Purpose

Plate-haptic rotationally asymmetric multifocal toric intraocular lenses (IOL) (Lentis Comfort Toric) occasionally rotate extensively after surgery. We conducted the current study to investigate the incidence of extensive misalignment of this IOL and its association with clinical parameters.

Study design

Retrospective case series.

Subjects and methods

Data were collected from patients who had undergone phacoemulsification and implantation of a plate-haptic multifocal toric IOL.

Results

Among 332 eyes, extensive misalignment of toric IOLs ≥ 45º occurred in 3.3% (11 eyes). The amount of misalignment in eyes with extensive misalignment was 81.6 ± 22.9º, while in those without extensive misalignment, it was 3.0 ± 2.7º. The eyes with extensive misalignment showed significantly greater axial length (p < 0.001), larger corneal diameter (p = 0.034), and flatter corneas (p = 0.044) than those without extensive misalignment. Repositioning surgery to correct toric IOL misorientation was conducted in 9 eyes between 7 and 28 days after cataract surgery. In 2 eyes, repositioning surgery was carried out twice.

Conclusions

In majority of cases plate-haptic multifocal toric IOLs showed satisfactory rotational stability, but extensive misalignment ≥ 45º occurred in 3.3% of cases.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1

Similar content being viewed by others

References

  1. Lee BS, Chang DF. Comparison of the rotational stability of two toric intraocular lenses in 1273 consecutive eyes. Ophthalmology. 2018;125:1325–31.

    Article  PubMed  Google Scholar 

  2. Kramer BA, Hardten DR, Berdahl JP. Rotation characteristics of three toric monofocal intraocular lenses. Clin Ophthalmol. 2020;14:4379–84.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  3. Oshika T, Inamura M, Inoue Y, Ohashi T, Sugita T, Fujita Y, et al. Incidence and outcomes of repositioning surgery to correct misalignment of toric intraocular lenses. Ophthalmology. 2018;125:31–5.

    Article  PubMed  Google Scholar 

  4. Lee BS, Onishi AC, Chang DF. Comparison of rotational stability and repositioning rates of 2 presbyopia-correcting and 2 monofocal toric intraocular lenses. J Cataract Refract Surg. 2021;47:622–6.

    Article  PubMed  Google Scholar 

  5. Takaku R, Nakano S, Iida M, Oshika T. Influence of frosted haptics on rotational stability of toric intraocular lenses. Sci Rep. 2021;11:15099.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  6. Osawa R, Sano M, Yuguchi T, Kaiya T, Oshika T. Effects of modified haptics on surgical outcomes and rotational stability of toric intraocular lens implantation. J Refract Surg. 2022;38:648–53.

    Article  PubMed  Google Scholar 

  7. Venter J, Pelouskova M. Outcomes and complications of a multifocal toric intraocular lens with a surface-embedded near section. J Cataract Refract Surg. 2013;39:859–66.

    Article  PubMed  Google Scholar 

  8. Garzón N, Poyales F, de Zárate BO, Ruiz-García JL, Quiroga JA. Evaluation of rotation and visual outcomes after implantation of monofocal and multifocal toric intraocular lenses. J Refract Surg. 2015;31:90–7.

    Article  PubMed  Google Scholar 

  9. Shodai R, Negishi K, Arai H, Toda I, Torii H, Tsubota K. Comparative analysis of the visual and refractive outcomes of a refractive segmented multifocal intraocular lens with and without toricity: 1-year results. Jpn J Ophthalmol. 2017;61:142–9.

    Article  PubMed  Google Scholar 

  10. Oshika T, Arai H, Fujita Y, Inamura M, Inoue Y, Noda T, et al. One-year clinical evaluation of rotationally asymmetric multifocal intraocular lens with + 1.5 diopters near addition. Sci Rep. 2019;9:13117.

    Article  PubMed  PubMed Central  Google Scholar 

  11. Zhu X, Meng J, He W, Rong X, Lu Y. Comparison of the rotational stability between plate-haptic toric and C-loop haptic toric IOLs in myopic eyes. J Cataract Refract Surg. 2020;46:1353–9.

    Article  PubMed  Google Scholar 

  12. Amigó A, Bonaque-González S. Is the capsular bag perimeter round or elliptical? J Ophthalmic Vis Res. 2016;11:159–61.

    Article  PubMed  PubMed Central  Google Scholar 

  13. Yao Y, Meng J, He W, Zhang K, Wei L, Cheng K, et al. Associations between anterior segment parameters and rotational stability of a plate-haptic toric intraocular lens. J Cataract Refract Surg. 2021;47:1436–40.

    Article  PubMed  Google Scholar 

  14. Lucisano A, Ferrise M, Balestrieri M, Busin M, Scorcia V. Evaluation of postoperative toric intraocular lens alignment with anterior segment optical coherence tomography. J Cataract Refract Surg. 2017;43:1007–9.

