Continuous cardiac output determination by thermodeprivation

  • Original Articles
  • Published:
Journal of Anesthesia Aims and scope Submit manuscript

Abstract

A modified thermodilution catheter (KATS catheter) capable of monitoring continuous cardiac output by thermodeprivation and preserving its conventional function was devised.

The KATS catheter has a thermistor incorporated closer to the tip of the catheter in addition to the usual thermistor used for conventional thermodilution. This additional thermistor is heated by a constant electric current but is capable of measuring its own temperature. The degree of heat deprivation is detected as the cooling of the thermistor, which is proportionally larger with larger blood velocity. Since blood flow is not the only source of heat deprivation, the actual formula was empirically derived by performing in vitro studies.

Cardiac output can be determined by assuming the cross sectional area of the pulmonary artery is stationary. Calibration can be derived from a cardiac output measurement by the usual thermodilution method with the same catheter. The KATS catheter readings correlated significantly with conventional thermodilution values and electromagnetic flowmeter readings in anesthetized dogs.

Continuous cardiac output measurement by the KATS catheter appears to be a promising technique.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  1. Swan HJC, Ganz W, Forrester J, Marcus H, Diamond G: Catheterization of the heart in man with the use of a flow-directed balloon tipped catheter. N Engl J Med 283:447–451, 1970

    CAS  PubMed  Google Scholar 

  2. Ganz W, Donoso R, Marcus HS, Forrester JS: A new technique for measurement of cardiac output by thermodilution in man. Am J Cardiol 27:392–396, 1971

    Article  CAS  PubMed  Google Scholar 

  3. Weisel RD, Berger RL, Hechtman HB: Measurement of cardiac output by thermodilution — current concepts. N Engl J Med 292:682–684, 1975

    Article  CAS  PubMed  Google Scholar 

  4. Weil MH: Measurement of cardiac output. Crit Care Med 5:117–119, 1977

    Article  CAS  PubMed  Google Scholar 

  5. Philip JH, Long MC, Quinn MD: Continuous thermal measurement of cardiac output. IEEE Transact Biomed Engineering BME-31 5:393–400, 1984

    Article  Google Scholar 

  6. Delaunois AL, Rovati LA: A new method for continuous measurement of cardiac out-put. Arch Int Pharmacodyn CXVI1-2:228–236, 1958

    Google Scholar 

  7. Barankay T, Jansco T, Nagay S, Petri G: Cardiac output estimation by a thermodilution method involving intravascular heating and thermistor recording. Acta Physiol Acad Sci Hungarie 38:167–173, 1970

    CAS  Google Scholar 

  8. Fegler G: Measurement of cardiac output in anaesthetized animals by a thermodilution method. Quart J Exp Physiol 39:153–164, 1954

    CAS  PubMed  Google Scholar 

  9. Khalii HH, Richardson Q, Guyton AC: Measurement of cardiac output by thermaldilution and direct Fick methods in dogs. J Appl Physiol 21:1131–1135, 1966

    Google Scholar 

  10. Gibbs FA: A thermoelectric blood fluid velocity recorder in the form of a needle. Proc Soc Exp BioI Med 31:141–146, 1933

    Google Scholar 

  11. Ferrari M, Giannini I, Carta F, Argiolas L, Capri A: Quantitative measurement of tissue blood fluid velocity by fast pulse heated thermistors. Physiol Chern Phys 14:553–560, 1982

    CAS  Google Scholar 

  12. Sykes MK, Vickers MD, Hull CJ: Principles of clinical measurement. Oxford, Blackwell Scientific Publications 1982, pp. 202–219

    Google Scholar 

  13. Williamson JR, Shanahan MO, Hochmuth RM: The influence of temperature on red cell deformability. Blood 46:611–624, 1975

    CAS  PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

About this article

Cite this article

Miyasaka, K., Volgyesi, G., Katayama, M. et al. Continuous cardiac output determination by thermodeprivation. J Anesth 5, 205–212 (1991). https://doi.org/10.1007/s0054010050205

Download citation

  • Received:

  • Accepted:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s0054010050205

Key words

Navigation