Screw-in force, torque generation, and performance of glide-path files with three rotation kinetics

  • Original Article
  • Published:
Odontology Aims and scope Submit manuscript

Abstract

We aimed to evaluate the screw-in force, torque generation, and performance of three nickel-titanium (NiTi) glide-path files with different rotational kinetics. ProTaper Ultimate Slider (PULS) and HyFlex EDM Glide-path (HEDG) files were used for canal shaping with constant rotation (CON) or the alternative rotation technique (ART). In the ART mode, the NiTi file was periodically rotated at a speed of 1.5 times faster than that in the CON mode. WaveOne Gold Glider was used with reciprocating motion (WOGG_RCP). Sixty J-shaped resin blocks were assigned to five groups: PULS_CON, PULS_ART, HEDG_CON, HEDG_ART, and WOGG_RCP (n = 12). Glide-path preparation was performed using an automated pecking device. During glide-path preparation, the screw-in force and clockwise and counterclockwise torques were recorded and the number of pecking motions required to reach the working length was determined. The centering ratio was calculated after glide-path preparation using stereomicroscopic images. Data were analyzed using one-way analysis of variance with the Games–Howell post hoc test and the Kruskal–Wallis test with Bonferroni correction. PULS_ART generated a lower maximum screw-in force than PULS_CON. The average number of pecking motions required to reach the working length by HEDG_ART was lower than that by HEDG_CON. The mean centering ratios of PULS_CON and HEDG_CON were − 0.04 and − 0.06, respectively, while those of PULS_ART, HEDG_ART, and WOGG_RCP were 0.09, 0.01, and 0.08, respectively. The ART mode reduced the screw-in force of PULS and enabled faster glide-path preparation with the HEDG file.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5
Fig. 6

Similar content being viewed by others

Data availability

The data that support the findings of this study are available from the corresponding author upon reasonable request.

References

  1. De-Deus G, Cardoso ML, Simões-Carvalho M, et al. Glide path with reciprocating driven pathfinding instrument: performance and fracture rate. J Endod. 2021;47:100–4.

    Article  PubMed  Google Scholar 

  2. West JD. The endodontic Glidepath: “Secret to rotary safety.” Dent Today. 2010;29:86 (8, 90–3).

    PubMed  Google Scholar 

  3. Alovisi M, Cemenasco A, Mancini L, et al. Micro-CT evaluation of several glide path techniques and ProTaper Next shaping outcomes in maxillary first molar curved canals. Int Endod J. 2017;50:387–97.

    Article  PubMed  Google Scholar 

  4. Biasillo V, Castagnola R, Colangeli M, et al. Comparison of shaping ability of the Reciproc Blue and One Curve with or without glide path in simulated S-shaped root canals. Restor Dent Endod. 2022;47:e3.

    Article  PubMed  PubMed Central  Google Scholar 

  5. van der Vyver PJ, Paleker F, Vorster M, de Wet FA. Micro-computed tomographic evaluation of two single rotary glide path systems. Int Endod J. 2019;52:352–8.

    Article  PubMed  Google Scholar 

  6. D’Amario M, Baldi M, Petricca R, De Angelis F, El Abed R, D’Arcangelo C. Evaluation of a new nickel-titanium system to create the glide path in root canal preparation of curved canals. J Endod. 2013;39:1581–4.

    Article  PubMed  Google Scholar 

  7. Uslu G, İnan U. Effect of glide path preparation with PathFile and ProGlider on the cyclic fatigue resistance of WaveOne nickel-titanium files. Restor Dent Endod. 2019;44:e22.

    Article  PubMed  PubMed Central  Google Scholar 

  8. Topçuoğlu HS, Topçuoğlu G, Düzgün S. Resistance to cyclic fatigue of PathFile, ScoutRaCe and ProGlider glide path files in an S-shaped canal. Int Endod J. 2018;51:509–14.

    Article  PubMed  Google Scholar 

  9. Martins JNR, Silva E, Marques D, et al. Characterization of the file-specific heat-treated ProTaper ultimate rotary system. Int Endod J. 2023;56:530–42.

