Introduction

The unique spiral-structure of the cochlea is essential to its function as the hearing sensory organ. It transduces physical fluid pressure waves into neural impulses that can be interpreted by the brain to sense and understand the acoustic environment. Although this shell-like structure is vital to one of our most essential senses, relatively little has been done to manufacture an artificial cochlea. In vivo experiments have provided a wealth of information about the mechanisms of hearing [9, 52, 57, 125]; however, artificial models could provide a platform to further understand and address many of the remaining challenges in repairing hearing impairments.

Hearing impairment is the most prevalent sensory deficit in humans, affecting 466 million people worldwide (World Health Organisation) [153]. Cochlear implants (CIs) have been transformative for those suffering from severe-to-profound hearing loss by bypassing normal acoustic hearing mechanisms and directly stimulating the cochlear nerve electrically. However, key limitations still constrain the clinical effectiveness and wider eligibility of these implants. Mechanical trauma generated during CI insertion and the resulting tissue trauma and chronic inflammatory response can damage residual acoustic hearing. Residual acoustic hearing can be beneficial when combined with CI electrical hearing (electro-acoustic hearing), and any interventions to preserve this would increase eligibility for CIs [37, 64, 66, 136]. Current means for detailed physical examination of the electrode-cochlear interactions involve animal [26, 86, 156] and human cadaveric testing [32, 72, 104]. However, these models present significant challenges as they cannot be easily instrumented or modified in shape and size to allow systematic testing of performance parameters. Furthermore, animal cochlea are very different anatomically from human cochleae [123] and also have restrictions on the availability and ethical considerations in their use. Only limited progress has been made towards other models, such as engineering realistic artificial cochleae [88].

A bio-mimetic cochlea has the potential to accelerate both the development of new treatments for hearing loss as well as optimise existing treatments. We evaluate requirements for a realistic model and assess current attempts to engineer artificial cochleae, their limitations, and indicate future directions for their development.

Fig. 1
figure 1

Diagram depicting the structure of the cochlea (right) and the insertion of the CI through the round window into the scala tympani. This is shown in relation to its position relative to the middle ear and external components of the CI (left). Note that spiral ganglion neurons from the cochlear nerve spiral around the 2.75 turns of the cochlea around the central axis (the modiolus)

Biological Background of Cochlea

Anatomy

Embedded in the temporal bone, the cochlea is a fluid-filled structure that is part of the osseous (bony) labyrinth, also referred to as the otic capsule. This consists of the semi-circular canals, responsible for sensing head rotation which is essential for balance, the vestibule, which houses the linear acceleration detectors (the otolith organs), and the cochlea itself (see Fig. 1).

Neural representation of frequencies in the normal cochlea are structured in a tonotopic manner, primarily by the intrinsic passive and active tuning of the basilar membrane, which maximises vibrations for particular frequencies in a graded manner from apex to base. This results in neurons (specifically spiral ganglion neurons localised in the Rosenthal’s canal) along the length of the cochlear spiral encoding low frequencies at the apex and high frequencies towards the base (in a range from 20 Hz to 20 kHz) [47]. CIs try to somewhat replicate this tonotopic representation, by presenting lower frequency signals to the apical electrodes and higher frequency ones to the basal electrodes.

A defining feature of the cochlea is its distinct ascending spiral geometry. One reason for the nautilus shell-like structure is thought to be due to spatial constraints in the temporal bone [113]. However, more recent studies have presented evidence suggesting that it also provides functional benefits. For instance, a study by Manoussaki et al. indicates that the graded curvature of the cochlea can aid the propagation of low frequencies similar to “whispering gallery nodes” and therefore influence low-frequency hearing limits [96].

Key parts of the cochlea for surgeons are the bony round window (RW) niche and the round window itself, as they are the most common entry portal for CI insertion (see Fig. 2a). The RW niche is a bony pouch of the tympanic cavity located anterior to the RW which is closed with a membrane. The niche has a width and depth of about \(1.66\pm 0.34\) mm and \(1.34\pm 0.25\) mm, respectively [137]. The RW is a small, circular opening with a transverse diameter of about \(1.65\pm 0.21\) mm positioned inferior and slightly posterior to the oval window at an average distance of 2–2.2 mm. It is covered by a thin membrane \((69.4\pm 4.3 {\upmu}\hbox {m}\)) [126] called the round window membrane (RWM), which enables fluid movement within the cochlea during auditory stimulation [68, 133, 137]. The oval window is closed by the stapes footplate, a part of one of the three ossicles in the middle ear which transfer vibrations of the eardrum to the inner ear fluids. The oval window is set in the bony vestibule. During auditory stimulation, the stapes footplate vibrates, creating inner ear pressure which is released by the compliant RW membrane.

