1 Introduction

Residual solvents or volatile organic solvents are used or formed during the manufacturing of pharmaceutical drug substances, intermediates, excipients or pharmaceutical drug product [1]. The solvents are toxic, have no therapeutic importance and affect the quality and stability of drug substances and drug products so they are not desirable in the final product [2,3,4]. Although it is difficult to remove completely with the common techniques used in practical manufacturing process such as increased process temperature or/and decreased pressure, they need to be minimized. However, depending on the nature of the API, residual solvents and drying condition of the process, some amount of residual solvents traces can be retained in the final drug substances or drug product. Thus, acceptable levels of many residual solvents are included in regulatory guideline; particularly in guideline Q3C issued by the International Conference on Harmonization of technical requirements for registration of pharmaceuticals for human use (ICH) [5, 6]. ICH has also included daily exposure limit of many solvents it has classified these solvents into four classes on the basis of the toxicity level and the degree to which they can be considered an environmental hazard [6]. Class I solvents (which covers 5 residual solvents) are known or suspected human carcinogens and environmental hazards, the use of these solvents should be avoided. Class I solvents should be identified and quantified. Class II solvents (which covers 29 residual solvents) are non-genotoxic animal carcinogens or possible causative agents of other irreversible toxicity such as neurotoxicity or teratogenicity. Use of these solvents should be limited. Class II solvents have individual limits. Class III solvents (which covers 26 residual solvents) having low toxic potential to man; no health-based exposure limit is needed. Class 3 solvents have PDEs of 50 mg or more per day. Finally, Class 4 solvents are those for which no adequate toxicological data have been found [7]. Therefore determination of residual solvents becomes a necessary procedure for quality control of drug substances and drug product to meet regulatory guideline and ensure patient safety [8, 9].

Head space gas chromatography (HSGC) is generally used to determine residual solvents because of its high separation efficiency and sensitivity for organic volatile solvents [10]. However head space bounds the analysis to those solvents being evaporated from HS only, it also requires larger sample load and analysis time should be longer due to sample equilibration. Headspace sampling is preferred because of its ability to avoid direct liquid or solid injection [11, 12]. HSGC methods minimize any possible interference caused by non-volatile substances or by the degradation/decomposition products of the non-volatile components. While comparing to headspace, direct injection method requires relatively low sample concentration, but the high boiling/melting point components of the sample may not be eluted through GC Column and they may contaminate the GC injection port and leads to poor chromatography. HSGC with FID detection has been mainly used for the analysis of organic volatile solvents present in the pharmaceutical drug substances and drug products [13,14,15,16]. Battu et al. have developed the residual solvent method where ethyl acetate and hexane has not been separated [17], while Adepu et al. have reported a method in which acetone and DCM has not been separated [18]. In addition, Puranik et al. have described the study on few residual solvents [19]. However, this method has some drawbacks, for example: dimethyl sulfoxide was used as diluent, which decomposes thermally to produce dimethyl sulfide; resolution between solvents is less than 1.5 and has been injected at low concentration; standard of residual solvents were not prepared as per ICH limit, etc.

This report describes the separation of six residual solvents, among which two pair of solvents were critical i.e. n-hexane and ethyl acetate, and DCM and acetone using headspace gas chromatography (HSGC) with flame ionization detector (FID) in Imatinib Mesylate API. Moreover, method was developed considering sample matrix and accuracy, precision and linearity were established.

Imatinib Mesylate is a tyrosine kinase inhibitor, used to treat patients with hematological malignancies or malignant sarcomas such as gastrointestinal stromal tumors, chronic myeloid leukemia, acute lymphoblastic leukemia, gastrointestinal stromal tumors [20, 21]. Imatinib Mesylate is designated chemically as 4-[(4-methylpiperazine-1-yl)methyl]-N-(4-methyl-3-{[4-(pyridin-3-yl)pyrimidin-2-yl]amino} phenyl) benzamide methanesulfonate (Fig. 1), is a specific inhibitor of the tyrosine kinase in BCR-ABL+ cell [22]. It is an approved drug for treatment of gastrointestinal stromal tumor since 2002 [21] and for chronic myelogenous leukemia since 2011 [23] from the United States food and drug administration. Six solvents are required for the synthesis of Imatinib Mesylate i.e. methanol, acetone, dichloromethane, n-hexane, ethyl acetate and pyridine [24] and these should be controlled in final API.