    Article  PubMed  Google Scholar 

  10. Ruddle CJ. The ultimate shaping system an opening for 3D cleaning and filling root canals. Advanced endodontics [Internet]. 2022 May 01, 2023. https://www.endoruddle.com/tc2pdfs/show/253/ProTaperUltimate_May2022.pdf

  11. Thu M, Ebihara A, Maki K, et al. Dynamic torque and screw-in force of four different glide path instruments assessed in simulated single- and double-curved canals: an in vitro study. J Dent Sci. 2023;18:1598–603.

    Article  PubMed  PubMed Central  Google Scholar 

  12. Htun PH, Ebihara A, Maki K, et al. Comparison of torque, screw-in force, and shaping ability of glide path instruments in continuous rotation and optimum glide path motion. J Endod. 2021;47:94–9.

    Article  PubMed  Google Scholar 

  13. Oh S, Seo JY, Lee JE, Kim HJ, Jang JH, Chang SW. Evaluation of design, mechanical properties, and torque/force generation of heat-treated NiTi glide path instruments. BMC Oral Health. 2022;22:528.

    Article  PubMed  PubMed Central  Google Scholar 

  14. Dias PS, Kato AS, Bueno C, et al. Comparative analysis of torsional and cyclic fatigue resistance of ProGlider, WaveOne Gold Glider, and TruNatomy Glider in simulated curved canal. Restor Dent Endod. 2023;48:e4.

    Article  PubMed  Google Scholar 

  15. Kobayashi C, Yoshioka T, Suda H. A new engine-driven canal preparation system with electronic canal measuring capability. J Endod. 1997;23:751–4.

    Article  PubMed  Google Scholar 

  16. Kimura S, Ebihara A, Maki K, Nishijo M, Tokita D, Okiji T. Effect of optimum torque reverse motion on torque and force generation during root canal instrumentation with crown-down and single-length techniques. J Endod. 2020;46:232–7.

    Article  PubMed  Google Scholar 

  17. Aydın ZU, Keskin NB, Özyürek T, Geneci F, Ocak M, Çelik HH. Microcomputed assessment of transportation, centering ratio, canal area, and volume increase after single-file rotary and reciprocating glide path instrumentation in curved root canals: a laboratory study. J Endod. 2019;45:791–6.

    Article  PubMed  Google Scholar 

  18. Campos DS, Rodrigues EA, Bueno C, et al. The ability of reciprocating glide path instruments to reach the full root canal working length. Aust Endod J. 2021;47:487–92.

    Article  PubMed  Google Scholar 

  19. Keskin C, İnan U, Demiral M, Keleş A. Cyclic fatigue resistance of R-Pilot, WaveOne Gold Glider, and ProGlider glide path instruments. Clin Oral Investig. 2018;22:3007–12.

    Article  PubMed  Google Scholar 

  20. Htun PH, Ebihara A, Maki K, et al. Comparison of torque, force generation and canal shaping ability between manual and nickel-titanium glide path instruments in rotary and optimum glide path motion. Odontology. 2020;108:188–93.

    Article  PubMed  Google Scholar 

  21. Kyaw MS, Ebihara A, Maki K, et al. Effect of kinematics on the torque/force generation, surface characteristics, and shaping ability of a nickel-titanium rotary glide path instrument: an ex vivo study. Int Endod J. 2022;55:531–43.

    Article  PubMed  Google Scholar 

  22. Metabiomed. EQ-M [Available from: https://www.meta- biomed.com/bbs/board.php?bo_table=dental&wr_id=41.

  23. Ha JH, Jo HJ, Kwak SW, Sigurdsson A, Kim HC. Effect of periodic changes in rotation speed on torsional stress and screw-in force by alternative rotation technique. J Endod. 2023;49:77–82.

    Article  PubMed  Google Scholar 

  24. Ajina M, Billis G, Chong BS. The effect of glide path preparation on root canal shaping procedures and outcomes. Eur Endod J. 2022;7:92–105.

    PubMed  PubMed Central  Google Scholar 

  25. Paleker F, van der Vyver PJ. Comparison of canal transportation and centering ability of K-files, ProGlider file, and G-files: a micro-computed tomography study of curved root canals. J Endod. 2016;42:1105–9.