    Article  PubMed  Google Scholar 

  15. Khng C, Osher RH. Evaluation of the relationship between corneal diameter and lens diameter. J Cataract Refract Surg. 2008;34:475–9.

    Article  PubMed  Google Scholar 

  16. Oh J, Shin HH, Kim JH, Kim HM, Song JS. Direct measurement of the ciliary sulcus diameter by 35-megahertz ultrasound biomicroscopy. Ophthalmology. 2007;114:1685–8.

    Article  PubMed  Google Scholar 

  17. Marchini G, Pedrotti E, Modesti M, Visentin S, Tosi R. Anterior segment changes during accommodation in eyes with a monofocal intraocular lens: high-frequency ultrasound study. J Cataract Refract Surg. 2008;34:949–56.

    Article  PubMed  Google Scholar 

  18. Shah GD, Praveen MR, Vasavada AR, Vasavada VA, Rampal G, Shastry LR. Rotational stability of a toric intraocular lens: influence of axial length and alignment in the capsular bag. J Cataract Refract Surg. 2012;38:54–9.

    Article  PubMed  Google Scholar 

  19. Zhu X, He W, Zhang K, Lu Y. Factors influencing 1-year rotational stability of AcrySof Toric intraocular lenses. Br J Ophthalmol. 2016;100:263–8.

    Article  PubMed  Google Scholar 

  20. Patnaik JL, Kahook MY, Berdahl JP, Hardten DR, Wagner BD, Seibold LK, et al. Association between axial length and toric intraocular lens rotation according to an online toric back-calculator. Int J Ophthalmol. 2022;15:420–5.

    Article  PubMed  PubMed Central  Google Scholar 

  21. Rastogi A, Khanam S, Goel Y, Kamlesh, Thacker P, Kumar P. Comparative evaluation of rotational stability and visual outcome of toric intraocular lenses with and without a capsular tension ring. Indian J Ophthalmol. 2018;66:411–5.

    Article  PubMed  PubMed Central  Google Scholar 

  22. Zhao Y, Li J, Yang K, Li X, Zhu S. Combined special capsular tension ring and toric IOL implantation for management of astigmatism and high axial myopia with cataracts. Semin Ophthalmol. 2018;33:389–94.

    Article  PubMed  Google Scholar 

  23. Hahn U, Krummenauer F, Schmickler S, Koch J. Rotation of a toric intraocular lens with and without capsular tension ring: data from a multicenter non-inferiority randomized clinical trial (RCT). BMC Ophthalmol. 2019;19:143.

    Article  PubMed  PubMed Central  Google Scholar 

  24. Ma D, Han X, Hua Z, Shen J, Zhang L, Qiu T, et al. Propensity-matched comparison of postoperative stability and visual outcomes of toric intraocular lens with or without a capsular tension ring and updated meta-analysis. Graefes Arch Clin Exp Ophthalmol. 2023;261:989–98.

    Article  PubMed  Google Scholar 

  25. Sasaki K, Eguchi S, Miyata A, Nishimura T, Miyata K, Hasegawa Y, et al. Anterior capsule coverage and rotational stability of an acrylic toric intraocular lens. J Cataract Refract Surg. 2021;47:618–21.

    Article  PubMed  Google Scholar 

  26. Prinz A, Neumayer T, Buehl W, Vock L, Menapace R, Findl O, et al. Rotational stability and posterior capsule opacification of a plate-haptic and an open-loop-haptic intraocular lens. J Cataract Refract Surg. 2011;37:251–7.

    Article  PubMed  Google Scholar 

  27. Ruhswurm I, Scholz U, Zehetmayer M, Hanselmayer G, Vass C, Skorpik C. Astigmatism correction with a foldable toric intraocular lens in cataract patients. J Cataract Refract Surg. 2000;26:1022–7.

    Article  CAS  PubMed  Google Scholar 

  28. Till JS, Yoder PR Jr, Wilcox TK, Spielman JL. Toric intraocular lens implantation: 100 consecutive cases. J Cataract Refract Surg. 2002;28:295–301.

    Article  PubMed  Google Scholar 

  29. Sun XY, Vicary D, Montgomery P, Griffiths M. Toric intraocular lenses for correcting astigmatism in 130 eyes. Ophthalmology. 2000;107:1776–82.

    Article  CAS  PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Tetsuro Oshika.

Ethics declarations

Conflict of interest

T. Oshika, Grant to the author’s institution (Santen), Consulting fees, Honoraria for lecture fee, Participation on Advisory Board (Santen); S. Nakano, None; M. Inamura, None; N. Ikai, None; Y. Kato, None; I. Izumi, None; K. Shimokawabe, None.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Corresponding Author: Tetsuro Oshika

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Oshika, T., Nakano, S., Inamura, M. et al. Extensive misalignment of plate-haptic rotationally asymmetric multifocal toric intraocular lens. Jpn J Ophthalmol 67, 560–564 (2023). https://doi.org/10.1007/s10384-023-01001-x

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s10384-023-01001-x

Keywords

Navigation