Abstract
Purpose
A faster time performance can be achieved by focusing on prompt emitted photons, such as Cherenkov photons. In this study, the time fluctuations of Cherenkov photons transmission in the three Cherenkov radiators were first simulated, and we evaluated the detector performance and discussed ways to improve the coincidence time resolution (CTR) results for the detector with the three radiators. Finally, we compared detection efficiency of Cherenkov radiation in three radiators.
Methods
In this work, we recorded the step-by-step information of Cherenkov photons and the electrons that produced them in the radiator, information of photons on the photon outputting surface, and the step-by-step information tracing of the parent charged particles (electrons) to evaluate the factors influencing the performance time of the detector based on Geant4. Specially, we usually use a photon amplitude timing in the experimental implements, and the arrival time of an event is defined as the arrival time of the first photon of the event accordingly.
Results
The time fluctuations of Cherenkov photons transmission in the three Cherenkov radiators with a 5 mm length were simulated as 28.5 ps for PbF, 40.1 ps for PbWO4, and 24.7 ps for Al2O3. After considering other factors such as quantum efficiency (QE) of 30% and transit time spread (TTS) of 10 ps from the process of practical electronics, the CTR of a pair of single-pixel detectors were 56.6 ps for PbF2, 78.8 ps for PbWO4, and 46.3 ps for Al2O3 with a thickness of 5 mm. The detection efficiency of PbF2 is 32.25%, PbWO4 of 31.91% and Al2O3 of 8.14% in the case of 5 mm length.
Conclusions
The simulation results indicated that using the Cherenkov photons produced in PbF2, PbWO4, Al2O3 for detector timing can obtain good CTR results. In this study, it demonstrates that it is possible to achieve a pretty good CTR with an improved QE and a reduced TTS of the photodetector.
Similar content being viewed by others
References
T. Pan, S.A. Einstein, S.C. Kappadath et al., Performance evaluation of the 5-Ring GE Discovery MI PET/CT system using the national electrical manufacturers association NU 2–2012 Standard. Med. Phys. 46(7), 3025–3033 (2019)
S. Chen et al., Performance characteristics of the digital uMI550 PET/CT system according to the NEMA NU2–2018 standard. EJNMMI Phys. 7(1), 1–14 (2020)
M. Conti, L. Eriksson, V. Westerwoudt, in Nuclear Science Symposium & Medical Imaging Conference (IEEE, 2011)
N. Mythra Varun et al., Sub-100 ps coincidence time resolution for positron emission tomography with LSO: Ce codoped with Ca. Phys. Med. Biol. 60(12), 4635–4649 (2015)
J.W. Cates, C.S. Levin, Advances in coincidence time resolution for PET. Phys. Med. Biol. 61(6), 2255–2264 (2016)
S. Gundacker et al., High-frequency SiPM readout advances measured coincidence time resolution limits. Phys. Med. Biol. 64(5), 055012 (2019)
R. Ota et al., in 2019 IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC), Manchester, United Kingdom, pp. 1–4 (2019)
S.E. Brunner, L. Gruber, J. Marton, K. Suzuki, A. Hirtl, Studies on the Cherenkov effect for improved time resolution of TOF-PET. IEEE Trans. Nucl. Sci. 61(1), 443–447 (2014)
S.E. Brunner, D. Schaart, BGO as a hybrid scintillator / Cherenkov radiator for cost-effective time-of-flight PET. Phys. Med. Biol. 62(11), 4421–4439 (2017)
R. Ota et al., Timing-performance evaluation of Cherenkov-based radiation detectors. Nucl. Instrum. Methods Phys. Res., Sect. A 923, 1–4 (2019)
R. Ota et al., Coincidence time resolution of 30ps FWHM using a pair of Cherenkov-radiator-integrated MCP-PMTs. Phys. Med. Biol. 64(7), 07LT01 (2019)
D. Yvon, V. Sharyy, M. Follin et al., Design study of a scintronic crystal targeting tens of picoseconds time resolution for gamma ray imaging: the ClearMind detector. J. Instrm. 15(7), P07029–P07029 (2020)
C. Dania et al., Simulation study to improve the performance of a whole-body scanner. Phys. Med. Biol. 65(5), 055013 (2020)
H.H. Ross, Measurement of beta-emitting nuclides using Cerenkov radiation. Anal. Chem. 41(10), 1260–1265 (1969)
J.H. Beaumont, A.J. Bourdillon, J. Bordas, Optical properties of PbI2 and PbF2. J. Phys. C Solid State Phys. 10(5), 761 (2001)
O.V. Buyanov, et al., Beam studies of EM-calorimeter prototype built of PbWO4 crystals (1994)
Saint-Gobain Crystals Technical Sheet. www.crystals.saint-gobain.com
A.S. Agostinelli et al., Geant4—a simulation toolkit. Nucl. Instrum. Methods Phys. Res. Sect. A 506(3), 250–303 (2003)
R.E. Bell, Comparison of leading-edge and crossover timing in coincidence measurements. Nucl. Inst. Methods 42(2), 211–212 (1996)
T. Paulus, Principles and Applications of Timing Spectroscopy. Application Note AN-42, EG&E. (1982)
R. Ota et al., Cherenkov radiation-based three-dimensional position-sensitive PET detector: A Monte Carlo study. Med. Phys. 45(5), 1999–2008 (2018)
Hamamatsu PMT Product Data Sheet. www.hamamatsu.com.cn
Acknowledgements
This research was supported by National Natural Science Foundation of China (No. 11675191), Science and Technology Innovation Project funded by Institute of High Energy(IHEP), Chinese Academy of Sciences (No. E05457U207).
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Han, X., Wei, L., Huang, X. et al. Simulation research on time resolution based on Cherenkov radiation. Radiat Detect Technol Methods 5, 421–429 (2021). https://doi.org/10.1007/s41605-021-00266-5
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s41605-021-00266-5