Variation in Cochlear Anatomy

Human cochleae display large variations in both size and shape [9, 140], which likely affects the clinical performance of CIs. It is, therefore, crucial to understand this variation when attempting to produce a representative artificial model. Indeed, by developing physical (i.e. model made of plastic or similar material that represents anatomically accurate “mechanical" twin of the human cochlea) or computational models that can represent the variation present in the human cochlea, we can try to understand the relationship between different structural features of the cochlea and the effectiveness of CIs. This could reveal possibilities for personalised/stratified medicine within this field by relating different device designs as being optimal for certain types of cochleae, particularly concerning the length of the cochlea, but also to other material properties such as stiffness of the implant in different dimensions for optimal, minimally traumatic full insertion.

Additionally, through a deeper comprehension of these disparities, researchers can enhance the utilisation of animal models by acknowledging the anatomical variations between the human and animal cochlea.

Characterisation of the Cochlea Size and Shape

The cochlea has a characteristic spiral geometry which flares out from a typical Archimedean spiral at the base. This has been characterised by either a piecewise function [24, 157], which separately describes the base and apex or a continuous double exponential function [16, 61]. Additionally, a function to determine the height of the cochlea is highly dependent on the reference frame used. For instance, when using a mid-modiolar axis, it is possible to determine that the height of the cochlear centerline increases linearly [24]. This can lead to the observation of three cochlea shape categories: sloping, intermediate, and ‘rollercoaster’ [9]. A sloping shape has an upward trajectory without any significant downward trend. The intermediate shape follows a slight upward trajectory after the RW, which then follows a slight decrease. Lastly, the ‘rollercoaster’ shape follows a downward trajectory after the RW, followed by an upward trajectory around 75–120°. Nevertheless, Gee et al. argue that by defining a basal plane sets the average height of the first 270° of the basal turn equivalent to the perspective of the inserted cochlear implant (CI) then the “rollercoaster” trajectory is not observed [43]. This definition results in a more sigmoidal increase of the cochlear height where the basal turn is rather flat, followed by a sharp defined rising of the cochlear spiral [43].

Size and Shape Variation of Cochlea

The variability of the cochlea is reflected in the range of cochlear duct lengths ranging from 30.8 to 43.2 mm [83, 155] and also, to a lesser extent, with the variable number of turns [9, 13, 49]. Hence, individual cochleae are not only a scaled-up version of the same basic shape but represent true morphologic variations. Therefore, different shapes and lengths of the electrode arrays should ideally be considered if atraumatic insertion, place-pitch matching (aligning the sound frequencies assigned to electrodes to the natural biologic tonotopic map), and preservation of residual hearing are of interest. Table 1 summarises some of the key anatomical features of the cochlea.

Furthermore, the height of the centre part of the scala tympani (ST) is larger than in the lateral and modiolar regions. The ST is the lumen into which CIs are placed [9]. Lateral wall height significantly decreases following the second turn (after 450°) [9, 74], which can imply a higher possibility of CI translocation through the basilar membrane from the ST (intended site) into the scala vestibuli or pushing against the basilar membrane. This likely increases the probability of residual hearing loss and perhaps also worse CI hearing outcomes, if the trauma affects the auditory neurons. Figure 2c depicts the changes in the cross-section of the cochlea and the width and thickness of its key membranes at the base and apex. The basilar membrane (excluding Organ of Corti) becomes thinner and wider towards the apex of the cochlea, ranging from 1.26 to 1.92 μm, thickness and ~ 202 μm width to 0.53–0.89 and 475 μm, respectively [99]. However, the thickness and width of the Reissner’s membrane remain consistent throughout the cochlear duct length at about \(6.4\pm 2.6\) μm and \(770\pm 180\) μm, respectively [31].

The basal turn represents a large part of the cochlear duct length. Its size variability is, therefore, a significant contributor to the overall CI insertion path and may drastically impact the angular insertion depth of the implant [38, 133]. Furthermore, the majority of insertion trauma is observable within this region (180–270°  from the RW) [1, 8, 33] which underlines its importance.

In addition to the major axis change between the first and ascending portions of the basal turn, several smaller anatomical peaks, dips and vertical jumps have been described in the vertical trajectory of the ST [9, 113]. These relatively sudden changes in the vertical trajectory of the ST can be critical when calculating the required implant insertion depth in 3D. Therefore, it is essential to replicate them in detail in the physical or computational artificial model as they may play a significant role in determining the insertion interactions between the cochlea and the implant. Each cochlear shape category is unique and, in theory, could require slightly different approaches in terms of insertion site and angle to minimise trauma. However, recent evidence shows a relative independence of the overall friction force experienced by the implant to the cochlea shape [61], although shape may still play a role in the local stresses and trajectories of the implant, and where this force is concentrated.