    Article  PubMed  Google Scholar 

  26. Turkistani AK, Gomaa MM, Shafei LA, Alsofi L, Majeed A, AlShwaimi E. Shaping ability of HyFlex EDM and ProTaper Next rotary instruments in curved root canals: a micro-CT study. J Contemp Dent Pract. 2019;20:680–5.

    Article  PubMed  Google Scholar 

  27. Ruddle CJ. Single-file shaping technique: achieving a gold medal result. Dent Today. 2016;35:98 (100, 2–3).

    PubMed  Google Scholar 

  28. Calhoun G, Montgomery S. The effects of four instrumentation techniques on root canal shape. J Endod. 1988;14:273–7.

    Article  PubMed  Google Scholar 

  29. Lee S, Lee DK. What is the proper way to apply the multiple comparison test? Korean J Anesthesiol. 2018;71:353–60.

    Article  PubMed  PubMed Central  Google Scholar 

  30. Kim HY. Statistical notes for clinical researchers: Nonparametric statistical methods: 2. Nonparametric methods for comparing three or more groups and repeated measures. Restor Dent Endod. 2014;39:329–32.

    Article  PubMed  PubMed Central  Google Scholar 

  31. Ha JH, Park SS. Influence of glide path on the screw-in effect and torque of nickel-titanium rotary files in simulated resin root canals. Restor Dent Endod. 2012;37:215–9.

    Article  PubMed  PubMed Central  Google Scholar 

  32. Ha JH, Kwak SW, Kim SK, Kim HC. Screw-in forces during instrumentation by various file systems. Restor Dent Endod. 2016;41:304–9.

    Article  PubMed  PubMed Central  Google Scholar 

  33. Kyaw MS, Ebihara A, Kasuga Y, et al. Influence of rotational speed on torque/force generation and shaping ability during root canal instrumentation of extracted teeth with continuous rotation and optimum torque reverse motion. Int Endod J. 2021;54:1614–22.

    Article  PubMed  Google Scholar 

  34. Thu M, Ebihara A, Kyaw MS, et al. Influence of different kinematics on stationary and dynamic torsional behavior of JIZAI nickel-titanium rotary instruments: an in vitro study. J Dent Sci. 2023;18:1170–6.

    Article  PubMed  Google Scholar 

  35. Kwak SW, Lee CJ, Kim SK, Kim HC, Ha JH. Comparison of screw-in forces during movement of endodontic files with different geometries, alloys, and kinetics. Materials (Basel). 2019;12.

  36. Ha JH, Cheung GS, Versluis A, Lee CJ, Kwak SW, Kim HC. “Screw-in” tendency of rotary nickel-titanium files due to design geometry. Int Endod J. 2015;48:666–72.

    Article  PubMed  Google Scholar 

  37. Kwak SW, Ha JH, Shen Y, Haapasalo M, Kim HC. Effects of root canal curvature and mechanical properties of nickel-titanium files on torque generation. J Endod. 2021;47:1501–6.

    Article  PubMed  Google Scholar 

  38. Boutsioukis C, Arias-Moliz MT. Present status and future directions–irrigants and irrigation methods. Int Endod J. 2022;55(Suppl 3):588–612.

    Article  PubMed  PubMed Central  Google Scholar 

  39. Marinho AC, Martinho FC, Zaia AA, Ferraz CC, Gomes BP. Influence of the apical enlargement size on the endotoxin level reduction of dental root canals. J Appl Oral Sci. 2012;20:661–6.

    Article  PubMed  PubMed Central  Google Scholar 

  40. Omori S, Ebihara A, Hirano K, et al. Effect of rotational modes on torque/force generation and canal centering ability during rotary root canal instrumentation with differently heat-treated nickel-titanium instruments. Materials (Basel). 2022;15.

  41. Pirani C, Iacono F, Generali L, et al. HyFlex EDM: superficial features, metallurgical analysis and fatigue resistance of innovative electro discharge machined NiTi rotary instruments. Int Endod J. 2016;49:483–93.

    Article  PubMed  Google Scholar 

  42. Ha JH, Kwak SW, Sigurdsson A, Chang SW, Kim SK, Kim HC. Stress generation during pecking motion of rotary nickel-titanium instruments with different pecking depth. J Endod. 2017;43:1688–91.

