Introduction

Many dynamic and quasi-dynamic experimental techniques for the determination of specific heat and other thermophysical properties of electrically conductive solid materials have been proposed and used since the beginning of twentieth century. They differ mostly depending on the applied power source which delivers the energy to the specimen and on the way of measurements of this energy and other specific quantities, especially the specimen temperature. Typically, the power sources have been either batteries or electrical capacitors, while the measurement of the specimen temperature has been direct, by using thermocouples or indirect, by using previously known function between the specimen electrical resistance and temperature. On the other hand, for applications at high and very high temperatures, due to increased heat transfer and dissipation, possible chemical reactions between the specimen and environment and mechanical stress of the specimen, the measurement time should be short as much as possible. In that sense, a method of fast specimen heating was proposed in 1970 by Cezairliyan et al. [1]. In their work, the specimen temperature was measured without contact, i.e., by a high-speed photoelectric pyrometer. The specimen used was in the form of a tube with a tiny centered hole through which radiation passed towards the pyrometer. Assuming a black body radiation from the hole, recorded radiance temperature was equaled to the absolute specimen temperature. The authors initially applied this method for the determination of specific heat and other thermophysical properties of molybdenum in the range from 1900 to 2800 K.

Although many high temperature materials have been characterized by pulse heating facilities based on the same or similar radiation thermometry as in Ref. [1], the contact thermometry, providing directly the data of absolute specimen temperature, has been still used in cases when the applied heating rate is not too fast (below millisecond resolution) and the temperature range not too high (above 2600 K). Application of such a measuring facility was reported in 1985 by Dobrosavljević and Maglić [2], who first proposed specific data reduction for the obtained thermocouple signals and later provided a corresponding error analysis [3]. Soon after, Maglić et al. [4] upgraded this facility and related measurement procedure in terms of applying both contact and noncontact measurements of the specimen temperature. This has been later used by different authors for the characterization of several high temperature metallic elements up to 2600 K, such as that of molybdenum [5], tantalum [6], hafnium [7], rhodium [8] and ruthenium [9].

In this research, an application of the subsecond calorimetry technique with both contact and radiance temperature measurement has been proposed above the temperature of 2600 K, which is the approximate limit of the standard contact thermometry. Measurements at very high temperatures have been made by using a high-speed pyrometer, but in contrast to techniques based on the method of Cezairliyan et al. [1] and that of Maglić et al. [4] this one uses the estimated values of spectral normal emissivity for each experimental run in order to convert the measured radiance temperature to the absolute temperature of the specimen. As a black-body approximation is not used in temperature measurements, the specimen does not need to have a hole in its center or to be in the form of a tube. Instead, it may take the form of a solid rod, which generally makes its manufacturing more convenient and cost effective.

This work is a part of a European joint research projectFootnote 1 aiming to establish reference facilities for measuring the specific heat of high temperature solids at temperatures up to about 3000 °C (~ 3300 K). The considered facilities are based on one stationary, the drop calorimetry method and two dynamic, the laser flash and the subsecond calorimetry techniques. In order to fulfill the goal regarding the temperature range, this work proposes a new measurement principle of the subsecond calorimetry technique and provides first results obtained on a pure tungsten specimen by using the related experimental facility of the Institute VINČA.

Principle

The purpose of the subsecond calorimetry technique is to measure the specific heat of an electro-conductive solid material at high temperatures and its principle is as follows: Heat rapidly a specimen with an electrical current pulse, i, and monitor simultaneously a voltage drop over the specimen, u, and its transient temperature, T. By knowing the specimen temperature and the total energy absorbed by the unit specimen mass, decreased by the total radiation heat losses, the specific heat of the specimen material can be determined, together with its specific electrical resistivity and total hemispherical emissivity.

The measurement of the specimen temperature is crucial here. It can be performed either by contact or noncontact thermometry. For the case of the contact temperature measurement only, a detailed measurement and data reduction procedure is given in Ref. [3]. However, in order to measure the properties at very high temperatures, but still keeping advantages of the contact thermometry, a simple principle in four stages is proposed here: Measure the specimen temperature by using simultaneously contact and noncontact thermometry and find all the three properties of the specimen, as well as its spectral normal emissivity, up to the limit of the used contact temperature sensor Tcmax (~ 2600 K for the C type thermocouple, for example) [4]; extrapolate the obtained values on spectral normal emissivity above this temperature limit for each experimental run; convert the measured radiance temperature, Tλ, to the absolute specimen temperature, T, and, finally, find the values of other three properties over the extended temperature range. A flowchart of such a measurement procedure is given in Fig. 1, where ρ is the specific electrical resistivity, εth total hemispherical emissivity, cp specific heat and ελn spectral normal emissivity of the specimen material.

