Abstract
Introduction
Real-time visualization of intraoperative electrocochleography (ECochG) potentials via a digital microscope during cochlear implantation can provide direct feedback during electrode insertion. The aim of this prospective, randomized study of 50 patients was to obtain long-term data with a focus on residual hearing preservation and speech understanding.
Material and methods
Cochlear implantations were performed in 50 patients (26 female, 24 male) with residual hearing using a digital microscope. Patients were randomized into two groups. Intraoperative ECochG potentials were either displayed directly in the surgeon’s field of view (picture-in-picture display, PiP) or not directly in the field of view (without picture-in-picture display, without PiP). Residual hearing preservation and speech comprehension were recorded within a 1-year follow-up period, compared between groups (PiP versus without PiP) and to a control group of 26 patients implanted without ECochG.
Results
Mean insertion time was significantly longer in the picture-in-picture group (p = 0.025). Residual hearing preservation after 6 weeks at 250 Hz was significantly better in the picture-in-picture group (p = 0.017). After one year, 76% of patients showed residual hearing in the picture-in-picture group (62% without picture-in-picture technique, p = n.s.). Use of the picture-in-picture technique resulted in better long-term pure tone residual hearing preservation at 250, 500, and 1000 Hz. Speech intelligibility improved by 46% in the picture-in-picture group (38% without picture-in-picture).
Discussion
This study is the first to describe long-term results in a large cohort of cochlear implant patients in whom digital visualization of intraoperative ECochG was used. Our results show that visualization of intraoperative ECochG has a positive effect on residual hearing preservation.
Similar content being viewed by others
Introduction
Cochlear implants (CI) are successfully implanted in patients with deafness or severe hearing loss and enable patients to receive sound. The surgically inserted electrode directly stimulates the human spiral ganglion cells [1]. Patients benefiting from cochlear implants are deafened infants [2, 3], elderly people [4], patients with Single-Sided Deafness [5] and, in the last decade, also patients with residual acoustic hearing. In the last years, hearing preservation during cochlear implantation has received more and more scientific and clinical attention. It could be shown that the combination of electric and acoustic stimulation of the cochlea by preserving residual hearing can have positive effects on speech understanding [6]. Additionally, perception of periodicity, sound localization [7, 8] and music perception [9] are improved. Potential mechanisms of hearing loss during cochlear implant surgery are suggested to be physical trauma in scala tympani and to the osseous spiral lamina, potential dislocations of the electrode to scala vestibuli and postoperative inflammation [10]. Therefore, methods to avoid intraoperative trauma are of high clinical interest. Advances in electrode design such as changes in stiffness, electrode tip, size and diameter, surface morphology or deployment mechanisms have contributed to a reduction of intraoperative damage to inner ear structures. Moreover, multiple intraoperative tests have been implemented. Impedance measurements, electrically evoked stapedius reflex threshold (eSRT) and electrically evoked compound action potentials (ECAP) or spread of neural excitation (SOE) provide useful and important information on correct position of electrode array and hearing nerve answers. These tests are broadly performed [11, 12]. However, a positive effect on hearing preservation or speech understanding has not been demonstrated [13] as they all measure the steady state after insertion of electrode and therefore don’t have a direct impact on the insertion itself.
A method enabling intraoperative monitoring of hearing preservation is electrocochleography (ECochG) [14]. In ECochG, electric potentials from inner and outer hair cells and auditory nerve caused by acoustic stimulation in the outer ear canal are detected [15, 16]. Measurement of those potentials can be performed in various ways, e.g. with an externally located electrode in the outer ear canal or a needle placed on the promontory [17, 18]. It is therefore a diagnostic instrument in preoperative screening before cochlear implantation. Used intraoperatively, it can give the surgeon direct feedback from the electrodes on the implant array monitoring the intracochlear activity. In case of deterioration, changes in insertion speed, insertion angle or applied pressure could be taken [19,20,21]. Moreover, retraction or stop of insertion is possible.
Digital microscopes such as the ARRISCOPE® (Munich Surgical Imaging, Munich, Germany) can provide direct visualization of intraoperative measurements such as EcochG potentials during electrode insertion by picture-in-picture technique (PiP). The microscope captures the surgical field with a high-resolution imaging sensor, processes the image in real-time and displays the image in the binocular based on two small OLED displays. This digital processing chain enables the integration of additional data into the digital binocular. Feasibility of intraoperative ECochG during cochlear implantation using the digital microscope has been demonstrated [22].
