The influence of force feedback and visual feedback in grasping tissue laparoscopicaUy

  • Published:
Surgical Endoscopy Aims and scope Submit manuscript

Abstract

Background: Due to the limited force feedback provided by laparoscopic instruments, surgeons may have difficulty in applying the appropriate force on the tissue. The aim of this study was to determine the influence of force feedback and visual feedback on the exerted pinch force.

Methods: A grasper with a force sensor in the jaws was developed. Subjects with and without laparoscopic experience grasped and pulled pig bowel with a force of 5 N. The applied pinch force was measured during tasks of 1-s and 1-min duration. Visual feedback was provided in half the measurements. Force feedback was adjusted by changing the mechanical efficiency of the forceps from 30% to 90%.

Results: The mean pinch force applied was 6.8 N (± 0.5), whereas the force to prevent slippage was 3.0 N ( ± 0.4). Improving the mechanical efficiency had no effect on the pinch force for the 1-s measurements. The amount of excessive pinch force when holding tissue for 1 min was lower at 30% mechanical efficiency compared with 90% (105% vs 131 %, p = 0.04). The tissue slipped more often when the subject had no visual feedback (2% vs 8%, p = 0.02).

Conclusion: Force feedback and visual feedbacK playa more limited role than expected in the task of grasping tissue with laparoscopic forceps.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Institutional subscriptions

Similar content being viewed by others

References

  1. Balázs M, Feussner H, Hirzinger G, Omote K, Ungeheuer A (1998) A new tool for minor-access surgery: replacing mechanical joints in laparoscopic forceps with elastic beams for improved pressure control and sensitivity. IEEE Eng Med Biol 17: 45–48

    Article  Google Scholar 

  2. Bholat OS, Haluck RS, Murray WB, Gorman PJ, Krummel TM (1999) Tactile feedback is present during minimally invasive surgery. J Am Coli Surg 189: 349–355

    Article  CAS  Google Scholar 

  3. den Boer KT, Herder JL, Sjoerdsma W, Meijer DW, Gouma DJ, Stassen HG (1999) Sensitivity of laparoscopic dissectors: what can you feel? Surg Endosc 13: 869–873

    Article  Google Scholar 

  4. de Visser H (2003) Grasping safely, instruments for bowel manipulation investigated [Dissertation]. Delft University of Technology, Delft, The Netherlands

    Google Scholar 

  5. de Visser H, Heijnsdijk EAM, Herder JL, Pistecky PV (2002) Forces and displacements in colon surgery. Surg Endosc 16: 142–1430

    Article  Google Scholar 

  6. Heijnsdijk EAM, Dankelman J, Gouma DJ (2002) Effectiveness of grasping and duration of clamping using laparoscopic graspers. Surg Endosc 16: 1329–1331

    Article  CAS  Google Scholar 

  7. Herder JL, Horward MJ, Sjoerdsma W (1997) A laparoscopic grasper with force perception. Min Invas Ther Allied Technol 6: 279–286

    Article  Google Scholar 

  8. Howe RD, Peine WJ, Kontarinis DA, Son JS (1995) Remote palpation technology. IEEE Eng Med Biol 14: 318–323

    Article  Google Scholar 

  9. Joice P, Hanna GB, Cuschieri A (1998) Errors enacted during endoscopic surgery—a human reliability analysis. Appl Erg 29: 409–414

    Article  CAS  Google Scholar 

  10. Rosen J, Hannaford B, MacFarlane MP, Sinanan MN (1999) Force controlled and teleoperated endoscopic grasper for minimally invasive surgery—experimental performance evaluation. IEEE Trans Biomed Eng 46: 1212–1221

    Article  CAS  Google Scholar 

  11. Ruurda JP, van Vroonhoven TJMV, Breeders TAMJ (2002) Robot-assisted surgical systems: a new era in laparoscopic surgery. Ann R Coli Surg Engl 84: 223–226

    Article  CAS  Google Scholar 

  12. Sjoerdsma W, Herder JL, Horward MJ, Jansen A, Bannenberg JJG, Grimbergen CA (1997) Force transmission of laparoscopic grasping instruments. Min Invas Ther Allied Technol 6: 274–278

    Article  Google Scholar 

  13. van der Piji AJ, Herder JL (2001) Development of 5-mm trocar laparoscopic forceps with mechanical force feedback. Proceedings of the Design Engineering Technical Conferences 2001, American Society of Mechanical Engineers, 3–12 September 2001, DETC 01, ASME Conference, Pittsburgh PA, USA

    Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

Reprints and permissions

About this article

Cite this article

Heijnsdijk, E.A.M., Pasdeloup, A., van der Pijl, A.J. et al. The influence of force feedback and visual feedback in grasping tissue laparoscopicaUy. Surg Endosc 18, 980–985 (2004). https://doi.org/10.1007/s00464-003-9244-0

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00464-003-9244-0

Key words

Navigation