Skip to main content
Log in

Contrast-detail evaluation and dose assessment of eight digital chest radiography systems in clinical practice

  • Chest
  • Published:
European Radiology Aims and scope Submit manuscript

Abstract

The purpose of this study was to assess contrast-detail performance and effective dose of eight different digital chest radiography systems. Digital chest radiography systems from different manufacturers were included: one storage phosphor system, one selenium-coated drum system, and six direct readout systems including four thin-film transistor (TFT) systems and two charge-coupled device (CCD) systems. For measuring image quality, a contrast-detail test object was used in combination with a phantom that simulates the primary and scatter transmission through lung fields (LucAl). Six observers judged phantom images of each modality by soft-copy reading in a four-alternative-forced-choice experiment. The entrance dose was also measured, and the effective dose was calculated for an average patient. Contrast-detail curves were constructed from the observer data. The blocked two-way ANOVA test was used for statistical analysis. Significant difference in contrast-detail performance was found between the systems. Best contrast-detail performance was shown by a CCD system with slot-scan technology, and the selenium-coated drum system was compared to the other six systems (p values ≤0.003). Calculated effective dose varied between 0.010 mSv and 0.032 mSv. Significant differences in contrast-detail performance and effective dose levels were found between different digital chest radiography systems in clinical practice.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Institutional subscriptions

Fig. 1
Fig. 2
Fig. 3

Similar content being viewed by others

References

  1. Kotter E, Langer M (2002) Digital radiography with large-area flat-panel detectors. Eur Radiol 12:2562–2570

    PubMed  Google Scholar 

  2. Schaefer-Prokop C, Uffmann M, Eisenhuber E, Prokop M (2003) Digital radiography of the chest: detector techniques and performance parameters. J Thorac Imaging 18:124–137

    Article  PubMed  Google Scholar 

  3. Huda W, Slone R. Review of radiologic physics 1995 Lippincott Wiliams & Wilkins, USA

    Google Scholar 

  4. International Commission on Radiation Units and Measurements (1996) Medical imaging—the assessment of image quality. ICRU Report no. 54 Bethesda, Md: International Commission on Radiation Units and Measurements, p. 54

    Google Scholar 

  5. Thijssen MA, Thijssen HO, Merx JL, van Woensel MP (1988) Quality analysis of DSA equipment. Neuroradiology 30:561–568

    Article  PubMed  Google Scholar 

  6. Conway BJ, Butler PF, Duff JE et al (1984) Beam quality independent attenuation phantom for estimating patient exposure from x-ray automatic exposure controlled chest examinations. Med Phys 11:827–832

    Article  PubMed  Google Scholar 

  7. AAPM Report No. 73, American Association of Physicists in Medicine, Quality Control in Diagnostic Radiology (2002) Diagnostic X-ray Imaging Committee Task Group No. 12, July 2002

  8. Peer S, Giacomuzzi SM, Peer R, Gassner E, Steingruber I, Jaschke W (2003) Resolution requirements for monitor viewing of digital flat-panel detector radiographs: a contrast detail analysis. Eur Radiol 13:413–417

    PubMed  Google Scholar 

  9. Servomaa A, Tapoivaara M (1998) Organ dose calculation in medical x-ray examinations by the program PCXMC. Radiat Prot Dosim 80:213–219

    Google Scholar 

  10. Veldkamp WJ, Thijssen MA, Karssemeijer N (2003) The value of scatter removal by a grid in full field digital mammography. Med Phys 30:1712–1718

    Article  PubMed  Google Scholar 

  11. Samei E, Saunders RS, Lo JY, Dobbins JT III, Jesneck JL, Floyd CE, Ravin CE (2004) Fundamental imaging characteristics of a slot-scan digital chest radiographic system. Med Phys 31:1298–2687

    Article  Google Scholar 

  12. Diekmann F, Diekmann S, Richter K, Bick U, Fischer T, Lawaczeck R, Press WR, Schon K, Weinmann HJ, Arkadiev V, Bjeoumikhov A, Langhoff N, Rabe J, Roth P, Tilgner J, Wedell R, Krumrey M, Linke U, Ulm G, Hamm B (2004) Near monochromatic X-rays for digital slot-scan mammography: initial findings. Eur Radiol 14:1641–1646

    Article  PubMed  Google Scholar 

  13. Veldkamp WJ, Kroft LJ, Mertens BJ, Geleijns J (2005) Comparison of image quality between a digital slot-scan CCD chest radiography system and AMBER and Bucky screen-film radiography chest systems. Radiology 235:857–866

    PubMed  Google Scholar 

  14. Kroft LJ, Geleijns J, Mertens BJ, Veldkamp WJ, Zonderland HM, de Roos A (2004) Digital slot-scan charged coupled device chest radiography versus AMBER and Bucky screen-film radiography: detection of simulated chest nodules and interstitial disease using a chest phantom. Radiology 231:156–163

    PubMed  Google Scholar 

  15. Kroft LJ, Veldkamp WJ, Mertens BJ, Boot MV, Geleijns J. Comparison of eight digital chest radiography systems: variation in detection of simulated chest disease. Am J Roengenol 185:339–346

