limitations due to the relaxation of the diffusion layer during repeated potential scanning. Furthermore, a robust protocol for the analysis of fast concentration transients was established, using the impulse and modulation transfer functions of the system. This method was found to be more precise than the mere analysis of undifferentiated traces in the time domain. As a proof of concept, the effect of increased viscosity was investigated, showing that AMP was more sensitive than CV to these variations. Overall, this analysis underlines further the enhanced temporal sensitivity of AMP over CV, at the expense of decreased chemical resolution, potentially having implications for in situ electrochemical detection of biologically relevant molecules.

Graphical abstract: Delayed voltammetric with respect to amperometric electrochemical detection of concentration changes in microchannels

Supplementary files

Article information

Article type
Paper
Submitted
25 Apr 2014
Accepted
05 Jun 2014
First published
05 Jun 2014

Lab Chip, 2014,14, 2929-2940

Author version available

Delayed voltammetric with respect to amperometric electrochemical detection of concentration changes in microchannels

R. Trouillon and M. A. M. Gijs, Lab Chip, 2014, 14, 2929 DOI: 10.1039/C4LC00493K

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements