Multicenter study on finger systolic blood pressure test for diagnosis of vibration-induced white finger

  • Original Article
  • Published:
International Archives of Occupational and Environmental Health Aims and scope Submit manuscript

Abstract

Objective

A multicenter study (six Rosai hospitals around Japan) was performed to investigate the diagnostic value of changes in finger systolic blood pressure (FSBP) after segmental local cooling for vibration-induced white finger (VWF).

Methods

Subjects were 154 men without exposure to vibration and 135 men with occupational vibration exposure. They were classified into four groups: Group A, 154 unexposed control cases; Group B, 21 exposed cases without VWF; Group C, 31 cases with a history of VWF but without any signs of VWF within the last year; and Group D, 83 cases with active VWF within the last year. FSBP% measurements were taken at room temperatures of 23 ± 1 and 21 ± 1°C, using a strain-gauge Digimatic 2000 plethysmograph (Medimatic).

Results

At a room temperature of 23 ± 1°C, there was a significant difference between Groups A and D, and B and D. At a room temperature of 21 ± 1°C, there was a significant difference between Groups A and C, A and D, and B and D. The values in Group D were the lowest at both room temperatures. Assuming a cut-off value of 75% at 23 ± 1°C, the sensitivity and specificity were 65.2 and 87.5%, respectively. Assuming the same cut-off value at 21 ± 1°C, the sensitivity and specificity were 73.9 and 82.5%, respectively. These values were not too high. Most of the subjects with WVF in this study were retired and had not used vibratory tools for many years. The situation of the subjects may affect the results of the FSBP test. Our data did not confirm a difference in diagnostic accuracy between room temperatures of 23 ± 1 and 21 ± 1°C.

Conclusions

Our study showed that the sensitivity and specificity of the FSBP test with a cut-off value of 75% at 23 ± 1°C, were 65.2 and 87.5%, respectively, and at 21 ± 1°C, they were 73.9 and 82.5%, respectively.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Institutional subscriptions

Fig. 1
Fig. 2

Similar content being viewed by others

References

  • Bovenzi M (1988) Vibration white finger, digital systolic blood pressure, and some biochemical findings on workers operating vibrating tools in the engine manufacturing industry. Am J Ind Health 14:575–584

    Article  CAS  Google Scholar 

  • Bovenzi M (1993) Digital arterial responsiveness to cold in healthy men, vibration white finger and primary Raynaud’s phenomenon. Scand J Work Environ Health 19:271–276

    PubMed  CAS  Google Scholar 

  • Bovenzi M (1998) A prospective study of the cold response of digital vessels in forestry workers exposed to saw workers. Int Arch Occup Environ Health 71:493–498

    Article  PubMed  CAS  Google Scholar 

  • Bovenzi M (2002) Finger systolic blood pressure indices for the diagnosis of vibration-induced white finger. Int Arch Occup Environ Health 75:20–28

    PubMed  Google Scholar 

  • Chang C (1976) Cold water immersion test in patients with vibration disease. Jpn J Ind Health 18:453–463

    CAS  Google Scholar 

  • Ekenvall L, Lindblad LE (1982) Digital blood pressure after local cooling as a diagnostic tool in traumatic vasospastic disease. Br J Ind Med 39:388–391

    PubMed  CAS  Google Scholar 

  • Ekenvall L, Lindblad LE (1986) Vibration white finger and digital systolic pressure during cooling. Br J Ind Med 43:280–283

    PubMed  CAS  Google Scholar 

  • International Organization for Standardization (2005) Mechanical vibration and shock—Cold provocation tests for the assessment of peripheral vascular function—part 2: measurement and evaluation of finger systolic blood pressure. ISO 14835-2, Geneva

  • Kurozawa Y, Nasu Y (2005) Factors influencing finger systolic blood pressure test for diagnosis of vibration-induced white finger. Environ Health Prev Med 10:366–370

    Article  Google Scholar 

  • Kurozawa Y, Nasu Y, Nose T (1991) Diagnostic value of finger systolic blood pressure in the assessment of vasospastic reactions in the fingerskin of vibration-exposed subjects after finger and body cooling. Scand J Work Environ Health 17:184–189

    PubMed  CAS  Google Scholar 

  • Kurozawa Y, Nasu Y, Oshiro H (1992) Finger systolic blood pressure measurement after finger cooling. Using the Laser-Doppler method for assessing vibration-induced white finger. J Occup Med 34:683–686

    PubMed  CAS  Google Scholar 

  • Hanley J, Mcneil B (1983) A method of comparing the areas under receiver operating characteristic curves derived from the same cases. Radiology 148:839–843

    PubMed  CAS  Google Scholar 

  • Nasu Y, Kurozawa Y (1995) Comparison of the diagnostic values of three provocation tests for vibration white finger. Arbete Och Hälsa 5:157–160

    Google Scholar 

  • Olsen N, Nielsen SL (1979) Diagnosis of Raynaud’s phenomenon in quarrymen’s traumatic vasospastic disease. Scand J Work Environ Health 5:249–256

    Google Scholar 

  • Olsen N, Hagberg M (1995) Report of working group 1—clinical and laboratory diagnostics of vascular symptom induced by hand–arm vibration. Arbete Och Hälsa 5:181–186

    Google Scholar 

  • Olsen N, Nielsen SL, Voss P (1982) Cold response of digital arteries in chain saw operators. Br J Ind Med 38:82–88

    Google Scholar 

  • Pyykkö I, Färkkilä M, Korhonen O, Starck J, Jäntti V (1986) Cold provocation tests in the evaluation of vibration-induced white finger. Scand J Work Environ Health 12:254–258

    PubMed  Google Scholar 

Download references

Acknowledgments

This research was funded as part of the Japanese Labor, Health and Welfare Organization’s research and development in the 13 fields of occupational injuries and resulting illnesses. We would also like to thank Professor Hideki Hashimoto of Tokyo University for his advice on analytical problems.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Yoshiro Nasu.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Nasu, Y., Kurozawa, Y., Fujiwara, Y. et al. Multicenter study on finger systolic blood pressure test for diagnosis of vibration-induced white finger. Int Arch Occup Environ Health 81, 639–644 (2008). https://doi.org/10.1007/s00420-007-0273-1

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00420-007-0273-1

Keywords

Navigation