  16. Bernhardt TM, Rapp-Bernhardt U, Hausmann T, Reichel G, Krause UW, Doehring W (2000) Digital selenium radiography: anti-scatter grid for chest radiography in a clinical study. Br J Radiol 73:963–968

    PubMed  Google Scholar 

  17. Neitzel U, Maack I, Gunther-Kohfahl S (1994) Image quality of a digital chest radiography system based on a selenium detector. Med Phys 21:509–516

    Article  PubMed  Google Scholar 

  18. Samei E, Flynn MJ (2003) An experimental comparison of detector performance for direct and indirect digital radiography systems. Med Phys 30:608–622

    Article  PubMed  Google Scholar 

  19. Awai K, Komi M, Hori S (2001) Selenium-based digital radiography versus high-resolution storage phosphor radiography in the detection of solitary pulmonary nodules without calcification: receiver operating characteristic curve analysis. Am J Roentgenol 177:1141–1144

    Google Scholar 

  20. Goo JM, Im J-G, Kim JH, et al (2000) Digital chest radiography with selenium-based flat-panel detector versus a storage phosphor system: comparison of soft-copy images. Am J Roentgenol 175:1013–1018

    Google Scholar 

  21. Mansson LG, Kheddache S, Lanhede B, Tylen U (1999) Image quality for five modern chest radiography techniques: a modified FROC study with an anthropomorphic chest phantom. Eur Radiol 9:1826–1834

    Article  PubMed  Google Scholar 

  22. Beute GH, Flynn MJ, Eyler WR, Samei E, Spizarny DL, Zylak CJ (1998) Chest radiographic image quality: comparison of asymmetric screen-film, digital storage phosphor, and digital selenium drum systems—preliminary study. Radiographics 18:745–754

    PubMed  Google Scholar 

  23. Borasi G, Nitrosi A, Ferrari P, Tassoni D (2003) On site evaluation of three flat panel detectors for digital radiography. Med Phys 30:1719–1731

    Article  PubMed  Google Scholar 

  24. Samei E (2003) Image quality in two phosphor-based flat panel digital radiographic detectors. Med Phys 30:1747–1757

    Article  PubMed  Google Scholar 

  25. Granfors PR, Aufrichtig R (2000) Performance of a 41×41-cm2 amorphous silicon flat panel x-ray detector for radiographic imaging applications. Med Phys 27:1324–1331

    Article  PubMed  Google Scholar 

  26. Volk M, Hamer OW, Feuerbach S, Strotzer M (2004) Dose reduction in skeletal and chest radiography using a large-area flat-panel detector based on amorphous silicon and thallium-doped cesium iodide: technical background, basic image quality parameters, and review of the literature. Eur Radiol 14:827–834

    Article  PubMed  Google Scholar 

  27. Bath M, Sund P, Mansson LG (2002) Evaluation of the imaging properties of two generations of a CCD-based system for digital chest radiography. Med Phys 29:2286–2297

    Article  PubMed  Google Scholar 

  28. Herrmann KA, Bonél H, Stäbler A et al (2002) Chest imaging with flat-panel detector at low and standard doses: comparison with storage phosphor technology in normal patients. Eur Radiol 12:385–390

    Article  PubMed  Google Scholar 

  29. Bacher K, Smeets P, Bonnarens K, De Hauwere A, Verstraete K, Thierens H (2003) Dose reduction in patients undergoing chest imaging: digital amorphous silicon flat-panel detector radiography versus conventional film-screen radiography and phosphor-based computed radiography. Am J Roentgenol 181:923–929

    Google Scholar 

  30. Goo JM, Im J-G, Lee HJ et al (2002) Detection of simulated chest lesions by using soft-copy reading: comparison of an amorphous silicon flat-panel detector system and a storage-phosphor system. Radiology 224:242–246

    PubMed  Google Scholar 

  31. Aufrichtig R (1999) Comparison of low contrast detectability between a digital amorphous silicon and a screen-film based imaging system for thoracic radiography. Med Phys 26:1349–1358

    Article  PubMed  Google Scholar 

  32. Sund P, Bath M, Kheddache S, Mansson LG (2004) Comparison of visual grading analysis and determination of detective quantum efficiency for evaluating system performance in digital chest radiography. Eur Radiol 14:48–58

    Article  PubMed  Google Scholar 

  33. Burgess AE, Jacobson FL, Judy PF (2001) Human observer detection experiments with mammograms and power-law noise. Med Phys 28:419–437

    Article  PubMed  Google Scholar 

Download references

Acknowledgements

The authors gratefully acknowledge the participation of the following persons in the panel of observers: J.P. van Delft, V. Schembri MSc, and D. Zweers BSc (Leiden University Medical Center, Department of Radiology).

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Wouter J. H. Veldkamp.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Veldkamp, W.J.H., Kroft, L.J.M., Boot, M.V. et al. Contrast-detail evaluation and dose assessment of eight digital chest radiography systems in clinical practice. Eur Radiol 16, 333–341 (2006). https://doi.org/10.1007/s00330-005-2887-6

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00330-005-2887-6

Keywords

Navigation