The aim of this prospective randomized study is to demonstrate the postoperative audiometric results in the first 50 patients operated with the digital microscope using intraoperative ECochG. We were specifically aiming at comparing audiometric results and insertion time between patients who were operated on with picture-in-picture technique on the one side, to patients who were implanted without the picture-in-picture technique on the other side. In the latter, the surgeon got the information about the intraoperative ECochG-potentials via verbal feedback from the audiologist.
Material and methods
Preliminary setup in the lab
Prior to the intraoperative trial, a collaborative laboratory test was conducted to define and test the image exchange interface. In addition, the size of the picture-in-picture insert was adopted. As graphical input, the Cochlear™ Research Platform 1.1 running on a standard laptop (ThinkPad, LenovoTM) was used. The laptop was connected to a Cochlear™ Nucleus® 6 sound processor. A Slim Modiolar Practice Electrode was inserted by an experienced surgical support specialist (Cochlear) into a pre-drilled PHACON Temporal Bone Model. The Hybrid™ component acoustically stimulates the inner ear and the electrode array records the inner ear potential. The image signal was taken from the HDMI output of the laptop and fed via a converter (Blackmagic—Mini Converter UpDownCross HD) as HD-SDI signal (1080p 30/60 Hz) into a fully digital surgical microscope. The image signal of the Cochlear™ Research Platform 1.1 was displayed on all displays, including the binocular of the surgical microscope in full-screen mode or in thumbnail view (Picture-in-Picture mode). The user was able to switch between these two display modes using the handgrip keys. Depending on the selected display mode, the live image of the microscope was simultaneously shown on the same displays, either in full-screen or picture-in-picture mode. According to the testing surgeons it was easily possible to briefly switch attention between the insertion and the display of the Cochlear Research Platform.
Patients
Fifty patients with residual hearing were included (26 female, 24 male). Mean age was 48.2 years (SD 21.4). Randomization in two groups was performed by a random generator (PiP: mean age 48.1 years, 15 female, 15 male; without PiP: mean age 44,5 years, 11 female, 9 male). Residual hearing preoperatively was defined as a hearing threshold of at least 80 dB HL in at least two tested frequencies (air conduction). Residual hearing postoperatively was defined as a hearing threshold of at least 115 dB HL in at least two tested frequencies (air conduction). Audiometric thresholds were measured in regular pre-diagnostic procedures before cochlear implantation. The mean pre-operative PTAlow was 68.3 dB HL (SD 23.7). Additionally, 26 patients (mean age: 48.1 years, 14 female, 12 male) were included for retrospective analysis of the data. All these patients were implanted with a CI622 electrode between 2019 and 2020 but without intraoperative ECochG measurements. The mean pre-operative PTAlow was 68.4 dB HL (SD 20.2).
Cochlear implants
Commercially available Cochlear™ Nucleus® Profile™ Plus with Slim Straight Electrode (CI622) (Cochlear®, Sydney, Australia) were used. Intraoperative ECochG was performed in all patients during electrode insertion. The patients were randomized into two groups: in group I potentials of intraoperative ECochG were visualized in the surgeon’s field of view. In group II potentials of intraoperative ECochG were not visualized in the surgeon’s field of view. Here, the audiologist gave acoustic feedback to the surgeon during insertion. In case of drops in the ECochG curve, the surgeon stopped the insertion and waited for the signal to recover.
Intraoperative setting (Fig. 1 )
The acoustic component of the sound processor was connected to a disposal sound tube (for 3 M E-A-RTONE™ insert earphones, 3 M AEARO) and a disposal ear-tip (3 M E-A-RLINK™, 3 M AEARO). The ear tip was positioned in the ear canal before surgery started. ECochG potentials were acoustically generated with a tone burst of 250 Hz, 112 dB HL and a duration of 7 ms. The screen of the Cochlear™ Research Platform 1.1 was transferred to the digital microscope ARRISCOPE® (Munich Surgical Imaging, Munich, Germany) and appeared in the surgeon´s field of view. During the insertion of the electrode array, the amplitude of cochlear microphonics was continuously measured with 3 updates per second. For recording, the most apical electrode (E22) and the case electrode were used. Additionally, the amplitude growth function (AGF) resembling the intraoperative residual hearing was determined at the end of insertion at 250, 500, 750 and 1000 Hz.
