Abstract
Background
Injection laryngoplasty (IL) is a common procedure performed by the laryngologist to treat glottal insufficiency. It can be performed under general anaesthesia (GA) or as an office-based procedure. One of the common problems faced during IL is the disconnection between the injection needle and injection material syringe due to high-pressure effect. This may lead to leakage of the injectates especially when injecting high-viscosity materials such as calcium hydroxylapatite (CaHa) or injecting through hard fibrotic scar of the vocal fold.
Purpose
Due to this recurring issue, we propose anti-reflux valve as a connector to link up these two devices. The function of anti-reflux valve is to ensure secure connection between these two devices and overcome the problem.
Methods
The anti-reflux valve that can be used is NeutraClearTM needle-free connector EL-NC1000 or MicroClaveTM clear connector. In our practice, these anti-reflux valves are used together with Integra MicroFrance straight malleable injection needle (0.5 mm diameter, 250 mm length) for IL under GA. However, any other injection needle suitable for IL may also be used with these anti-reflux valves.
Results
Our 3 years’ experience has shown good outcome with no reported incidence of device detachment or leakage of injectates during IL procedure.
Conclusions
Anti-reflux valve is a readily available device in the operating theatre or clinic and only involve simple preparation prior to IL. It is beneficial as an added device in IL procedure.
Similar content being viewed by others
Data availability
Not applicable.
References
Bruening W (1911) Uber eine neue behandlungsmethode der rekurrensslahmung. Ver Deutsch Laryng 18:93–151 (Google Scholar)
Mallur PS, Rosen CA (2010) Vocal fold injection: review of indications, techniques, and materials for augmentation. Clin Exp Otorhinolaryngol 3(4):177–182. https://doi.org/10.3342/ceo.2010.3.4.177
Berteau C, Filipe-Santos O, Wang T, Rojas HE, Granger C, Schwarzenbach F (2015) Evaluation of the impact of viscosity, injection volume, and injection flow rate on subcutaneous injection tolerance. Med Devices (Auckl) 8:473–484. https://doi.org/10.2147/MDER.S91019. (published 2015 Nov 11)
Chhetri DK, Jamal N (2014) Percutaneous injection laryngoplasty. Laryngoscope 124(3):742–745. https://doi.org/10.1002/lary.24417
Kamarudin N, Abdul Rahim N, Azman M et al (2022) Subchordal cyst containing non-degraded hyaluronic acid: a rare complication of injection laryngoplasty. Cureus 14(8):e27801. https://doi.org/10.7759/cureus.27801
Chhetri DK, Jahan-Parwar B, Hart SD, Bhuta SM, Berke GS (2004) Injection laryngoplasty with calcium hydroxylapatite gel implant in an in vivo canine model. Ann Otol Rhinol Laryngol 113(4):259–264
Tabak YP, Jarvis WR, Sun X, Crosby CT, Johannes RS (2014) Meta-analysis on central line-associated bloodstream infections associated with a needleless intravenous connector with a new engineering design. Am J Infect Control 42(12):1278–1284. https://doi.org/10.1016/j.ajic.2014.08.018. (Epub 2014 Nov 25, PMID: 25465257)
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
The authors declare that they have no conflict of interest.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Springer Nature or its licensor (e.g. a society or other partner) holds exclusive rights to this article under a publishing agreement with the author(s) or other rightsholder(s); author self-archiving of the accepted manuscript version of this article is solely governed by the terms of such publishing agreement and applicable law.
About this article
Cite this article
Abdul Rahim, N., Laurent, I. & Remacle, M. Securing injection laryngoplasty with anti-reflux valve. Eur Arch Otorhinolaryngol 280, 3027–3029 (2023). https://doi.org/10.1007/s00405-023-07969-z
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00405-023-07969-z