Fig. 2
figure 2

Limited surgical view of the round window during implantation (a) and Otic Capsule with cochlear duct length measurement (b). Cross-section of the cochlea at base and apex highlighting key anatomical features (c). Th thickness, W width

Variations of Round Window

The RW also demonstrates large variability in shapes between implant recipients with majority being of oval-shape (see Table 1) [29, 95, 130, 132]. A sharp bony crest called the crista fenestra, which occupies an extensive area projecting into the RW, may play a significant role as a barrier to the ST [29]. Different RW niche morphologies produce various sizes of crests [6]. In some situations, removing the crest is necessary to introduce the implant successfully. If the RW is unreachable or the entry angle is not satisfactory, a cochleostomy (separate hole drilled into the ST) is typically considered. Additionally, the surgical view of the RW is limited by the margins of facial recess, i.e. facial nerve posterior, chorda tympani, eardrum anterior, and incus buttress superiorly (see Fig. 2a). Therefore, the surgeon’s manoeuvrability is restricted and could benefit from the experience acquired by training on a physical phantom of the human cochlea.

Table 1 Summary of the size and shape of various features of human cochleae

Reproduction of Cochleae

Models of the cochlea can be broadly classified into four categories: physical models, animal models, tissue engineering models, and computational models. Physical models are useful for investigating the mechanical aspects of CI insertion and the electrical properties of the electrode-nerve interface [74, 88, 105]. These models are often created using materials that mimic the mechanical and electrical properties of the cochlea and can be modified to study the influence of specific parameters on CI behaviour. Animal models, such as guinea pigs [4, 51, 58] and chinchillas [145], are prevalent in CI research due to their ability to replicate physiological responses to stimulation, or damage. Tissue engineered models are created using living cells and could be potentially used to study the effects of CIs on cells and tissues in more biologically realistic environments than 2D cultures on tissue culture plastic. In general, tissue-engineered models are often constructed using scaffolds that support the growth and organisation of cells into tissues. For cochleae, there have been examples of organoid cultures [84] and decellularised tissues [100] for cochlear tissue engineering although more extensive studies of producing replacement tissues have been discussed for the middle and outer ear [3, 5, 22]. Lastly, computational models are increasingly being used to study the electrical and mechanical properties of the cochlea and the effects of CIs on the auditory system [7, 14, 110, 118]. These models can be used to replicate the complex biological environment of the cochlea in a more controlled and reproducible manner and one that can be easily modified to study specific parameters. Computational models could also be used to predict the performance of CIs in different scenarios, such as differing electrode configurations or stimulation strategies.

The aforementioned models often work in tandem as the acquired data from physical, animal, and tissue-engineered models can be subsequently fed into computational simulation models to accelerate research and examine variables that would otherwise be substantially time-consuming to study. These models play an imperative role in validating computational simulations, as the simulations fail without their tangible data and confirmation.

The following sections discuss various techniques to produce physical artificial cochlea models and the utilisation of animal models, tissue engineering models, and computational simulations.

Physical Models

The development of an anatomically accurate model of the human cochlea is of great interest to researchers studying the mechanical aspects of CI insertion (e.g. atraumatic implantation and insertion trajectories) and electrical properties in optimising the electrode-nerve interface (e.g. simulating nerve activation with different stimulation strategies and electrode positions). For example, having a model that can reliably measure insertion forces and register it with implant position over time (e.g. insertion depth) delivers information that can influence surgical practice. Studying the behaviour of electrode arrays within the cochlea can improve CI design and introduce an individualised approach by understanding how a specific implant might behave for a given recipient based on their cochlear shape and size, allowing a personalised selection of implants to possibly minimise insertion forces, optimised electrode position, or avoid basilar membrane contact.

The properties of the artificial cochlea model may vary based on the experiment. For instance, a transparent physical model with a smooth intracochlear lumen with embedded sensors is required to evaluate the intracochlear pressure or insertion forces [54, 102, 105, 124]. The transparency of the model is essential as it allows direct visualisation of the implant behaviour during insertion [89]. The insertion forces can be measured using a force sensor that can be attached to the cochlear model (often multi-axis measurement) [62, 89, 109], between the electrode array and the insertion device (one-axis) [82, 85]. Alternatively, in the case of an “open-channel" artificial model (a model with only the basal turn of the ST fully open at the top surface), the overall force at the location of the basilar membrane/ osseous spiral lamina can be measured with flat force sensor or membrane [54].

Nevertheless, in order to create these models, it is vital to understand the mechanical and electrical properties of the cochlea. Table 2 displays key mechanical data that should be considered when fabricating an artificial cochlea. For instance, the load required to rupture the basilar membrane, which is in the basal turn and apical turn about 35 and 26 mN [67], respectively, is of great interest when mimicking CI insertion as its penetration can result in the translocation of the CI into the scala media and scala vestibuli, which is frequently associated with the loss of residual hearing, and worse CI function [57, 74, 81, 111, 150, 154].