Audiometric tests
Preoperatively, all patients underwent standard tone and speech audiometry as well as objective hearing tests like auditory brainstem response (ABR) and cochlear microphonics (CM) in routine cochlear implant screening by professional audiometrists using Auritec audiometer (Auritec GmbH, Hamburg, Germany) and BERA Eclipse EP 25/ECochG/ASSR (Interacoustics A/S, Assens, Denmark). Additionally, air and bone conduction were measured one day before and one day after surgery. Postoperative monitoring of residual hearing and speech perception was performed at 6 weeks, 3 months, 6 months and 12 months postoperatively. All tests were performed according to the declaration of Helsinki and were approved by the local ethics committee. Informed written consent was obtained from all patients.
Additional intraoperative monitoring during cochlear implant surgery
The electrode impedance mirrors the electrical resistance and gives additional information about a successful electrode array insertion and the technical integrity of the device. Electrical Stapedius Reflex Thresholds (eSRTs) are measured by visually observing the contraction of the stapedius muscle in the implant ear. Moreover, the function of CI and nerve was monitored by Electric Compound Action Potentials (ECAPs). Spread of Excitation (SOE) [11] and Trans-Impedance Matrix (TIM) measurement [23] identify and characterize the normal electrode position. Intraoperatively, the correct electrode position was examined with 3D C-arm X-ray-unit.
Data analysis
Graphs and statistical analysis were performed using GraphPad Prism 6 (GraphPad Software Inc., LA Jolla, USA), SPSS (Ver. 27, IBM, Ehningen, Deutschland), MATLAB (Ver. R2022a, MathWorks Inc., USA) and Microsoft Excel (Redmond, USA). In the case of normal distribution, t-tests were obtained. If normal distribution could not be assumed, a Mann–Whitney-Wilcoxon test was applied.
Results
Insertion time
Insertion time was defined as the time between the first intracochlear electrode contact and the end of insertion. Mean insertion time in the picture-in-picture group was significantly longer than in the without picture-in-picture group (101 s versus 77.6 s, p = 0.025, z = 2.23). Accordingly, the insertion speed was 0.1 mm/s slower with the picture-in-picture method.
Hearing preservation
Patients
Residual hearing was measured at 6 weeks, 3 months, 6 months and 12 months postoperatively. After 6 weeks postoperatively, hearing preservation was significantly better in the picture-in-picture group (96% versus 72% in the without picture-in-picture group, p = 0.017). Hearing preservation during the follow-up period is shown in Fig. 2. Residual hearing preservation is not significantly better in patients implanted with intraoperative ECochG compared to the control group of patients implanted regularly without ECochG. However, after 3 months postoperatively there is a visible trend that the hearing thresholds are worse for patients without ECochG at 1000 Hz (Fig. 4).
Amplitude growth function of ECochG potentials
No significant differences between groups were detected with slightly better AGF thresholds in the PiP group. AGF-potentials at different frequencies over the follow-up-period are shown in Fig. 3.
Comparison of pure tone threshold and amplitude growth function
For the evaluation and validation of AGF thresholds, a comparative analysis between the detected pure tone thresholds and the AGF results was calculated (Fig. 4). In addition, mean pure tone thresholds of 26 patients operated without intraoperative ECochG were used as a control group. It was shown that at 250 Hz, AGF results significantly differ from the pure tone thresholds of subjective audiometry at all times of evaluation (p < 0.01, p = 0.04, respectively). For 500 and 1000 Hz, AGF results only differed significantly from the pure tone thresholds after 6 months postoperatively (p = 0.01, p = 0.05 respectively). Without the use of intraoperative ECochG, residual hearing after 3 months is worse than with the use of intraoperative ECochG measurements, a statistical significance could not be detected (Fig. 4).
