Signal processing methods for pulse oximetry

https://doi.org/10.1016/0010-4825(95)00049-6Get rights and content

Abstract

Current signal processing technology has driven many advances in almost every aspect of life, including medical applications. It follows that applying signal processing techniques to pulse oximetry could also provide major improvements. This research was designed to identify and implement one or more techniques that could improve pulse oximetry oxygen saturation (SpO2) measurements. The hypothesis was that frequency domain analysis could more easily extract the cardiac rate and amplitude of interest from the time domain signal. The focus was on the digital signal processing algorithms that had potential to improve pulse oximetry readings, and then test those algorithms. This was accomplished using the Fast Fourier Transform (FFT) and the Discrete Cosine Transform (DCT). The results indicate that the FFT and DCT computation of oxygen saturation were as accurate without averaging, as weighted moving average (WMA) algorithms currently being used, and directly indicate when erroneous calculations occur.

References (37)

  • G.B. Drummond et al.

    Arterial oxygen saturation before intubation of the trachea

    Brit. J. Anaesthesia

    (1984)
  • Z. Wang

    Fast discrete sine transform algorithms

    Signal Processing

    (1990)
  • J.W. Severinghaus et al.

    History of blood gas analysis

    Int Anesthesiology Clinics

    (1987)
  • J.W. Severinghaus et al.

    Recent developments in pulse oximetry

    Anesthesiology

    (1992)
  • J.T. Moller et al.

    Randomized evaluation of pulse oximetry in 20,802 patients: II

    Anesthesiology

    (1993)
  • E.H. Wood et al.

    Photoelectric determination of arterial oxygen saturation in man

    (1943)
  • F.W. Cheney

    The ASA closed claims study after the pulse oximeter

    ASA Newsletter

    (1990)
  • J.A. Pologe

    Pulse oximetry: Technical aspects of machine design

    Int Anesthesiology Clinics

    (1987)
  • K.K. Tremper et al.

    Pulse oximetry and oxygen transform

  • J. A. H. Bos, W. Schelter, W. Gumbrecht, B. Montag, E. P. Eijking, S. Armbruster, W. Erdmann and B. Lachmann....
  • K.K. Tremper et al.

    Pulse oximetry: applications

  • J.C.S. Lee et al.

    Measurement of percent carboxyhemoglobin with pulse oximetry technique

  • J.C.S. Lee et al.

    Simultaneous measurement of percent carboxyhemoglobin and functional oxygen saturation

  • M.R. Neuman et al.

    Motion artifact in pulse oximetry

  • L.K.L. Lum et al.

    A controlled motion artifact study of EKG synchronization on pulse oximetry

  • L.K.L. Lum et al.

    Evaluation of pulse oximetry with EKG synchronization

  • J.E. Scharf et al.

    Pulse oximetry through spectral analysis

  • M.L. Heller et al.

    Polarographic study of arterial oxygenation during apnea in man

    The New England J. Medicine

    (1961)
  • Cited by (161)

    • Data-driven assessment of cardiovascular ageing through multisite photoplethysmography and electrocardiography

      2019, Medical Engineering and Physics
      Citation Excerpt :

      Although being a volumetric measure and not a direct measurement of the pulse pressure, thanks to the strict relationship between changes in arterial pressure and volume and its non-invasivity, it is commonly employed for arterial stiffness evaluation. PPG is often performed in a transmission modality on the fingers or toes, allowing the assessment of both arterial oxygen saturation and single wavelength pulse wave contour [17]. Thanks to the high diffusive properties of biological tissues within NIR spectral range [18], PPG measurements can also be performed in a back-reflection modality where the light source and detector are located on the same surface at a certain distance from one another (few centimeters).

    • The Correlation Between Finger Mean Arterial Pressure and Blood Pressure Measurement

      2023, International Journal of Engineering Trends and Technology
    • In-Ear SpO for Classification of Cognitive Workload

      2023, IEEE Transactions on Cognitive and Developmental Systems
    View all citing articles on Scopus

    This research work was supported in part by a grant from Group Technologies, Inc.