Optimization of multidetector array CT acquisition parameters for CT colonography

  • Published:
Abdominal Imaging Aims and scope Submit manuscript

Abstract

Background: Multidetector array computed tomography technology has increased the number of acquisition parameters that a user must select. This article examines the criteria by which detector collimation, reconstructed scan width, reconstruction scan interval, pitch, image noise, and patient dose can be optimized.

Methods: A water phantom containing tissue-equivalent polyps in an air-filled tube was scanned using multiple acquisition parameter combinations. X-ray tube current was changed to independently match dose and noise to a single-detector array technique. Images were assessed in axial and three-dimensional formats.

Results: All polyps were visible in axial and three-dimensional images for all scans. Helical artifacts were noted at higher table speed values. The 5-mm-wide scans spaced at 3-mm intervals allowed visualization of the smallest polyps: 5 × 5 or 7 × 3 (diameter × height; mm).

Conclusion: Our results indicate that thin-scan and low-noise techniques need not be used to visualize polyps of clinically relevant sizes.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

Author information

Authors and Affiliations

Authors

Rights and permissions

Reprints and permissions

About this article

Cite this article

McCollough, C. Optimization of multidetector array CT acquisition parameters for CT colonography. Abdom Imaging 27, 253–259 (2002). https://doi.org/10.1007/s00261-001-0166-9

Download citation

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00261-001-0166-9

Navigation