Fig. 1
figure 1

Chemical structure of Imatinib Mesylate

In this research article, we have been described a development and validation of an HS-GC analytical method for determination of six residual solvents used in the route of synthesis to produce the Imatinib Mesylate drug substances. Incorporated Residual solvents Permissible daily exposure limit and ICH class categories are described in Table 1.

Table 1 Class of the residual solvents of Imatinib Mesylate as per the ICH guidelines

2 Experimental

2.1 Chemical, material, and reagents

Imatinib Mesylate API was synthesized in Cadila Healthcare Ltd. (Ahmedabad, Gujarat, India) [25]. Methanol, acetone, dichloromethane (DCM), n-hexane, ethyl acetate and pyridine were purchased from Merck (India). 1-Methyl-2-Pyrrolidinone (NMP) was purchased from Spectrochem (Mumbai, India).

2.2 Instrumentation

HSGC system of an Agilent technologies 6890N equipped with a flame ionization detector with a headspace sampler (Agilent technologies G1888) was used for method development and method validation studies. A 2 mm I.D. deactivated direct split liner was used as an inlet liner and chemstation software was used for data acquisition and chromatographic data integration. A Sartorius micro balance CP225D (Germany) and micro-pipette (100–1000 µL from Borosil) were used.

2.3 Gas chromatographic conditions

30 m × 0.53 mm I.D., 3.0 µm film thickness DB-624 (bonded 6% cyanopropylphenyl-94% dimethylpolysiloxane) capillary GC column to achieve proper Separation in a developed method. DB-624 column was manufactured by J&W Scientific (Agilent Scientific Technologies, Wilmington, DE, USA). The GC method parameters, headspace sampler conditions, and oven temperature programming of the method have been mentioned in Table 2.

Table 2 Experimental condition for determination of residual solvents of Imatinib Mesylate API

2.4 Preparation of standard and sample solution

Standard solution of residual solvent was prepared according to respective ICH limit. A composite standard stock solution of all the known residual solvent was prepared in such a way that it contains a final concentration of 300 ppm for methanol, 500 ppm for acetone, 60 ppm for dichloromethane, 29 ppm of n-hexane, 500 ppm for ethyl acetate and 20 ppm for pyridine in NMP.

The blank vial was prepared with 1.0 mL of NMP, the standard vial was prepared with 1.0 mL of the standard solution and the sample vials were prepared with 100 mg per 1.0 mL NMP.

2.5 Method validation

The method validation was performed by evaluating specificity, limit of detection (LOD), limit of quantitation (LOQ), linearity, accuracy, intermediate precision, system suitability and method precision of residual solvents as specified in the ICH harmonized tripartite guideline (2005).

3 Result and discussion

3.1 Gas chromatographic analytical method development

Method development by HSGC involves critical parameter such as selection of sample solvent, detector, column, carrier gas, optimization of headspace condition and chromatographic conditions. Developed method should be specific, sensitive, robust and QC friendly. The critical parameter of the developed method is discussed below.

3.1.1 Selection of detector and carrier gas

A flame ionization detector (FID) was used for this method because FID has good sensitivity. The carrier gas was selected as nitrogen because it is economical as compared to helium.