Speech understanding
For the evaluation of speech understanding Freiburger monosyllables were used (Fig. 5). Speech understanding at 65 dB improved by 38% in the without-picture-in-picture group and by 46% in the picture-in-picture group after one year compared to the best-aided condition preoperatively (p < 0.01, p = 0.017 respectively). There was so significant difference between both groups (p = 0.65, z = -0.48).
Discussion
This is the first study analyzing digital real-time visualization of intraoperative ECochG potentials during cochlear implantation in a large patient cohort. The goal of this randomized prospective trial was to evaluate patients´ hearing outcomes focusing on the preservation of residual hearing. We were specifically aiming at comparing postoperative audiometric results between two groups. In group I intraoperative ECochG signals were visualized in the surgeon´s field of view (picture-in-picture). In group II intraoperative ECochG signals were only acoustically reported by the audiologist (without picture-in-picture). In this study, residual hearing preservation in low-frequency range was achieved in 76% of the patients. Thereby, direct visualization of EcochG led to an improved residual hearing preservation. Taken together we think that digital real-time visualization of intraoperative EcochG during cochlea-implantation is a promising tool to improve patients´ hearing outcome.
Different methods to test for the correct position of the electrode immediately after or during surgery have been established. Electrically evoked compound action potentials such as e.g. NRT® (neural response telemetry) are commonly performed but do not provide real-time information during insertion. They are mostly used as a post-insertion control instrument and for objective fitting. Moreover, stimulation is not performed via tone burst but by electric activation of the implant [13] and therefore is not helpful to preserve residual acoustical hearing. In contrast, ECochG can provide real-time feedback to the surgeon during electrode insertion. However, until now, there are two major problems with ECochG performed during cochlear implant surgery.
Firstly, the majority of the present studies assessing intraoperative ECochG measurement are not based on real-time feedback during insertion but rather provide post-insertional feedback although potentials are measured immediately after insertion [20, 24]. With this approach, relevant information during the insertion period may be lost. In a model experiment performed by a senior cochlear implant surgeon, the mean insertion speed was 0.48 mm/sec with a mean insertion time between 44.4 and 48.6 s [25]. Another retrospective analysis of 54 uneventful insertions revealed a mean insertion speed of 96.5 mm/min [26]. Although the insertion period seems relatively short, manual insertion by the surgeon at a constant speed without trembling is limited. In our opinion, the insertion period needs to be considered as the key time window during which changes in insertion speed or -angle can be taken if drops in ECochG potentials occur aiming at avoiding intracochlear trauma with the need for immediate feedback. Secondly, even if direct feedback is given, this needs to be done – up to now—verbally by the audiologist or acoustically by the machine performing the measurement. Considering external factors such as human reaction rate, high noise levels and the presence of several team members in the operating room, the surgeon operating via microscope has multiple tasks to fulfill simultaneously in the moment of electrode insertion. Thus, relevant information come along with a considerable time delay which in turn prohibits adequate reaction, e.g. short pause of electrode insertion or gentle pull back.
In this analysis, it was demonstrated that digital real-time visualization of intraoperative ECochG with a digital microscope facilitates this feedback and can add to hearing preservation. An important finding was, that in the picture-in-picture group the insertion speed was lower than in the without picture-in-picture group. This can be explained by continuous monitoring of the ECochG-curve during electrode insertion leading to a slower insertion. Residual hearing preservation in low-frequency range was possible in 76% of the patients (picture-in-picture group). These data are roughly in line with a recent investigation assessing residual hearing after cochlear implantation with a new lateral wall electrode in a cohort of 20 patients where intraoperative ECochG was used [27]. Another study of 17 patients focusing on residual hearing preservation obtained comparable results [28]. However, there are data that indicate a lower rate of residual hearing preservation than in our investigation [29]. Of note, in that study, intraoperative ECochG was not performed. We think that the use of intraoperative ECochG and especially its direct visualization will emerge as an essential part of modern cochlear implant surgery. This is underlined by our comparative analysis, which yielded better hearing preservation in patients operated with ECochG compared to no use of ECochG at all.
