Evaluation of surgical performance with standard rigid and flexible-tip laparoscopes

  • Original article
  • Published:
Surgical Endoscopy And Other Interventional Techniques Aims and scope Submit manuscript

Abstract

Background

Flexible-tip laparoscopes have recently been introduced into clinical practice, with the goal of improving surgeon performance during complex laparoscopic procedures. We used objective and subjective performance parameters to compare standard rigid 0° and 30° lens laparoscopes two flexible-tip laparoscopes in an in vitro model.

Methods

Twenty-nine subjects with varied levels of surgical experience performed complex laparoscopic tasks in three different models simulating (a) prostate dissection from the rectum, (b) cystic duct clipping, and (c) distal posterior rectum dissection. Each task was performed using two Storz rigid laparoscopes (0° and 30°) and two flexible-tip laparoscopes, the Olympus LTF-V3 and the Fujinon EL2-TF310. The sequence of application of the two flexible-tip laparoscopes was randomized. In each case, an experienced laparoscopic camera driver controlled the field of vision. Time to complete each task, operative precision, and subjective surgeon rating scores were compared. Statistical analysis was performed with analysis of variance (ANOVA) and a two-sided fisher’s exact test.

Results

In all three models, the flexible laparoscopes offered no advantage in terms of procedure time, surgical precision, or subjective surgeon rating score when compared with the 30° lens rigid laparoscope. The 30° rigid lens laparoscope and the two flexible-tip laparoscopes were superior to the 0° lens rigid laparoscope for all parameters evaluated, with the exception of subjective rating in the cystic duct model and procedure time in the colorectal model.

Conclusion

In this in vitro experimental model, the flexible-tip laparoscopes found to have no advantage over the standard rigid 30° lens laparoscope. These models were validated, as the 0° lens rigid laparoscope was surpassed by the 30° lens rigid laparoscope and the flexible-tip laparoscopes. Both flexible-tip laparoscopes produced similar results and excellent image quality, but some experience is required before their smooth application can be achieved.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Institutional subscriptions

Figure 1
Figure 2
Figure 3

Similar content being viewed by others

References

  1. Berber E, Siperstein AE (2001) Understanding and optimizing laparoscopic videosystems. SurgEndosc 15: 781–787

    Google Scholar 

  2. Berci G, Forde KA (2000) History of endoscopy: what lessons have we learned from the past? Surg Endosc 14: 5–15

    Article  PubMed  Google Scholar 

  3. Boppart SA, Deutsch TF, Rattner DW (1999) Optical imaging technology in minimally invasive surgery, current status and future directions. Surg Endosc 13: 718–722

    Article  PubMed  Google Scholar 

  4. Cuschieri A (1999) Technology for minimal access surgery. BMJ 319: 1304–1310

    PubMed  Google Scholar 

  5. Kourambas J, Preminger GM (2001) Advances in camera, video, and imaging technologies in laparoscopy. Urol Clin North Am 28: 5–14

    PubMed  Google Scholar 

Download references

Acknowledgments

Fujinon, Olympus, and Storz each supplied equipment for this study.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to J. Landman.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Perrone, J., Ames, C., Yan, Y. et al. Evaluation of surgical performance with standard rigid and flexible-tip laparoscopes. Surg Endosc 19, 1325–1328 (2005). https://doi.org/10.1007/s00464-004-8282-6

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00464-004-8282-6

Keywords

Navigation