3.1.2 Selection of column

The GC Column is a crucial parameter for developing an efficient and sensitive HSGC method. The residual solvents were commonly determined by bonded 6% cyanopropylphenyl-94%dimethylpolysiloxane (624) column due to its mid-polar nature, higher temperature limit than other polar columns (DB-Wax), and higher retention of polar solvents compare to non-polar column (DB-1). The retention of polar solvents on a DB-624 column is relatively stronger at high temperatures and could provide high separation between solvents which has minor differences in boiling points. Since the objective was to develop an efficient HSGC method, a DB-624 column (30 m × 0.53 mm, 3.0 µm film thickness) was selected, which is commonly used for residual solvent determination [26,27,28,29]. As shown in Fig. 2 of the representative chromatograms for the method using DB-624 column, all residual solvents present in Imatinib Mesylate API were well separated from each other and from diluent also. To separate dichloromethane and acetone was highly critical by HSGC. Resolutions of this critical pairs were 6.3 in a developed method. To define the effect of the stationary phase on the resolution, a non-polar GC column DB-1 (100% dimethyl polysiloxane) used with the same chromatographic conditions of the developed method. Resolution of a critical pair was decreased to 1.2 from 6.3 with DB-1 Column. These results clearly indicated that the DB-624 column is the best choice for the separation of all six residual solvents in Imatinib Mesylate API.

Fig. 2
figure 2

Typical chromatogram of the six composite standard solution

3.1.3 Selection of sample solvent

Several solvents were tried mainly DMF, DMSO and NMP for sample solvents and it was observed that NMP gave smooth baseline with no interference at the retention times of the targeted solvents. DMS Peak was formed during heating in an oven when using DMSO as sample solvents. DMF gave baseline interference at the retention time of Pyridine. So finally NMP was used as the standard and sample solvent because of its ability to dissolve a wide variety of drug substances. It has a high boiling point (202 °C) that does not interfere with the analysis of volatile solvents.

3.1.4 Chromatographic conditions

To develop an HSGC method, there are two strategies for selecting oven programs. The first strategy was to keep initial oven temperature low and then gradient elution and in second strategy isothermal elution at relatively high oven temperatures. We wanted to increase the retention time of methanol, so we choose the first strategy to start our method development and finalized 35 °C Temp as initial oven temp with Hold time for 2 min. After that in ramp 1, a slow gradient was applied i.e. 4 °C/min to 80 °C to resolve the critical pairs, dichloromethane and acetone. To elute other high boiling unknown impurities and NMP, which is having 202 °C boiling point, oven temp set at 40 °C/min increment to 230 °C with 12 min Final hold time. The flow rate of nitrogen was finalized at 3.3 mL/min which is equivalent to 24 cm/s of linear velocity. Finally, the method was developed with a total run time of about 29 min.

3.1.5 Optimization of headspace parameters

The sensitivity of the HSGC method was directly impacted by headspace oven temperature, Headspace oven Temperature should be kept same or above the boiling point of the residual solvents but below the boiling point of the sample solvents. To minimize the carryover problems Loop temperature has been kept 10–15 °C higher than oven temperature and the transfer line temperature also has been kept 10–15 °C higher than the loop temperature [30]. Headspace oven temperature was kept at 80 °C because among all residual solvents pyridine has a highest boiling point of 80 °C. The boiling point of the NMP is 202 °C. Therefore, the headspace oven, loop, and transfer line temperatures were selected at 80 °C, 90 °C and 110 °C, respectively. The vial equilibration time was set to 30 min. Other headspace parameter has been described in Table 2.

3.2 Method validation

The final method was validated per requirements prescribed in ICH guidelines [31].

3.2.1 Specificity

The method specificity was demonstrated by injecting the Blank, individual residual solvents standard solution (chromatogram has been attached in Fig. S-2) and specificity solution (composite standard solution of all residual solvents). In the developed chromatographic method, no interference was observed at the retention time of targeted residual solvents from each other and from sample solvents or other unknown peaks. The retention time of methanol, acetone, dichloromethane (DCM), n-hexane, ethyl acetate and pyridine was found to be 4.10, 6.07, 6.92, 7.99, 9.61 and 14.69 min, respectively. The retention time of sample solvent (NMP) was found to be 18.0 min. A typical chromatogram of the standard solution is shown in Fig. 2.