Table 2 Approximate values for the mechanical properties of different components of the cochlea

Additionally, if we are able to replicate the electrical properties (e.g. resistivity) of the cochlear bone and its contained fluids (see Table 3), this would greatly help in understanding current spread during cochlear electric stimulation. Having artificial models can vary significantly increase the number of repeated experiments that can be performed in a standardised model [61, 88], as opposed to biological tissues, and thus the robustness of any conclusions; these models are mechanically more durable than cadaveric specimens and do not degrade quickly over time. Furthermore, the use of fixed cadaveric tissues may alter the electrical and mechanical properties of the sample, and the availability of fresh cadavers is often limited. Artificial models can also be altered to change one parameter at a time in order to explore the influence of specific parameters. However, replicating the electrical properties of a cochlea has proven to be difficult. Artificial cochleae could be created from bone-like material to better mimic in vivo tissue; nevertheless, these materials are yet to be fully characterised for high-resolution additive manufacturing. One interesting solution is that 3D printing can be used to fabricate models with an appropriate size of embedded pores filled with a conductive solution that would enable fine-tuning the electrical properties of the material [28, 88, 106].

Table 3 Approximate values for the electrical properties of selected anatomical elements of the cochlea

Several methods can be utilised for the development of artificial cochleae (see Figs. 3 and 4).

Fig. 3
figure 3

Workflow of corrosion casting method. This method uses curable resins that fill the hollow otic capsule within the temporal bone, which is digested to leave the cured resin that replicates the otic capsule space. Negative moulds utilise a double casting method, using the initial cast as a mould which is subsequently removed to leave the hollow lumen of the cochlea

Fig. 4
figure 4

Workflow of 3D printing and CNC machining artificial cochlea. a Registration and segmentation of the otic capsule from microCT scans of the temporal bone can be used to generate CAD files of the cochlear structure. b CAD files can be manipulated to make various geometries for 3D printing. For resin-based 3D printing, scaffolding and slicing the model are required prior to printing shape-accurate models. c CNC—compatible files can be derived from cochlear CAD models that can be programmed to machine planar cochlear models

Casting

For a long time, casting has been the prevalent technique for developing artificial cochleae. Several studies have utilised polymethylmethacrylate (PMMA, often known as acrylic glass or plexiglass) as the casted material for the fabrication of 2D [109, 122] and 3D [18, 82, 89, 102, 103, 122] cochleae models. The advantage of this material is its transparency which provides good visualisation of the array’s behaviour during insertion. However, the 2D models offer only limited information as they lack the true three-dimensional form [24].

Rebscher et al. described a multi-casting process exploiting low-melting-point alloy (LMA) and PMMA for the development of an artificial ST (see Fig. 3) [122]. They produced multiple high-accuracy replicas of the same temporal bone by utilising vulcanising silicone rubber, which functioned as a mould. Firstly, the PMMA was injected into a dissected temporal bone and subsequently cooled down. To remove the ST cast from the cadaveric specimen, the temporal bone was decalcified. Following that, the PMMA cast was then covered by vulcanising silicone creating mould. After the curing process of the silicone, the mould was carefully divided into two parts to release the original PMMA cast. In the next step, LMA was poured into the silicone mould to create LMA-casted replicas. Lastly, LMA casts were covered by PMMA. Once PMMA cooled down, the LMA was released (lower melting point than PMMA) to fabricate a “block" model of the ST.

This process of multi-casting is relatively complicated and time-consuming, in addition to the work needed for the dissection of temporal bones from cadaveric specimens. Moreover, the number of fabricated models using Rebscher’s method is limited by the number of dissected samples, and the models’ shape cannot be easily modified or adjusted, as it is based on the anatomical specimen itself. Many studies have exploited casted cochlea models from companies such as Advanced Bionics (AB), MED-EL or Cochlear [24, 82, 89, 93]. However, dimensions and anatomic accuracy of these models can be sub-optimal compared to other manufacturing techniques, although this has not been systematically quantified yet [24, 89]. As multi-casting involves many steps, each inherently introducing a level of variability, it can be assumed that the anatomic accuracy will suffer as small features could be difficult to reproduce.

In addition to producing cochlear models, corrosion casting has been used to study various anatomical features of the cochlea. For example, Carraro et al. have used partial corrosion casting to study the vasculature within the cochlea [20, 21]. By perfusing the vasculature with a castable commercial resin, Mercox II, and digesting tissue, it was possible to preserve the vascular structure in mouse cochleae and study it with scanning electron microscopy.