    Article  PubMed  Google Scholar 

  43. Hsu YY, Kim S. The ProFile system. Dent Clin North Am. 2004;48:69–85.

    Article  PubMed  Google Scholar 

  44. Keskin C, Sivas Yilmaz Ö, Inan U, Özdemir Ö. Postoperative pain after glide path preparation using manual, reciprocating and continuous rotary instruments: a randomized clinical trial. Int Endod J. 2019;52:579–87.

    Article  PubMed  Google Scholar 

  45. Li UM, Lee BS, Shih CT, Lan WH, Lin CP. Cyclic fatigue of endodontic nickel titanium rotary instruments: static and dynamic tests. J Endod. 2002;28:448–51.

    Article  PubMed  Google Scholar 

  46. Araújo LBB, Calefi PHS, Alcalde MP, Gavini G, Vivan RR, Duarte MAH. Impact of kinematics on the efficiency and safety of an engine-driven file for glide path preparation in MB2 canals of maxillary molars. Clin Oral Investig. 2023;27:1153–8.

    Article  PubMed  Google Scholar 

  47. Silva E, Lima CO, Barbosa AFA, Lopes RT, Sassone LM, Versiani MA. The impact of TruNatomy and ProTaper Gold instruments on the preservation of the periradicular dentin and on the enlargement of the apical canal of mandibular molars. J Endod. 2022;48:650–8.

    Article  PubMed  Google Scholar 

  48. Lopes HP, Ferreira AA, Elias CN, Moreira EJ, de Oliveira JC, Siqueira JF Jr. Influence of rotational speed on the cyclic fatigue of rotary nickel-titanium endodontic instruments. J Endod. 2009;35:1013–6.

    Article  PubMed  Google Scholar 

  49. Carlesi T, Plotino G, Colangeli M, Pappalardo A, Malagnino VA. Influence of rotation speed of Mtwo files on root canal instrumentation time with different canal curvatures. Iran Endod J. 2021;16:38–42.

    PubMed  PubMed Central  Google Scholar 

  50. Ajuz NC, Armada L, Gonçalves LS, Debelian G, Siqueira JF Jr. Glide path preparation in S-shaped canals with rotary pathfinding nickel-titanium instruments. J Endod. 2013;39:534–7.

    Article  PubMed  Google Scholar 

  51. Berutti E, Cantatore G, Castellucci A, et al. Use of nickel-titanium rotary PathFile to create the glide path: comparison with manual preflaring in simulated root canals. J Endod. 2009;35:408–12.

    Article  PubMed  Google Scholar 

  52. Kwak SW, Ha JH, Cheung GS, Kim HC, Kim SK. Effect of the glide path establishment on the torque generation to the files during instrumentation: an in vitro measurement. J Endod. 2018;44:496–500.

    Article  PubMed  Google Scholar 

  53. Sobotkiewicz T, Huang X, Haapasalo M, et al. Effect of canal curvature location on the cyclic fatigue resistance of reciprocating files. Clin Oral Investig. 2021;25:169–77.

    Article  PubMed  Google Scholar 

  54. Oh S, Kum KY, Kim HJ, et al. Bending resistance and cyclic fatigue resistance of WaveOne Gold, Reciproc Blue, and HyFlex EDM instruments. J Dent Sci. 2020;15:472–8.

    Article  PubMed  Google Scholar 

Download references

Acknowledgements

This study was supported by a grant from the National Research Foundation of the Republic of Korea (NRF-2021R1G1A1006751).

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Soram Oh.

Ethics declarations

Conflict of interest

The authors have no competing interests to declare that are relevant to the content of this article.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Springer Nature or its licensor (e.g. a society or other partner) holds exclusive rights to this article under a publishing agreement with the author(s) or other rightsholder(s); author self-archiving of the accepted manuscript version of this article is solely governed by the terms of such publishing agreement and applicable law.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Woo, JY., Jang, JH., Chang, S.W. et al. Screw-in force, torque generation, and performance of glide-path files with three rotation kinetics. Odontology (2023). https://doi.org/10.1007/s10266-023-00879-w

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1007/s10266-023-00879-w

Keywords

Navigation