Fig. 1
figure 1

Flowchart of the proposed measurement procedure

As short wavelengths (λ < 1 μm) are usual for a high temperature detection of the radiance temperature, the spectral normal emissivity, derived from Planck’s law and the spectral radiance of a gray body [10], may be reduced to \(\varepsilon_{{\lambda {\text{n}}}} \approx c_{2} \lambda^{ - 1} {\text{e}}^{{c_{2} /\lambda \left( {T^{ - 1} - T_{\lambda }^{ - 1} } \right)}}\), where c2 is the second radiation constant. From this formula, if both measured and extrapolated spectral normal emissivity values are written as \(\varepsilon^{\prime}_{{\lambda {\text{n}}}} \left( T \right)\) and the measured radiance temperature in function of time as \({T}_{\lambda }\left(t\right)\), the conversion to the absolute temperature above Tcmax is made by:

$$ T\left( t \right) = \frac{{T_{\lambda } \left( t \right)}}{{1 + T_{\lambda } \left( t \right)\frac{\lambda }{{c_{2} }}\ln \left[ {\varepsilon^{\prime}_{{\lambda {\text{n}}}} \left( T \right)} \right]}} $$
(1)

As this equation is intrinsic in regards to the absolute temperature T, it must be solved numerically. If all the spectral normal emissivity values is fitted, for example, by a linear function

$$ \varepsilon^{\prime}_{{\lambda {\text{n}}}} \left( T \right) = a_{0} + a_{1} T $$
(2)

the iterative equation for computing the absolute temperature values from Eq. (1) may be found by the Newton–Rapson method [11] as

$$ T_{\text {k} + 1} \left( {t_{\text i} } \right) = T_{\text k} \left( {t_{\text i} } \right) - \frac{{a_{0} + a_{\text 1} T_{\text k} \left( {t_{\text i} } \right) - {\text{e}}^{{\frac{{{\text c}_{2} }}{\uplambda }\left[ {\frac{1}{{T_{\text k} \left( {t_{\text i} } \right)}} - \frac{1}{{T_{\uplambda } \left( {t_{\text i} } \right)}}} \right]}} }}{{a_{\text 1} + \frac{{\text c_{2} }}{{\lambda T_{\text k}^{2} \left( {t_{\text i} } \right)}}{\text{e}}^{{\frac{{\text c_{2} }}{\uplambda }\left[ {\frac{1}{{T_{\text k} \left( {t_{\text i} } \right)}} - \frac{1}{{T_{\uplambda } \left( {t_{\text i} } \right)}}} \right]}} }} $$
(3)

where k is the iteration number, and i is the time step above Tcmax. For higher order fittings of \(\varepsilon^{\prime}_{{\lambda {\text{n}}}}\) values, related iterative equations may be accordingly derived.

Experimental

Setup

The proposed measurement principle has been applied by using an experimental facility based on the subsecond calorimetry with both contact and noncontact thermometry. The facility consists of special specimen holders, cooling unit, environmental chamber, high current pulse circuitry, sensing and detection unit and data acquisition and control system. Schematic of the facility is presented in Fig. 2, and its detailed description is given in Refs. [3, 4].

Fig. 2
figure 2

Schematic of the experimental setup

Specimen

A specimen of pure polycrystalline tungsten was provided by Goodfellow Cambridge Ltd. It had the form of a thin rod, 3 mm in diameter and 200 mm in length, nominally. Its chemical purity was claimed to be greater than 99.95% by mass. The specimen was manually polished by using fine grade sandpapers. According to the final measured dimensions and mass of the specimen and their related uncertainties, the specimen density was estimated to be (19.27 ± 0.03) kg m−3 in comparison with 19.25 kg m−3 of pure tungsten [12].

Measurements

The specimen was placed between the clamps inside the chamber where the air was evacuated to about 10–4 mbar. Its stationary specimen temperature was maintained by conduction from the coolers to about 0 °C. For both transient and stationary contact temperature measurement, thermocouple leads of C type (W95%/Re5%–W74%/Re26%) and 70 µm in diameter were intrinsically welded on the specimen center. The voltage drop was measured by thin wires of pure tungsten, also 70 µm in diameter and welded at the distance of about 20 mm, which was the actual length of the specimen effective part. The radiance temperature was measured by a high-speed pyrometer, having the nominal wavelength of 900 nm and a powerful focusing optics of a less than 2 mm field of view at a distance of 400 mm. The pyrometer was focalized in a close vicinity to the welded thermocouple leads. As heat radiation from the specimen was detected from outside the chamber, a window’s measured transparency of 0.58 at 900 nm was included in the computation of the specimen radiance temperature. One and two in series high current, 12 V dc power sources were used for the specimen heating with approximate heating rates of 700 K s−1 and 2000 K s−1, respectively.