Additive Manufacturing

Additive manufacturing, such as 3D printing, is now a well-proven technology that has the potential to develop highly accurate artificial models of human cochleae. Nevertheless, not all 3D printing techniques are suitable for developing such models. Firstly, some 3D printing technologies, such as fused deposition modelling (FDM) or selective laser sintering (SLS), do not produce products with the level of detail required for cochlear research. This is a result of the materials used as well as the nature of the technique. SLS exploits a laser to sinter plastic particles (often nylon-based material) into a solid structure. Although SLS does not require support generation during printing as the powder supports the print, its limitation is the printing resolution, which is suboptimal for prints of the inner ear very small size. FDM may utilise transparent materials (e.g. PLA); however, it fabricates products by heating a filament and building it layer-by-layer, which may generate a step-like finish with a low level of smoothness. The printing resolution is, therefore, significantly dependent on the layer’s height. This limitation may further result in suboptimal transparency of the product as each layer scatters light. Secondly, the cochlea anatomy is complex as it contains overhangs, tunnels, and hollow structures, and its fabrication may require temporary supportive scaffolds. If the supports are erected within the cochlea lumen, the smoothness of the inside structure may be compromised. Following that, after the removal of the supportive scaffold, the print is frequently polished to obtain better surface smoothness. However, due to the complex anatomy and inaccessibility; this is not easily achievable in the cochlea.

Some 3D printing techniques such as stereolithography (SLA, see Fig. 4b) [18, 39, 54, 89, 108], digital light processing (DLP) [61], and polyjet printing (PJP) [24, 89] have been already exploited for the fabrication of the ST models with much better accuracy (<40 \({\upmu }\hbox {m}\) [61]). SLA and DLP are photopolymer-based technologies that use ultraviolet light to cure resin (liquid plastics) into 3D prints. In the case of SLA, a system of mirrors focusses the laser into a small spot that is subsequently moved over the printing plane to cure each layer. A single accurate light source provides good smoothness as each printed layer is merged with the previous one. In addition, the layer-merging process also decreases the number of needed supports. DLP uses a projector with UV light to cure the whole layer at once, which enables faster printing but might produce a step-like finish with a too-high layer height. For SLA and DLP, the printed product must be further processed after the printing by washing in isopropanol and curing using a UV-light chamber to obtain the highest possible quality. The optimal printing quality of these techniques comes with a trade-off, as the only supported materials are photopolymer-based. Furthermore, SLA and DLP enable the printing of the product from one material only. However, for example, the material’s conductive properties might be tuned by introducing microchannels or pores for modifying the electrical conductivity of the construct, used when studying CI stimulation electrical properties [88].

PJP techniques, on the other hand, allow the printing of multiple materials at once and, thus, fabricating products with various mechanical properties (e.g. flexible and rigid). It exploits thermoplastics that are heated and then deposited on a platform layer-by-layer in the form of droplets using multiple print heads. Moreover, the supportive scaffolds can be printed out of soluble materials, which are easier to remove than SLA and DLP supports which are made of the same materials as the print.

Leon et al. used both PJP and SLA printing techniques to fabricate an artificial model of ST and compared them to models from CI companies MED-EL (SLA printed), AB (multi-step casting) and Cochlear (2D planar model) [89]. They observed that the measured insertion forces in the SLA model were similar to Cochlear’s model, which demonstrates good inner surface smoothness. However, the PJP model demonstrated even lower insertion forces than the SLA printed model, implying an improved internal surface finish. The disadvantage of the PJP model was its semi-transparency which was not optimal for successful visualisation of the implant behaviour during insertion. Hence, they recommend the SLA technique due to its ability to fabricate transparent models. However, PJP technique can achieve enhanced transparency with the use of appropriate material, for example, VeroClear (Stratasys) [112].

It is important to bear in mind that 3D printing capabilities progress rapidly. Several new materials are developed each year, enhancing 3D printing abilities and providing new avenues for prototyping [23, 116]. Hence, different 3D printing techniques can be recommended for the fabrication of artificial cochlea each year as the material’s selection vary.

Other Manufacturing Methods

Some studies have used 2D polytetrafluorethylene (PTFE) artificial cochleae [85, 124]. This material was used due to its low friction coefficient, which was found to be comparable to the slippery endosteum of the ST [79, 144]. These models can be prepared by using computer numerical control (CNC, see Fig. 4c) machines that carve the material in a 2D plane using precise drills. Although the PTFE material provides optimal smoothness of the internal surface of the model, the lack of the third dimension significantly alters the electrode array’s behaviour during insertion [24, 89].

