The goal of this study was to evaluate the performance of a computer-aided detection (CAD) system in full-field digital mammography (Senographe 2000D, General Electric, Buc, France) in finding out carcinomas depending on the parenchymal density. A total of 226 mediolateral oblique (MLO) and 186 craniocaudal (CC) mammographic views of histologically proven cancers were retrospectively evaluated with a digital CAD system (ImageChecker V2.3 R2 Technology, Los Altos, CA, USA). Malignant tumors were detected correctly by CAD in MLO view in 84.85% in breasts with parenchymal tissue density of the American College of Radiology (ACR) type 1, in 70.33% of the ACR type 2, in 68.12% of the ACR type 3, and in 69.70% of the ACR type 4. For the CC view, similar results were found according to the ACR types. Using the chi-square and McNemar tests, there was no statistical significance. However, a trend of better detection could be seen with decreasing ACR type. In conclusion, there seems to be a tendency for breast tissue density to affect the detection rate of breast cancer when using the CAD system.
Similar content being viewed by others
References
M Moskowitz (1983) ArticleTitleScreening for breast cancer: how effective are our test? A critical review CA Cancer J Clin 33 26–39 Occurrence Handle6401574 Occurrence Handle1:STN:280:DyaL3s7gt1CrsA%3D%3D
J Rennie R Rusting (1996) ArticleTitleMaking headway against cancer Sci Am 3 56–59 Occurrence Handle10.1038/scientificamerican0996-56
RE Bird (1990) ArticleTitleProfessional quality assurance for mammography screening programs Radiology 77 8–10
HP Chan K Doi CJ Vyborny RA Schmidt CE Metz KL Lam T Ogura YZ Wu H MacMahon (1990) ArticleTitleImprovement in radiologists' detection of clustered microcalcifications on mammograms. The potential of computer-aided diagnosis Invest Radiol 25 1102–1110 Occurrence Handle2079409 Occurrence Handle1:STN:280:DyaK3M7nslKrsg%3D%3D Occurrence Handle10.1097/00004424-199010000-00006
B Zheng YH Chang M Staiger W Good D Gur (1995) ArticleTitleComputer-aided detection of clustered microcalcifications in digitized mammograms Acad Radiol 2 655–662 Occurrence Handle9419620 Occurrence Handle1:STN:280:DyaK1c%2FotVyqsg%3D%3D Occurrence Handle10.1016/S1076-6332(05)80431-3
HP Chan B Sahiner MA Helvie N Petrick MA Roubidoux TE Wilson DD Adler C Paramagul JS Newman S Sanjay-Gopal (1999) ArticleTitleImprovement of radiologists' characterization of mammographic masses by using computer-aided diagnosis: an ROC study Radiology 212 817–827 Occurrence Handle10478252 Occurrence Handle1:STN:280:DyaK1MvgslKqsg%3D%3D
EL Thurfjell KA Lernevall AA Taube (1994) ArticleTitleBenefit of independent double reading in a population-based mammography screening program Radiology 191 241–244 Occurrence Handle8134580 Occurrence Handle1:STN:280:DyaK2c7otFOrtw%3D%3D
A Malich C Marx M Facius T Boehm M Fleck WA Kaiser (2001) ArticleTitleTumour detection rate of a new commercially available computer-aided detection system Eur Radiol 11 2454–2459 Occurrence Handle11734939 Occurrence Handle1:STN:280:DC%2BD3MnovVaitw%3D%3D Occurrence Handle10.1007/s003300101079
CJ Vyborny (1994) ArticleTitleCan computers help radiologists read mammograms? Radiology 191 315–317 Occurrence Handle8153298 Occurrence Handle1:STN:280:DyaK2c3gsVWqtA%3D%3D
K Doi H MacMahon S Katsuragawa RM Nishikawa Y Jiang (1999) ArticleTitleComputer-aided diagnosis in radiology: potential and pitfalls Eur Radiol 31 97–109 Occurrence Handle1:STN:280:DC%2BD3c%2FivFaguw%3D%3D Occurrence Handle10.1016/S0720-048X(99)00016-9
M Funovics S Schamp TH Helbich B Lackner P Wunderbaldinger M Fuchsjäger G Lechner G Wolf (2001) ArticleTitleEvaluation of a computer-assisted diagnosis system in breast carcinoma Röfo Fortschr Geb Röntgenstr Neuen Bildgeb Verfahr 173 218–223 Occurrence Handle11293863 Occurrence Handle1:STN:280:DC%2BD3M3ivV2lsA%3D%3D Occurrence Handle10.1055/s-2001-11608