Results

In total, 40 experimental runs have been performed on the same specimen. Heating time and, therefore, the maximum specimen temperature, was gradually increased in order to determine an adequate number of the total hemispherical emissivity values. First 23 runs were made up to Tcmax ≈ 2600 K, which is the limit of the C type thermocouple standard function, while the rest of 17 runs were all recorded at temperatures above 2600 K, i.e., up to about 3700 K.

By first using the data reduction procedure from Ref. [4], the non-averaged values of ρ obtained from the measurement of voltage drop, current and absolute specimen temperature (first 23 runs, up to Tcmax) are given in Fig. 3, while those of εth, together with their linear fit are given in Fig. 4 (Stage I, black circles). The latter ones were used for the calculation of cp (Fig. 5, black circles). On the other hand, the values of ελn from a typical set of absolute and radiance temperature data up to Tcmax and their linear extrapolation above this temperature, Eq. (2), are presented in Fig. 6 (Stage II).

Fig. 3
figure 3

The values of ρ up to and above Tcmax

Fig. 4
figure 4

The values of εth up to and above Tcmax

Fig. 5
figure 5

The values of cp up to and above Tcmax

Fig. 6
figure 6

The typical values of ελn up to Tcmax and their linear extrapolation above Tcmax

Having the estimated values of ελn and radiance temperature converted to the absolute temperature for each experimental run above Tcmax from Eq. (3) (Stage III), the values of specific electrical resistivity and total hemispherical emissivity were computed in the higher temperature range (Stage IV, Figs. 3 and 4, gray circles). At last, the values of εth were re-smoothed with a new polynomial fitFootnote 2 and used for the determination of specific heat in the whole temperature range (Stage IV). The final non-averaged values of cp in the higher temperature range are presented in Fig. 5, gray circles.

Discussion

In order to evaluate correctness of the proposed principle and measurement, the obtained results are fitted and compared with the available high temperature literature data of pure tungsten. First, a comparison among the fitted data of specific electrical resistivityFootnote 3 with those recommended by Desai et al. [13] is given in Fig. 7. A high level of coincidence can be observed within the whole temperature range, except at highest temperatures, where a difference of about 3% emerges between the recommended and actual results. Such the difference, however, may be overridden by the final total measurement uncertainty of ρ at these temperatures.

Fig. 7
figure 7

Comparison among the obtained values of ρ and those recommended

Considering the total hemispherical emissivity, both its non-averaged and fitted values are compared with several sets of available literature data [14,15,16,17,18,19,20,21,22,23,24,25,26], as presented in Fig. 8. It can be seen a good agreement among actual and several sets of literature data, especially in their tendency with a temperature change. Finally, for the specific heat, a comparison among the obtained values and the available literature data [27,28,29,30,31,32,33,34,35,36,37,38,39,40,41] is presented in Fig. 9. Most literature data reveal an uprise of specific heat values with temperature, so do the actual measurements. After their further reduction with a polynomial fit,Footnote 4 it can be observed that the final results of specific heat coincide very well with most literature data, i.e., within less than 10% at highest temperatures.

Fig. 8
figure 8

Comparison among the obtained values of εth and those from the literature

Fig. 9
figure 9

Comparison among the obtained values of cp and those from the literature

With regard to the measurement uncertainty, Stage II is crucial: Results of the extrapolation directly influence the converted temperature values of Stage III, which further makes a direct impact to the final values of Stage IV. In other words, uncertainty of the extrapolated spectral normal emissivity above Tcmax directly contributes to the total horizontal uncertainty of three measured properties, i.e., to their temperature uncertainty. As this contribution may be particularly important in cases when specific heat and specific electrical resistivity values strongly depend on temperature, special care should be taken in practice for the realization of Stage II of the proposed measurement procedure.

Conclusions

The standard use of the subsecond calorimetry technique implies the temperature measurements either by radiation or contact thermometry. While the both methods of temperature measurements have their advantages and disadvantages, this work proposes their simultaneous use which allows a measurement of the specimen spectral normal emissivity up to the temperature limit of the contact thermometry. Then, above that limit, the obtained values of the spectral normal emissivity are extrapolated, letting the recorded radiance temperature to be converted to the absolute specimen temperature. The latter values are finally used to estimate the specific electrical resistivity, total hemispherical emissivity and specific heat at very high temperatures.

The proposed principle has been tested on a specimen of pure polycrystalline tungsten. Obtained results of all the measured properties show good agreement with related literature data, but also suggest that radiance to absolute temperature conversion needs to be considered carefully due to its potentially significant influence on the final temperature uncertainty of presented results.