Electro-Acoustic Models

As there are many aspects of cochlear physiology that are of interest to researchers, there are a variety of different models to replicate these different aspects. In addition to the physical mentioned above, some have investigated modelling electro-acoustic aspects of the cochlea. These typically attempt to replicate the sensory epithelium of the cochlea using electro-active materials such as piezoelectric membranes and micromechanical systems [63, 65, 70, 146, 159]. Replicating the high-frequency selectivity (20 Hz–20 kHz) and sensitivity, sound pressure level range (0–140 dB SPL) as well as the small size and power requirements of the human cochlea, presents a substantial technical challenge; however, work continues towards the aim of restoring the range and specificity of natural hearing with CIs [19, 107]. This may involve atraumatic insertion covering the full extent of cochlear spiral and an increasing specificity of the neural activation. Currently, devices have demonstrated some limited tonotopy within the range of human speech, typically in the ~ 1.4 to 14 kHz range with rather high 70 dB+ sound pressure levels [70, 159]. Other reports have used alternative methods such as triboelectric devices to detect lower frequency ranges from ~ 300 to 2000 Hz in in vivo conditions with Guinea pigs [69] with idealised conditions in other studies lowering the minimum frequency to the tens of hertz [91].

Animal Models

Animal models are well established in CI research and have benefits as models in replicating the complex structure of the cochlea and its constituent tissues. By conducting in vivo experiments, it is possible to use the features of intact myelinated primary auditory neurons/ spiral ganglion neurons within the cochlea, an immunological response which is important in understanding chronic issues such as fibrosis and ossification [25, 26, 131], and potential for conducting some behavioural studies and measuring electrically evoked potentials [35, 46, 50, 80] to conduct a wide range of CI studies to understand the CI-nerve interface.

There are, however, several considerations in the applicability of different animal models for human CI research in terms of both their anatomy and physiology. The cochlea varies in size, shape, and complexity amongst different species, with differences in the length, width, and number of turns [73, 78, 123]. Although the overall scalar structure is largely conserved in mammals, the overall shape and size do not scale linearly with overall body size, indicating that factors other than body size determine the cochlea’s structure [78]. As most animals would have much smaller cochleae than humans, smaller custom-made small CIs will need to be used, which increases the complexity of conducting and extrapolating results from animal models [86, 92, 121]. Furthermore, the structure of the cochlea will determine the spread of the electric field from a CI and, hence, the neural activation which limits their effectiveness in answering some of the key unanswered questions in the CI field such as the spread of neural activation to different CI parameters.

Several different animal models have been used for studies that are interested in structure-related parameters and CI implantation. Rodent models such as mice, rats, gerbils, guinea pigs, ferrets, and chinchillas have been widely used in CI research due to their availability, potential for instrumentation, and established gene-editing tools in the case of mice and rats; as reviewed by Reiss [123]. Guinea pigs, in particular, have been extensively used due to their wide availability, their inner ears being easy to access, and their cochleae being more comparable to the human cochlea in several physiological aspects [4, 51, 58, 86, 147]. Chinchillas have been considered an even better model due to the similarities of their cochlea to humans with regards to its number of turns, hearing range, and sensitivity [145]. Cats are the most popular non-rodent model due to their basal turn being of a similar size to human cochlear [123], although, the overall shape and size of their cochleae differ significantly from humans. Moreover, the lack of myelin in the soma regions of human type I primary auditory neurons causes a delay in spike conduction compared to cat neurons [119], which can impact the transmission of temporal fine structure of auditory signals in the human cochlea. Larger animal models include sheep [128] and miniature pigs [156] as well as marmoset [73] and rhesus macaque monkeys [97, 123]. These are closer to the size of human cochleae, with the marmoset cochlear shape being most similar when scaled up by a factor of 2.5 [73], and allow the partial implantation of clinical electrodes although these model are much less available than small rodents and do not well replicate many of the specific features of human cochleae which are not even featured in other primates [151]. For comparison, the ratio of the volume of the human cochlea to different animal cochleae is as follows: mouse 80–100:1, rat 20–25:1, gerbil 5–6:1, cat 2–3:1, macaque 2–3:1, and sheep ~ 1.7:1 [123].

It is important to consider that the use of animal models is complementary to measures that can be conducted in human patients and cadaveric tissues. For instance, several studies have characterised the electrical properties [32, 72] and conducted CI insertion studies [18, 75, 77] in both fresh-frozen and fixed human cochleae. Furthermore, many electrophysiological and psychoacoustic measures have been developed to test the CI-nerve interface in humans, which include contact impedance and trans-impedance measurements [59, 87, 138], electrically evoked compound action potentials (eCAPS) [27, 41, 42], and electrically evoked auditory brainstem responses (eABR) [17, 36] which enable the evaluation of CI electrical characteristics (including fibrosis, positioning, and electrical faults), cochlear neural activation patterns, and propagation the CI stimulation to the brain, respectively. However, human studies do not allow us to systematically test numerous parameters in the same experiment, such as electrode position, stimulation parameters, pulse shapes, the geometry of the cochlea, and electrode design, size, and shape, which are much more possible in other models.

