Effects of specimen variables and stress amplitude on the S-N analysis of two PMMA based bone cements

https://doi.org/10.1016/j.ijfatigue.2017.08.019Get rights and content

Highlights

  • Changing sample cross section from circular to rectangular decreases fatigue life.

  • Changing sample surface preparation from directly moulded to a machined surface decreases fatigue life.

  • Cement formulation produces different responses to changes in fatigue stress amplitudes.

Abstract

The fatigue performance of bone cement is influenced by the testing parameters. In previous in vitro fatigue studies, different testing conditions have been used leading to inconsistencies in the findings between the studies, and consequent uncertainties about the effects of testing specimen specifications and stress parameters. This study evaluates the role of specimen variables (namely; specimen cross-section shape, surface production method and cement composition) in a range of in vitro stress amplitudes (±12.5, ±15, ±20, ±30 MPa), using S-N (Wöhler) analysis. The two main findings are: while specimen cross-section configuration and fabrication method (specimen type) played a key role in controlling the fatigue longevity of the same cement, the stress amplitude was seen as the dominant controlling variable to affect the fatigue behaviour of different cements when using the same specimen type. Thus, considering the effect of specimen type, testing at high stress amplitudes should be treated with caution, particularly in tension-compression loading, to ensure fatigue failure occurs due to mechanical rather than thermal effects and thus models the in vivo behaviour.

Introduction

PMMA based bone cements are used for the fixation of orthopaedic implants, thus their fatigue failure leads to major clinical problems in the form of implant loosening, pain and ultimately clinical failure of the device. The stress levels in the bone cement mantle around a joint replacement range from 3 to 11 MPa [1]. Depending on the patient’s age and activity, a hip or a knee replacement faces up to 2 million load cycles yearly [2], therefore these stress levels are required to be transmitted by the mantle of cemented joint replacements for the rest of the patient’s life, thus of the order of tens of millions of load cycles. In vitro, however, fatigue characterisation of bone cement has been performed using higher stress levels; mostly within the range of 10–30 MPa [3] to shorten testing time and reduce the number of test run-outs.

A review paper by Lewis [4] examined the “intrinsic” and “extrinsic” factors that can affect the measured fatigue performance of bone cement, concluded that there are “only a few areas of agreement” and “many areas of disagreement”. The effects of specimen preparation variables (cross section shape and surface production technique) on fatigue behaviour at a specific constant-amplitude stress have been examined in previous studies considering the influence of specimen shape in tension-compression only [5], surface preparation method in tension-tension only [6] or the effects of both configuration and fabrication methods in tension-compression [7], [8] and tension-tension loading [8], [9]. Due to testing at only one stress amplitude, the fatigue data has been analysed using Weibull relationships in most of these studies [5], [7], [8], [9] and only one study [8] has compared two testing methods: tension-only of rectangular moulded specimens at a single stress level versus tension-compression at multiple stress levels of circular machined specimens analysing the results using Weibull and Wöhler approaches, respectively. Fatigue testing at various stress amplitudes has been considered in some studies, but using only a single specimen configuration and modification technique within the same study (e.g. [10], [11], [12]). It has been suggested that using different fatigue testing methods makes it inappropriate to compare the findings from different studies [13], [14]. In addition to using a range of testing conditions, it has been pointed out that many of these studies “have employed inappropriate statistical methods” [3] and “have not addressed the issue of possible interactions between the parameters being investigated” [3]. Additionally, the cement formulation [13], methods of mixing, whether or not a partial vacuum was used to reduce porosity [6], [10], [11], [12], amount and type of opacifier and/or antibiotic [13], [15], all affect the fatigue life and have been discussed extensively [e.g. 15].

