Cobalt-60 therapy unit for large field irradiation

https://doi.org/10.1016/0360-3016(81)90461-2Get rights and content

Abstract

A cobalt-60 therapy unit designed and constructed by The Ontario Cancer Institute is described. This unit is capable of treating large fields up to 50 × 160 cm2 at a conventional source to surface distance of 90 cm by using large collimator openings. Three selectable filters were incorporated. They all act as electron filters to improve the build-up characteristics of these beams. Two of the three filters also act as beam flatteners. The longitudinal collimators are individually driven so that it is possible to produce a vertical beam edge by closing one of the collimators completely. This condition is very useful for matching adjacent fields. The output of this unit, at its operating distance, ranges from 100–200 rad/min depending on which of the filters is chosen, which makes the treatment time shorter than most known facilities for very large field irradiation. This unit has been in routine use since 1977.

References (11)

  • R.S. Bush et al.

    Treatment of epithelial carcinoma of the ovary: Operation, irradiation, and chemotherapy

    Am. J. Obsiet. Gynecol.

    (1977)
  • R.S. Bush et al.

    Primary lymphoma of gastrointestinal tract

    Radiology

    (1969)
  • J.R. Cunningham et al.

    A simple facility for whole body

    Radiology

    (1962)
  • A.J. Dembo et al.

    Whole abdominal irradiation by a moving-strip technique for patients with ovarian cancer

    Int. J. Radiat. Oncol. Biol. Phys.

    (1979)
  • P.J. Fitzpatrick et al.

    Half-body radiotherapy

    J. Can. Assoc. Rad.

    (1976)
There are more references available in the full text version of this article.

Cited by (0)

Support by a grant from the National Cancer Institute and the Medical Research Council both of Canada.

Department of Physics.

Department of Radiation Oncology.

View full text