In conclusion, animal models are valuable in conducting validation of CI techniques and the systemic responses to CI implantation. However, the limitations of these different models should be considered, especially in light of reducing animal use and used in conjunction with human measures and other models, as will be explored below.

Tissue Engineering

Several animal models have been utilised to study the cochlea and develop strategies to improve CI performance. Yet, it remains unclear how to improve electrical stimulation and how different stimulation strategies could affect neural excitation. This necessity led to the use of tissue engineering in the hearing field.

Tissue engineering is a set of methods that can replace or repair damaged or diseased tissues with natural, synthetic, or semi-synthetic tissues which can be fully functional or will grow into the required functionality [143]. These methods could, in theory, be utilised for replicating the complex three-dimensional cellular architecture of the cochlea in vitro. Furthermore, they could serve as useful platforms for studying cellular viability and expression in various conditions.

Two important cell types in the cochlea are hair cells (HCs) and primary auditory neurons (PANs), also known as spiral ganglion cells (SGNs). In the mammalian cochlea, HCs serve to sense the mechanical movement, amplify it and transmit this signal to the auditory nerve [44, 135]. PANs act as the neural conduit transmitting cochlear HC signals to the brain [30].

Some research has focussed on culturing auditory cells obtained from animals or differentiated from induced pluripotent stem cells (iPSCs). Much work has been conducted in vitro, which mostly use mouse, rat, and guinea pig sources for auditory HCs and PANs. Although it has been quite challenging to obtain these cells in significant numbers and maintain them over time, especially when requiring specific purified cell populations, recent studies have demonstrated some success [98, 117]. A few studies have progressed to generating human inner ear cells (e.g. IHCs and OHCs) from iPSCs [45, 60, 71]. It has been more challenging to differentiate PANs since the concerns would involve the electrical activity, firing potentials, and the possession of appropriate ion channels as well as the gene expression profiles.

Whilst in vitro models have facilitated the understanding of cellular mechanisms within the cochlea, they are limited in replicating the complexity of the in vivo micro-environment.

By combining data from the flexibility and specificity of in vitro experiments, systematic effects and replication of live structures of in vivo studies, and the clinical relevance of cadaveric studies, much has been learned about the cochlea and the impact of CIs. However, as all of these approaches have their limitations, there is an unmet need for in vitro platforms for hearing research. The cellular and molecular aspects of the cochlea could be integrated into a 3D model, which would complement the limitations of the previous models. This in vitro platform could mimic the main functional aspects of the cochlea, including the current spread profile. If incorporated as a host to human iPSC-derived cells, this model would not only reduce the time and cost required for testing but also eliminate the need for experiments on living creatures to study cochlear biology and determine the efficiency and reliability of new drugs or technologies, e.g. CIs, for hearing research.

Computational Models

There is a large field of computational audiology that has been used to effectively model several aspects of the cochlea which we will briefly overview in this review. In terms of modelling physical aspects of the cochlea, these can broadly be categorised as electrical and mechanical models.

Electrical models of the cochlea are focussed on optimising the electrical implant-nerve interfaces that underlie the function of a CI and have been reviewed extensively by others [2, 48, 76]. These models primarily consist of two main aspects: (1) modelling of the electrical voltage spread within the cochlea, and (2) biophysical and phenomenological models of the neural excitation of auditory nerve fibres.

For the 3D electrical characteristics of the cochlea, there has been extensive work in developing finite element models of the electrical stimulation of CIs that have been established by the groups of Frijns and Rattay [14, 118]. These have gradually increased in complexity from simpler parametric representations of the cochlear spiral to microCT-based models that also incorporate the trajectories of auditory neurons [11, 55, 115]. As well as understanding the electrical properties of the cochlea, these finite element models have also been utilised for impedance-guided insertion to determine the CI positioning within the cochlea from electrical measurements [127]. These finite element models can be coupled to multi-physics simulations such as thermal safety analyses of intracochlear heating with magnetically steered CIs [39]. As an alternative to finite element models, simpler circuit models of the cochlea have been developed, such as ladder network models, to model specific phenomena [148].

Biophysical models of neural activation are extensions of the foundational work of Hodgkin and Huxley [56]. As discussed by Bachmaier [10], the use of multi-compartmental biophysical models of myelinated nerve fibres is able to replicate many phenomena observed in patients such as the sensitivity of the auditory nerves to the polarity of stimulation [15, 118, 134].

