Orientation of the intermediate-vision zone of rotationally asymmetric multifocal intraocular lenses and photic phenomena

  • Clinical Investigation
  • Published:
Japanese Journal of Ophthalmology Aims and scope Submit manuscript

Abstract

Purpose

To evaluate the association between the orientation of the intermediate-vision zone and postoperative photic phenomena in eyes implanted with rotationally asymmetric multifocal intraocular lenses with + 1.5 diopters near addition.

Study design

Retrospective comparative case series.

Methods

We performed the photic phenomena test (PPT) to quantitatively evaluate photic phenomena at 1–3 months after cataract surgery with Lentis Comfort LS-313 MF15 or Lentis Comfort Toric LS-313 MF15T insertion (Oculentis; Santen Pharmaceutical). In cases of bilateral surgeries, only the right eye was included in the analysis. We also conducted a questionnaire survey about the photic phenomena.

Results

The study included 96 eyes from 96 patients. The intermediate-vision zones of 44, 19, 19, and 14 eyes were fixed in the inferior, superior, nasal, and temporal directions, respectively. The measured PPT values for halo, glare, and starburst were equivalent in all the fixed directions. Eighty-four eyes (87.5%) perceived a triangle-shaped halo, and the direction of fixation for the intermediate-vision zone and the extension direction of this photic phenomenon were consistent in 94.0% of the eyes. Less than 10% of the patients responded “moderate” or “severe” for the triangle-shaped halo, glare, and starburst in fixed directions with low frequency in the questionnaire.

Conclusion

No qualitative or quantitative difference was found in photic phenomena in eyes implanted with the LS-313 MF15 or the MF15T regardless of the fixation direction. Our study revealed that many patients were aware of a triangle-shaped halo extending in the direction of the intermediate-vision zone.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Institutional subscriptions

Fig. 1
Fig. 2
Fig. 3
Fig. 4

Similar content being viewed by others

References

  1. Yoo A, Kwag JY, Song IS, Kim MJ, Jeong H, Kim JY, et al. Comparison of visual function after implantation of inferior sector-shaped intraocular lenses: low-add + 1.5 D vs + 3.0 D. Eur J Ophthalmol. 2016;26:607–11.

    Article  PubMed  Google Scholar 

  2. Vounotrypidis E, Diener R, Wertheimer C, Kreutzer T, Wolf A, Priglinger S, et al. Bifocal nondiffractive intraocular lens for enhanced depth of focus in correcting presbyopia: clinical evaluation. J Cataract Refract Surg. 2017;43:627–32.

    Article  PubMed  Google Scholar 

  3. Oshika T, Arai H, Fujita Y, Inamura M, Inoue Y, Noda T, et al. One-year clinical evaluation of rotationally asymmetric multifocal intraocular lens with + 1.5 diopters near addition. Sci Rep. 2019;9:13117.

    Article  PubMed  PubMed Central  Google Scholar 

  4. Pedrotti E, Mastropasqua R, Bonetto J, Demasi C, Aiello F, Nucci C, et al. Quality of vision, patient satisfaction and long-term visual function after bilateral implantation of a low addition multifocal intraocular lens. Int Ophthalmol. 2018;38:1709–16.

    Article  PubMed  Google Scholar 

  5. Kim KH, Kim WS. Visual outcome and patient satisfaction of low-power-added multifocal intraocular lens. Eye Contact Lens. 2018;44:60–7.

    Article  PubMed  Google Scholar 

  6. Song IS, Yoon SY, Kim JY, Kim MJ, Tchah H. Influence of near-segment positioning in a rotationally asymmetric multifocal intraocular lens. J Refract Surg. 2016;32:238–43.

    Article  PubMed  Google Scholar 

  7. Bala C, Meades K. Improvement in vision with inverted placement of an asymmetric refractive multifocal intraocular lens. J Cataract Refract Surg. 2014;40:833–5.

    Article  PubMed  Google Scholar 

  8. Ukai Y, Okemoto H, Seki Y, Nakatsugawa Y, Kawasaki A, Shibata T, et al. Quantitative assessment of photic phenomena in the presbyopia-correcting intraocular lens. PLoS One. 2021;16:e0260406.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  9. Son HS, Khoramnia R, Yildirim TM, Baur I, Labuz G, Auffarth GU. Functional outcomes and reading performance after combined implantation of a small-aperture lens and a segmental refractive bifocal lens. J Refract Surg. 2019;35:551–8.

    Article  PubMed  Google Scholar 

  10. Buckhurst PJ, Naroo SA, Davies LN, Shah S, Drew T, Wolffsohn JS. Assessment of dysphotopsia in pseudophakic subjects with multifocal intraocular lenses. BMJ Open Ophthalmol. 2017;1:e000064.

    Article  PubMed  PubMed Central  Google Scholar 

  11. Hayashi K, Hayashi H, Nakao F, Hayashi F. Correlation between pupillary size and intraocular lens decentration and visual acuity of a zonal-progressive multifocal lens and a monofocal lens. Ophthalmology. 2001;108:2011–7.

    Article  CAS  PubMed  Google Scholar 

  12. Koch DD, Samuelson SW, Haft EA, Merin LM. Pupillary size and responsiveness: implications for selection of a bifocal intraocular lens. Ophthalmology. 1991;98:1030–5.

    Article  CAS  PubMed  Google Scholar 

  13. Kawamorita T, Uozato H, Aizawa D, Kamiya K, Shimizu K. Optical performance in rezoom and array multifocal intraocular lenses in vitro. J Refract Surg. 2009;25:467–9.

    Article  PubMed  Google Scholar 

  14. Kanellopoulos AJ, Asimellis G. Clear-cornea cataract surgery: pupil size and shape changes, along with anterior chamber volume and depth changes. A Scheimpflug imaging study. Clin Ophthalmol. 2014;24:2141–50.

    Article  Google Scholar 

  15. Montés-Micó R, López-Gil N, Pérez-Vives C, Bonaque S, Ferrer-Blasco T. In vitro optical performance of nonrotational symmetric and refractive-diffractive aspheric multifocal intraocular lenses: impact of tilt and decentration. J Cataract Refract Surg. 2012;38:1657–63.

    Article  PubMed  Google Scholar 

  16. Alio JL, Plaza-Puche AB, Javaloy J, Ayala MJ, Moreno LJ, Piñero DP. Comparison of a new refractive multifocal intraocular lens with an inferior segmental near add and a diffractive multifocal intraocular lens. Ophthalmology. 2012;119:555–63.

    Article  PubMed  Google Scholar 

  17. de Wit DW, Diaz J, Moore TC, Moutari S, Moore JE. Effect of position of near addition in an asymmetric refractive multifocal intraocular lens on quality of vision. J Cataract Refract Surg. 2015;41:945–55.

    Article  PubMed  Google Scholar 

Download references

Acknowledgements

We would like to thank Editage (www.editage.com) for English language editing.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Tsuyoshi Mito.

Ethics declarations

Conflicts of interest

T. Mito, None; Y. Ukai, None; Y. Nakatsugawa, None; T. Shibata, None; S. Shibata, None; E. Kubo, None; H. Sasaki, None.

Additional information

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Corresponding Author: Tsuyoshi Mito

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Mito, T., Ukai, Y., Nakatsugawa, Y. et al. Orientation of the intermediate-vision zone of rotationally asymmetric multifocal intraocular lenses and photic phenomena. Jpn J Ophthalmol 67, 318–325 (2023). https://doi.org/10.1007/s10384-023-00988-7

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s10384-023-00988-7

Keywords

Navigation