Currently, various standards are available for the fatigue testing of bone cement and have been used to various extents. ASTM F2118 was originally published in 2001, with various subsequent revisions, leading to ASTM F2118-14 being the current version [16]. This standard uses cylindrical dumbbell specimens, with fully reversed fatigue in phosphate buffered saline, either three stress levels are used, or the fatigue life for 5 million load cycles is found. The frequency is required to be constant and if above 5 Hz should be checked to ensure that no frequency effects occur. The suggested stress levels are ±15, ±12.5 and ±10 MPa which can be varied for hip and knee replacement applications or reduced to ±5, ±7 and ±9 MPa for spinal applications with a minimum of 15 speciemns used per load level. The specimens should be manufactured by injecting into a silicone mould which has been produced by moulding the silicone around machined metal blanks. The other major fatigue of bone cement standard is ISO 16402:2008 which was reconfirmed in 2013, and uses four point bending of rectangular bars [17]. The four point bending loads are from 5 N to the force which produces the required maximum stress. The specimens have moulded top and bottom surfaces and their sides may be either moulded or machined. So both standards require that the loaded surfaces are moulded, rather than machined. However, over the years many studies have used machined surfaces, particularly for cylindrical dumbbells, since it is easier to mould cylinders and then machine to shape. Furthermore, being fully reversed tension-compression, ATSM F2118-14 [16] leads to all areas of the cement undergoing the full stress range in both tension and compression, while the 4 point bending in ISO 16402:2008 [17] means that only the upper and lower surfaces are exposed to the maximum stresses and undergo either only tensile or only compressive loading, which may not be what occurs in vivo.

However, as both machined and moulded surfaces and rectangular and circular cross-sections have been used, it is important to examine the influence of specimen preparation method on more than one cement and one stress level [7], [9]. This requires the comparison of the fatigue performance of different specimen types at various stress amplitudes and examining whether testing specimen variables, along with changing the stress amplitude, can affect the fatigue behaviour of different specimens and cements and their relative ranking.

Hence the aim of the current study is to compare the fatigue results, with different specimen surface preparation methods and shapes, of two different bone cement formulations with the same viscosity classification (high viscosity) but different filler content. The fatigue life results are compared by the commonly used S-N (Wöhler) analysis (e.g. [8], [12], [18], [19]) as recommended in ASTM F2118-14. Furthermore, it has been reported that fatigue damage accumulation in vivo shows that failure progress is affected by the stress amplitude, thus using a single stress amplitude would be a “misleading measure of durability” [20].

Section snippets

Materials

Two brands of high viscosity bone cements were tested: SmartSet GHV Gentamicin and DePuy CMW1 (both produced and supplied by DePuy CMW, Blackpool, UK). The powder component of SmartSet GHV is a methyl methacrylate/methyl acrylate copolymer and contains 14.37 wt% zirconium dioxide as radiopaque agent and 4.22 wt% Gentamicin as antibiotic filler whereas the CMW1 polymeric powder is solely polymethylmethacrylate and contains 9.1 wt% barium sulphate as a radiopacifier with no antibiotic added. The

General comparison of S-N curves of various specimen types

The S-N curves (Fig. 1) show obvious differences between the fatigue behaviour of the two cements. For both bone cements, the effect of specimen shape in conjunction with surface preparation method is clear. Consistantly, the longest fatigue life was associated with the circular moulded (CMM) specimens at all stress amplitudes. Excluding this specimen type, the results appeared to be more dependent on the change in chemical composition where divergence in the trends of S-N curves was seen

Effects of specimen cross-sectional shape and size

The analysis of fatigue results using S-N curves has shown that specimen shape and surface production method (specimen type) can significantly alter the measured fatigue life of bone cement, supporting the trends of our previous findings using tension-tension and tension-compression both at a single stress amplitude and using the Weibull functions for data analysis [7], [9]. The effect of specimen shape on fatigue behaviour can be influenced by the material, thus two different cement

Conclusions

  • For the same cement composition, the response of fatigue crack initiation and growth appears to be controlled by the specimen configuration and fabrication method used. In general, moulded specimens, particularly with circular cross sections, tend to provide significantly longer fatigue lives than the machined samples at all stress amplitudes. Thus supporting the requirements of the current standards to have moulded surfaces.