In contrast to the biophysical approach, phenomenological models do not rely on specific biophysical mechanisms and derive empirical relationships based on neurophysiological and psychophysical observations [141]. Due to the much-reduced parameter space, this approach allows the efficient modelling of complex phenomena that can be adjusted to individual CI patients and has proven effective at predicting and explaining a diverse range of auditory phenomena [139, 141, 142].

Combining the 3D volume conduction models with neuronal models can be a compelling method to investigate the effect of various parameters of the electrode-nerve interface for CIs. These enable the investigation of the effect of different stimulation parameters and positioning on auditory nerve fibre activation [11, 55, 94, 115]. Recent studies have demonstrated the coupling of neural activation from these models to an automatic speech recognition neural network to predict phoneme-level speech perception and information transmission [16].

The mechanics of the cochlea have been extensively studied since the pioneering work of von Békésy [149]. The extensive work in the mathematical and computational modelling of the basilar membrane micromechanics that underlie the mechanism of acoustic hearing is reviewed by Ni and colleagues [110]. Finally, mechanical models can provide insight into the insertion forces during cochlear implantation, which can lead to significant trauma and inflammatory response, damaging residual hearing [7].

In conclusion, computational models can be powerful tools to facilitate the understanding of the physical phenomena within the cochlea. However, these models require the correct inputs for measured quantities that can often be difficult to derive and need validation with experimental data to ensure that the model is accurate.

Tables 12, and 3 summarise some of the essential information regarding cochlea physical, electrical, and mechanical attributes that can be used for computational simulations. Table 4 compares the merits and drawbacks of the examined cochlea models.

Table 4 Advantages and disadvantages of discussed types of cochlea models

Future Perspectives

Future research will focus on developing anatomically accurate artificial cochlea with embedded force and pressure sensors to detect insertion forces that arise during CI implantation. Preserving the residual hearing will aid the further development of EAS implants as natural acoustic stimulation is yet to be exceeded in performance by the electrical stimulation. Additionally, these models could also be utilised for studying inner ear therapeutics and drug delivery systems. Accurate, transparent cochlea models could help precisely determine the pharmacokinetics of drugs delivered inside the inner ear and their spread over time.

Combining cell-based models with animal models can lead to a more comprehensive understanding of CIs and improve the design of safe and effective treatments for auditory disorders. Cell-based models can be used to simulate the electrical and mechanical properties of the cochlea and to study how different stimulation parameters affect the auditory nerve. This information can then be used to guide the design of animal experiments, such as determining the optimal stimulation parameters to use in vivo. Animal models can be used to validate and refine the cell-based models and verify their accuracy in replicating the in vivo response of the cochlea to CIs. A cell-based 3D model of the cochlea could also play an important role in understanding the pathophysiology and aetiology of auditory disorders as well as allowing the interpretation of electric fields of the electrode arrays of CIs in the cochlea by bio-mimicking the true cochlear physiology. Despite the limitations of animal models, they still have advantages in certain areas, such as tracking the systemic response to cochlear implantation and aiding in the development of new therapeutic approaches to mitigate potential adverse effects.

The data from in vivo and in vitro experiments enables us to validate and inform the design of computational models to understand the mechanisms of the CI-auditory nerve interface. These computational models have increased in complexity over the last 20 years in development to combine finite element models with auditory nerve models to test a variety of clinically relevant parameters and help devise new stimulation strategies. Further development in this field may enable personalised approaches to replicate an individual’s specific cochlear anatomy and CI interface to improve their performance rather than generic procedures. Additionally, the dynamic time-dependent component of CI stimulation, rather than purely resistive finite element models, could allow further insights into the validity of using specific stimulation parameters to improve focussed auditory nerve stimulation.

Ultimately, by combining the insights from patients, cadavers, animals, in vitro experiments, physical models, and computational models it is possible to account for their individual limitations and build a more comprehensive understanding of optimal CI application for patient benefit.

Conclusion

The elaborate and intricate structure of the human auditory system is a marvel that cannot yet be matched by modern engineering. However, understanding the cochlear structure and how to interact with the delicate system is crucial in addressing huge challenges in otology and audiology. Improving models and understanding the cochlea will lay the foundation for developing the next generation of CIs and future inner ear therapies. These implants and treatments should address the major challenges of insertion trauma and current spread to preserve cochlear health and residual hearing whilst conveying high sound fidelity by improving the spatial selectivity of stimulation.

Furthermore, understanding the variability of cochlea’s anatomy and its effect on insertion parameters and CI performance could open up the capability of personalised approaches for individual cases to deliver optimal patient outcomes. Addressing these challenges will widen the eligibility for CIs and improve the lives of the growing proportion of people suffering from hearing loss.

Ultimately, a 3D in vitro model of the cochlea with integrated auditory cells would revolutionise the study of various features of the inner ear to support the development of new technologies and the validation of computational simulations and drug-based therapies.