Objectives
Peripheral percutaneous vaECMO cannulation in acute situations, such as cardiopulmonary resuscitation (eCPR), can be highly demanding due to calcification and anatomical variations of the femoral vessels [1]. We introduce a method of accessing peripheral vessels utilizing a hydrophilic introducer sheath (Gore Dry Seal Flex).
Methods
We report on a 71-year-old multimorbid male patient after OHCA, who was transferred to our center for eCPR. Despite several attempts by experienced colleagues, it was not possible to pass the femoral artery using the standard 17-Fr arterial cannula. Therefore, an 18-Fr hydrophilic introducer sheath (Gore Dry Seal Flex) was implanted, in which a 17-Fr arterial ECMO-cannula was positioned.
Results
A 71-year-old man was transferred to the catheter laboratory under mechanical CPR, after OHCA and 5-min low-flow time. Coronary angiography showed no coronary artery stenosis. Arterial pH was 7.2 and lactate 10 mmol/L on arrival. The patient’s medical history revealed a pulmonary vein isolation procedure earlier that year, planned CRT-D implantation, and an aortic aneurysm of 5.7 cm. A severe arrhythmic event was assumed. We decided to establish a vaECMO in the heart catheter lab.
After placing an 8-Fr introducer sheath to the left the femoral artery, further dilatation of the vessel was impossible due to bending and snapping of the wire and dilatators. After accessing the contralateral femoral artery, the same difficulties occurred due to rigid and highly calcified vessels. As a bailout solution, an 18-Fr dry seal introducer sheath (Gore Dry Seal Flex) was easily introduced into the femoral artery and a 17-Fr arterial cannula was inserted right into the hydrophilic sheath (Fig. 1).
The Dry Seal Flex Introducer Sheath by Gore is a hydrophilic coated sheath that provides enhanced flexibility and, therefore, improved access to branched or highly calcified vessels in difficult anatomy. Sheath sizes are available from 12–26 Fr and lengths of 33–65 mm. Due to the challenging access of the peripheral arterial vessels, we decided to use a Dry Seal Introducer Sheath, to establish arterial ECMO flow.
After that, venous cannulation was successfully performed in a standard way, whereas the vein also showed extreme rigidity. A flow of 6 L/min could be achieved without any problems. During the stay in the intensive care unit, we found elevated measures of neuron-specific enolase (NSE), due to missing neurological reactions, the patient underwent cranial-CT, where edema with unfavorable prognosis was seen. Due to the severity of the cranial imaging, we decided to terminate any further therapy. In total, the ECMO-therapy was five full days with no technical problems concerning the elevated pressure of the arterial cannula through the introducer sheath.
Conclusion
In cases of difficult access to peripheral vessels, due to anatomical variants or calcification, using a hydrophilic introducer sheath may be lifesaving and time-saving.
Reference
Rupprecht L, Lunz D, Philipp A, et al. Pitfalls in percutaneous ECMO cannulation. Heart Lung Vessel. 2015;7:320–6.
Acknowledgements
Open Access funding provided by Projekt DEAL.
Author information
Authors and Affiliations
Corresponding authors
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Gaisendrees, C., Suhr, L., Ivanov, B. et al. Using hydrophilic introducer sheath for peripheral vaECMO cannulation in highly calcified vessels: the bailout solution. J Artif Organs 23, 405–406 (2020). https://doi.org/10.1007/s10047-020-01169-w
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10047-020-01169-w