  • The same specimen configuration and fabrication of different cement

Acknowledgements

The Libyan Government is thanked for funding the research (PhD studentship ref 6978) and DePuy CMW is acknowledged for providing the bone cement materials and mixing devices.

References (45)

  • W.R. Krause et al.

    Fatigue properties of acrylic bone cements: review of the literature

    J Biomed Mater Res

    (1988)
  • N. Wallbridge et al.

    The walking activity of patients with artificial hip joints

    Eng Med

    (1982)
  • G. Lewis et al.

    Toward standardization of methods of determination of fracture properties of acrylic bone cement and statistical analysis of test results

    J Biomed Mater Res B Appl Biomater

    (2000)
  • G. Lewis

    Fatigue testing and performance of acrylic bone-cement materials: state-of-the-art review

    J Biomed Mater Res B Appl Biomater

    (2003)
  • Paravic V, Nobel P, Alexander JW, Leibs TR, Elliot A. Effect of specimen preparation on porosity and fatigue life: a...
  • J.P. Davies et al.

    Comparison of the fatigue characteristics of centrifuged and uncentrifuged Simplex P bone cement

    J Orthop Res

    (1987)
  • E.J. Harper et al.

    Tensile characteristics of ten commercial acrylic bone cements

    J Biomed Mater Res (Appl Biomater)

    (2000)
  • J.A. Johnson et al.

    Fatigue of acrylic bone cement - effect of frequency and environment

    J Biomed Mater Res

    (1989)
  • C. Persson et al.

    Mechanical effects of the use of vancomycin and meropenem in acrylic bone cement

    Acta Orthop

    (2006)
  • ASTM F2118-14. Standard test method for constant amplitude of force controlled fatigue testing of acrylic bone cement...
  • ISO 16402:2008. Implants for surgery – acrylic resin cement – flexural fatigue testing of acrylic resin cements used in...
  • T.A. Freitag et al.

    Fracture characteristics of acrylic bone cements. II. Fatigue

    J Biomed Mater Res, Part A

    (1977)
  • Cited by (3)

    • Long-term mechanical properties of a novel low-modulus bone cement for the treatment of osteoporotic vertebral compression fractures

      2021, Journal of the Mechanical Behavior of Biomedical Materials
      Citation Excerpt :

      After injection into a fractured vertebral body, the bone cement will experience dynamic cyclical loading, mainly in compression (Wilke et al., 1999; Kazarian and Graves, 1977; Callaghan and McGill, 1995) and therefore, such loading condition is most relevant to replicate in an in vitro setting. Fatigue properties of standard bone cements alone have already been reported (Robo et al., 2018b; Ajaxon and Persson, 2014; Sheafi and Tanner, 2017, 2019; Schönning et al., 2020; Panpisut et al., 2019), as well as studies that evaluated cements in an in vitro or ex vivo augmented models (Lewis et al., 2008; Aghyarian et al., 2015, 2017a, 2017b). Moreover, unreacted monomer release is a recognized unavoidable effect after implantation of PMMA-based bone cements, which is rarely reported but is believed to have an effect on the initial mechanical properties of the material (López et al., 2014; Vallo et al., 1998).

    • Relationship between fatigue parameters and fatigue crack growth in PMMA bone cement

      2019, International Journal of Fatigue
      Citation Excerpt :

      In two of our previous studies [35,36] the differences in fatigue lives were attributed to variations in specimen production method and resultant surface morphology, leading to changes in the stress concentration factors and also considered the role of cement composition with limited discussion of crack propagation mechanisms. In our subsequent study [37], the specimen variables were examined further in a range of stress amplitudes, reporting a particular influence of the stress amplitude used in describing the fatigue longevity of different cements, without considering in depth the crack initiation and growth mechanisms. In the current study, the changes in absorbed energy per loading cycle are compared for different specimen types and the results have shown interesting findings regarding the effect of specimen type which are also controlled by cement composition.

    • Master S-N curve fitting method of welded structure and software development

      2020, Hanjie Xuebao/Transactions of the China Welding Institution
    View full text