Regarding the results of the AGF of ECochG-potentials in the follow-up period, it was shown that AGF potentials were lower in the picture-in-picture group over all frequencies. This also speaks for real-time visualization of ECochG-potentials during insertion. However, so far, there are still lots of unanswered questions concerning AGF-potentials during cochlear implantation especially due to the heterogeneity of ECochG monitoring technique [30]. Comparing AGF thresholds with pure tone audiometry, AGF potentials at 250 Hz were significantly lower over the whole follow-up period than thresholds obtained by pure tone audiometry. Studies investigating on AGF potentials showed that they indicate neural and hair cell function and that postoperative ECochG thresholds positively correlate with behavioral thresholds after surgery [31]. However, in the respected study, this could only be confirmed for 500 and 1000 Hz but not for 250 Hz. Although a specific explanation for the phenomenon cannot be given, these new data could add to a better understanding of intraoperative ECochG potentials. A better speech understanding in the picture-in-picture group further underlined the positive effect of direct visualization of intraoperative ECochG enabling the surgeon to directly change insertion speed or angle if drops in the curve occur.
Taken together, this study demonstrates valid data that indicate a positive effect on residual hearing preservation in cochlear-implant surgery when digital live imaging of intraoperative ECochG is performed. In the future, ongoing close cooperation of scientists, surgeons and technicians will contribute to technical progress and digitalization in otologic surgery. Robot-supported cochlear implant surgery providing a slow insertion process at constant speed is an emerging field that might benefit from digital live visualization of intraoperative ECochG in the future [32].
Data availability
Not applicable.
References
Li H, Schart-Moren N, Rohani SA et al (2020) Synchrotron radiation-based reconstruction of the human spiral ganglion: implications for cochlear implantation. Ear Hear 41(1):173–181
Teschendorf M, Arweiler-Harbeck D, Bagus H (2010) Speech development after cochlear implantation in children with bilingual parents. Cochlear Implants Int 11(Suppl 1):386–389
Holman MA, Carlson ML, Driscoll CL et al (2013) Cochlear implantation in children 12 months of age and younger. Otol Neurotol 34(2):251–258
Buchsenschutz K, Arnolds J, Bagus H et al (2015) Surgical risk profile and audiological outcome in the elderly after cochlea-implantation. Laryngorhinootologie 94(10):670–675
Zeitler DM, Sladen DP, DeJong MD et al (2019) Cochlear implantation for single-sided deafness in children and adolescents. Int J Pediatr Otorhinolaryngol 118:128–133
Gifford RH, Dorman MF, Skarzynski H et al (2013) Cochlear implantation with hearing preservation yields significant benefit for speech recognition in complex listening environments. Ear Hear 34(4):413–425
Dunn CC, Perreau A, Gantz B, Tyler RS (2010) Benefits of localization and speech perception with multiple noise sources in listeners with a short-electrode cochlear implant. J Am Acad Audiol 21(1):44–51
Dorman MF, Gifford RH (2010) Combining acoustic and electric stimulation in the service of speech recognition. Int J Audiol 49(12):912–919
Yuksel M, Meredith MA, Rubinstein JT (2019) Effects of low frequency residual hearing on music perception and psychoacoustic abilities in pediatric cochlear implant recipients. Front Neurosci 13:924
Eshraghi AA, Gupta C, Van De Water TR et al (2013) Molecular mechanisms involved in cochlear implantation trauma and the protection of hearing and auditory sensory cells by inhibition of c-Jun-N-terminal kinase signaling. Laryngoscope 123(Suppl 1):S1-14
Grolman W, Maat A, Verdam F et al (2009) Spread of excitation measurements for the detection of electrode array foldovers: a prospective study comparing 3-dimensional rotational x-ray and intraoperative spread of excitation measurements. Otol Neurotol 30(1):27–33
Cohen LT, Richardson LM, Saunders E, Cowan RS (2003) Spatial spread of neural excitation in cochlear implant recipients: comparison of improved ECAP method and psychophysical forward masking. Hear Res 179(1–2):72–87
Miller CA, Brown CJ, Abbas PJ, Chi SL (2008) The clinical application of potentials evoked from the peripheral auditory system. Hear Res 242(1–2):184–197