3.2.2 Linearity and range

The linearity of the method was determined using 7 concentration level over the range 20–150% of ICH Limit Level. The calibration curve was found to be linear within the range and correlation coefficient (r2) values for all six residual solvents were found to be higher than 0.99. Linearity curve and values for the residual solvents have been provided in Fig. 3 and Table 3. Each residual solvent was easily passed acceptance criteria for accuracy, system precision, method precision and linearity from the low concentration to high concentration, therefore the range of the method was 60–450 ppm for methanol, 100–750 ppm for acetone, 15–100 ppm for dichloromethane, 6–50 ppm for n-hexane, 100–750 ppm for ethyl acetate and 5–35 ppm for pyridine.

Fig. 3
figure 3

Linearity plot of residual solvents

Table 3 Linearity sample concentration and results

3.2.3 Method sensitivity

The LOQ were determined based on a signal-to-noise ratio of 3:1 and 10:1 respectively. Based on validation results, LOQ limit was found to be 60.3 ppm for methanol, 100.6 ppm for acetone, 15.1 ppm for dichloromethane, 6.7 ppm for n-hexane, 99.6 ppm for ethyl acetate and 4.8 ppm for pyridine. Accuracy at LOQ level was confirmed by injecting three different preparations of Imatinib spiked with each residual solvents at LOQ level and calculated % recoveries of each solvent.

3.2.4 Accuracy (recovery)

The accuracy of the method was determined by spiking of all solvents at four different level i.e. LOQ Level, 50% level, 100% level, and 150% level of ICH limit in a triplicate analysis. Recovery for all six solvents was found within the range of 80–120%. Recovery study and method precision results were reported in Table 4 and indicate that the method was accurate.

Table 4 Accuracy and method precision data

3.2.5 Precision

The precision of the method was determined by system precision (six replicate injection of standard solution) and method precision (six different preparation of spike solution) studies. In both the studies %Relative Standard Deviation of peak areas for all the solvents were less than 5.0%. The results were provided in Table 5. These results proved that the system suitability was passed and method is precise (Table 5).

Table 5 System precision and system suitability parameter

3.2.6 System suitability

The system suitability criterion was taken to be the resolution between the closely eluting pairs, i.e. acetone and dichloromethane. The system suitability was checked by injecting the standard solution before starting any analysis of residual solvents. The criterion for system suitability was that the resolution between the closely eluting pair should not be less than 1.5 and it was found that optimum resolution 6.3 was achieved (Table 5). Other system suitability parameter chromatogram has been attached in Fig. S-1.

3.2.7 Robustness

Robustness of the method was performed by slightly changing carrier gas flow rate, oven temperature and split ratio. The following parameters were changed: column oven temperature ± 3 °C from the ideal conditions (initial column oven temperature at 32 °C and 38 °C), the flow rate ± 5% from the ideal conditions (flow rate 2.0 psi and 2.5 psi), the split ratio ± 10% from the ideal conditions (the split ratio of 1:1.8 and 1:.2.2. The results have been mentioned in Table 6. No significant difference in % RSD, resolution and elution order which indicate that the method is robust. Results of robustness study were reported in Table 6.

Table 6 Results of robustness study

4 Result

Residual solvents analysis was performed on developed and validated method for commercial batch of Imatinib Mesylate API in triplicate. Results were reported in Table 7 and chromatogram has been attached in Fig. S-3.

Table 7 Residual solvents contents in the commercial batch of Imatinib Mesylate API

5 Conclusion

A selective and sensitive fast static HSGC method has been successfully developed for the determination of methanol, acetone, dichloromethane, n-hexane, ethyl acetate and pyridine in Imatinib Mesylate API through consideration of route of synthesis and solvents nature. The developed method was successfully validated as per regulatory guideline and found to be precise, accurate, linear, robust and specific. Additionally, our method is suitable for analysis of pyridine and other solvents in one single method, which is accurate, precise and linear in presence of sample matrix. However only a limited number of solvents are used in Imatinib Mesylate API, this method may be used to separate the residual solvents present in other drug substances and can be used for routine analysis to monitor in-process drying and in quality control for bulk drug manufacturing. Taken together, our developed HSGC method demonstrated precise, economical and commercially viable quantitative technique for residual solvents determination in Imatinib Mesylate API which will also be advantageous for industrial scale manufacturing.