S Muller (1999) ArticleTitleFull field digital mammography designed as a complete system Eur Radiol 31 25–34 Occurrence Handle1:STN:280:DyaK1MvgsVOntw%3D%3D Occurrence Handle10.1016/S0720-048X(99)00066-2
U Bick (2000) ArticleTitleFull-field digital mammography Röfo Fortschr Geb Röntgenstr Neuen Bildgeb Verfahr 173 957–964 Occurrence Handle10.1055/s-2000-9210
S Obenauer S Luftner-Nagel D Heyden Particlevon U Munzel F Baum E Grabbe (2002) ArticleTitleScreen-film vs full-field digital mammography: image quality, detectability and characterization of lesions Eur Radiol 12 1697–1702 Occurrence Handle12111060 Occurrence Handle1:STN:280:DC%2BD38zms1Ghuw%3D%3D Occurrence Handle10.1007/s00330-001-1269-y
F Baum U Fischer S Obenauer E Grabbe (2002) ArticleTitleComputer-aided detection in direct digital full-field mammography: initial results Eur Radiol 12 3015–3017 Occurrence Handle12439584 Occurrence Handle1:STN:280:DC%2BD38nnvVWrsw%3D%3D
S Satija RH Moore JS Michalson et al. (2000) ArticleTitleBreast tissue patterns as a risk factor for developing cancer (abstract) Radiology 217 446
InstitutionalAuthorNameAmerican College of Radiology (ACR) (1998) Illustrated breast imaging reporting and data system (BI-RADS) EditionNumber3 American College of Radiology Reston, VA
MT Mandelson N Oestreicher PL Porter D White CA Finder SH Taplin E White (2000) ArticleTitleBreast density as a predictor of mammographic detection: comparison of interval- and screen-detected cancers J Natl Cancer Inst 92 1081–1087 Occurrence Handle10880551 Occurrence Handle1:STN:280:DC%2BD3cvivFyrtA%3D%3D Occurrence Handle10.1093/jnci/92.13.1081
LJ Warren Burhenne SA Wood CJ D'Orsi SA Feig DB Kopans KF O'Shaughnessy EA Sickles L Tabar CJ Vyborny RA Castellino (2000) ArticleTitlePotential contribution of computer-aided detection to the sensitivity of screening mammography Radiology 215 554–562 Occurrence Handle10796939 Occurrence Handle1:STN:280:DC%2BD3c3lslSjtw%3D%3D
RF Brem J Baum M Lechner S Kaplan S Souders LG Naul J Hoffmeister (2003) ArticleTitleImprovement in sensitivity of screening mammography with computer-aided detection: a multiinstitutional trial Am J Roentgenol 181 687–693
JA Baker EL Rosen JY Lo EI Gimenez R Walsh MS Soo (2003) ArticleTitleComputer-aided detection (CAD) in screening mammography: sensitivity of commercial CAD systems for detecting architectural distortion Am J Roentgenol 181 1083–1088
WT Ho PW Lam (2003) ArticleTitleClinical performance of computer-assisted detection (CAD) system in detecting carcinoma in breasts of different densities Clin Radiol 58 133–136 Occurrence Handle12623042 Occurrence Handle1:STN:280:DC%2BD3s7htVCmsw%3D%3D Occurrence Handle10.1053/crad.2002.1131
RL Birdwell DM Ikeda KF O'Shaugnessy EA Sickles (2001) ArticleTitleMammographic characteristics of 115 missed cancers later detected with screening mammography and the potential utility of computer-aided detection Radiology 219 192–202 Occurrence Handle11274556 Occurrence Handle1:STN:280:DC%2BD3M7oslKgsQ%3D%3D
RF Brem JW Hoffmeister JA Rapelyea G Zisman K Mohtashemi G Jindal MP Disimio SK Rogers (2005) ArticleTitleImpact of breast density on computer-aided detection for breast cancer Am J Roentgenol 184 439–444
A Malich DR Fischer M Facius A Petrovich J Boettcher C Marx A Hansch WA Kaiser (2005) ArticleTitleEffect of breast density on computer aided detection J Digit Imaging 18 227–233 Occurrence Handle15827823 Occurrence Handle10.1007/s10278-004-1047-x
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Obenauer, S., Sohns, C., Werner, C. et al. Impact of Breast Density on Computer-Aided Detection in Full-Field Digital Mammography. J Digit Imaging 19, 258–263 (2006). https://doi.org/10.1007/s10278-006-0592-x
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10278-006-0592-x