O’Leary S, Briggs R, Gerard JM et al (2020) Intraoperative observational real-time electrocochleography as a predictor of hearing loss after cochlear implantation: 3 and 12 month outcomes. Otol Neurotol 41(9):1222–1229
Verpy E, Weil D, Leibovici M et al (2008) Stereocilin-deficient mice reveal the origin of cochlear waveform distortions. Nature 456(7219):255–258
Patuzzi RB, Yates GK, Johnstone BM (1989) Outer hair cell receptor current and sensorineural hearing loss. Hear Res 42(1):47–72
Zhang M (2012) High-frequency hearing impairment assessed with cochlear microphonics. Acta Otolaryngol 132(9):967–973
Radeloff A, Shehata-Dieler W, Scherzed A et al (2012) Intraoperative monitoring using cochlear microphonics in cochlear implant patients with residual hearing. Otol Neurotol 33(3):348–354
Bester C, Razmovski T, Collins A et al (2020) Four-point impedance as a biomarker for bleeding during cochlear implantation. Sci Rep 10(1):2777
O’Connell BP, Holder JT, Dwyer RT et al (2017) Intra- and postoperative electrocochleography may be predictive of final electrode position and postoperative hearing preservation. Front Neurosci 11:291
O’Leary S, Mylanus E, Venail F et al (2023) Monitoring cochlear health with intracochlear electrocochleography during cochlear implantation: findings from an international clinical investigation. Ear Hear 44(2):358–370
Arweiler-Harbeck D, D’Heygere V, Meyer M et al (2021) Digital live imaging of intraoperative electrocochleography—first description of feasibility and hearing preservation during cochlear implantation. Otol Neurotol 42(9):1342–1346
Klabbers TM, Huinck WJ, Heutink F, Verbist BM, Mylanus EAM (2020) Transimpedance matrix (TIM) measurement for the detection of intraoperative electrode tip foldover using the slim modiolar electrode: a proof of concept study. Otol Neurotol. https://doi.org/10.1097/MAO.0000000000002875
Adunka OF, Giardina CK, Formeister EJ et al (2016) Round window electrocochleography before and after cochlear implant electrode insertion. Laryngoscope 126(5):1193–1200
Mittmann P, Mittmann M, Ernst A, Todt I (2017) Intracochlear pressure changes due to 2 electrode types: an artificial model experiment. Otolaryngol Head Neck Surg 156(4):712–716
Kontorinis G, Lenarz T, Stover T, Paasche G (2011) Impact of the insertion speed of cochlear implant electrodes on the insertion forces. Otol Neurotol 32(4):565–570
Lenarz T, Buechner A, Lesinski-Schiedat A, Timm M, Salcher R (2020) Hearing preservation with a new atraumatic lateral wall electrode. Otol Neurotol 41(8):e993–e1003
Iso-Mustajarvi M, Sipari S, Lopponen H, Dietz A (2020) Preservation of residual hearing after cochlear implant surgery with slim modiolar electrode. Eur Arch Otorhinolaryngol 277(2):367–375
Van Abel KM, Dunn CC, Sladen DP et al (2015) Hearing preservation among patients undergoing cochlear implantation. Otol Neurotol 36(3):416–421
Yin LX, Barnes JH, Saoji AA, Carlson ML (2021) Clinical utility of intraoperative electrocochleography (ECochG) during cochlear implantation: a systematic review and quantitative analysis. Otol Neurotol 42(3):363–371
Koka K, Saoji AA, Litvak LM (2017) Electrocochleography in cochlear implant recipients with residual hearing: comparison with audiometric thresholds. Ear Hear 38(3):e161–e167
Vittoria S, Lahlou G, Torres R et al (2020) Robot-based assistance in middle ear surgery and cochlear implantation: first clinical report. Eur Arch Otorhinolaryngol. https://doi.org/10.1007/s00405-020-06070-z
Acknowledgements
We thank the following companies (named alphabetically) for supplying test materials and enabling this study: Cochlear®, Munich Surgical Imaging®
Funding
Open Access funding enabled and organized by Projekt DEAL. Cochlear, IIR-2324, Diana Arweiler-Harbeck.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
This research was funded through a grant provided by Cochlear Research & Development Ltd under reference number IIR-2324.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Höing, B., Eichler, T., Juelly, V. et al. Digital live imaging of intraoperative electrocochleography during cochlear implantation: the first 50 patients. Eur Arch Otorhinolaryngol 281, 1175–1183 (2024). https://doi.org/10.1007/s00405-023-08197-1
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00405-023-08197-1