The impact of stent design on the structural mechanics of the crossing Y-stent: an in vitro study

  • Interventional Neuroradiology
  • Published:
Neuroradiology Aims and scope Submit manuscript

Abstract

Introduction

Crossing Y-stent technique is a valid option for coiling wide-necked bifurcation aneurysms. Two main designs of stents, the closed- (CCS) and open-cell (OCS), are used in combination for a crossing Y-construct. This in vitro study was conducted to assess the mechanical characteristics of each stent-combination and to suggest an optimal combination for clinical practice.

Methods

The Enterprise and the Neuroform3 stents were used as closed-cell and open-cell stents, respectively. Four different Y-stent combinations; double CCSs (CCS-CCS; Enterprise-Enterprise), OCS followed by CCS (OCS-CCS; Neuroform-Enterprise), CCS followed by OCS (CCS-OCS; Enterprise-Neuroform), and double OCSs (OCS-OCS; Neuroform-Neuroform) were tested in a bifurcation aneurysm model. Images of the model were obtained by digital subtraction angiography (DSA), and morphological changes of Y-constructs caused by interaction between stents were compared.

Results

Double OCSs showed the best stent apposition to model branches and no collapse of the second stent at the intersection, whereas Y-stent combinations using a CCS as the second stent showed tubular collapse of the second stent at the crossing point. These combinations revealed unsatisfactory apposition to the model branch in which the second stent was deployed. Most narrowing of the second stent was noted in the double CCSs construct, which resulted in poorest stent apposition.

Conclusion

Based on a simple in vitro experiment, we suggest that double OCSs Y-construct is optimal for achieving best stent-wall apposition. Furthermore, our findings suggest that Y-stent combinations utilizing a CCS as the second stent may increase the risk of thromboembolic complications due to poor stent-wall apposition.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2

Similar content being viewed by others

References

  1. Chow MM, Woo HH, Masaryk TJ, Rasmussen PA (2004) A novel endovascular treatment of a wide-necked basilar apex aneurysm by using a Y-configuration, double-stent technique. AJNR Am J Neuroradiol 25:509–512

    PubMed  Google Scholar 

  2. Rohde S, Bendszus M, Hartmann M, Hähnel S (2010) Treatment of a wide-necked aneurysm of the anterior cerebral artery using two Enterprise stents in “Y”-configuration stenting technique and coil embolization: a technical note. Neuroradiology 52:231–235

    Article  PubMed  Google Scholar 

  3. Sani S, Lopes DK (2005) Treatment of a middle cerebral artery bifurcation aneurysm using a double neuroform stent “Y” configuration and coil embolization: technical case report. Neurosurgery 57(1 Suppl):E209

    Article  PubMed  Google Scholar 

  4. Spiotta AM, Gupta R, Fiorella D et al (2011) Mid-term results of endovascular coiling of wide-necked aneurysms using double stents in a Y configuration. Neurosurgery 69:421–429

    Article  PubMed  Google Scholar 

  5. Kanaan H, Jankowitz B, Aleu A et al (2010) In-stent thrombosis and stenosis after neck-remodeling device-assisted coil embolization of intracranial aneurysms. Neurosurgery 67:1523–1532

    Article  PubMed  Google Scholar 

  6. Mocco J, Fargen KM, Albuquerque FC et al (2011) Delayed thrombosis or stenosis following enterprise-assisted stent-coiling: is it safe? Midterm results of the interstate collaboration of enterprise stent coiling. Neurosurgery 69:908–913

    Article  CAS  PubMed  Google Scholar 

  7. Lee DH, Arat A, Morsi H, Shaltoni H, Harris JR, Mawad ME (2008) Dual antiplatelet therapy monitoring for neurointerventional procedures using a point-of-care platelet function test: a single-center experience. AJNR Am J Neuroradiol 29:1389–1394

    Article  CAS  PubMed  Google Scholar 

  8. Yamada NK, Cross DT 3rd, Pilgram TK, Moran CJ, Derdeyn CP, Dacey RG Jr (2007) Effect of antiplatelet therapy on thromboembolic complications of elective coil embolization of cerebral aneurysms. AJNR Am J Neuroradiol 28:1778–1782

    Article  CAS  PubMed  Google Scholar 

  9. Heller RS, Miele WR, Do-Dai DD, Malek AM (2011) Crescent sign on magnetic resonance angiography revealing incomplete stent apposition: correlation with diffusion-weighted changes in stent-mediated coil embolization of aneurysms. Neurosurgery 115:624–632

    Article  Google Scholar 

  10. Buchanan GL, Basavarajaiah S, Chieffo A (2012) Stent thrombosis: incidence, predictors and new technologies. Thrombosis 2012:956962

    Article  PubMed Central  PubMed  Google Scholar 

  11. Capranzano P, Dangas G (2012) Late stent thrombosis: the last remaining obstacle in coronary interventional therapy. Curr Cardiol Rep 14:408–417

    Article  PubMed  Google Scholar 

  12. Cook S, Wenaweser P, Togni M et al (2007) Incomplete stent apposition and very late stent thrombosis after drug-eluting stent implantation. Circulation 115:2426–2434

    Article  CAS  PubMed  Google Scholar 

  13. Tanaka N, Martin JB, Tokunaga K et al (2004) Conformity of carotid stents with vascular anatomy: evaluation in carotid models. AJNR Am J Neuroradiol 25:604–607

    PubMed  Google Scholar 

  14. Cekirge HS, Yavuz K, Geyik S, Saatci I (2011) A novel “Y” stent flow diversion technique for the endovascular treatment of bifurcation aneurysms without endosaccular coiling. AJNR Am J Neuroradiol 32:1262–1268

    Article  CAS  PubMed  Google Scholar 

  15. Lozen A, Manjila S, Rhiew R, Fessler R (2009) Y-stent-assisted coil embolization for the management of unruptured cerebral aneurysms: report of six cases. Acta Neurochir (Wien) 151:1663–1672

    Article  Google Scholar 

  16. Akgul E, Aksungur E, Balli T et al (2011) Y-stent-assisted coil embolization of wide-neck intracranial aneurysms. A single center experience. Interv Neuroradiol 17:36–48

    CAS  PubMed Central  PubMed  Google Scholar 

  17. Kono K, Shintani A, Yoshimura R et al (2013) Triple antiplatelet therapy with addition of cilostazol to aspirin and clopidogrel for Y-stent-assisted coil embolization of cerebral aneurysms. Acta Neurochir (Wien) 155:1549–1557

    Article  Google Scholar 

  18. Thorell WE, Chow MM, Woo HH, Masaryk TJ, Rasmussen PA (2005) Y-configured dual intracranial stent-assisted coil embolization for the treatment of wide-necked basilar tip aneurysms. Neurosurgery 56:1035–1040

    PubMed  Google Scholar 

  19. De Bock S, Iannaccone F, De Santis G et al (2012) Our capricious vessels: the influence of stent design and vessel geometry on the mechanics of intracranial aneurysm stent deployment. J Biomech 45:1353–1359

    Article  PubMed  Google Scholar 

  20. Ebrahimi N, Claus B, Lee CY, Biondi A, Benndorf G (2007) Stent conformity in curved vascular models with simulated aneurysm necks using flat-panel CT: an in vitro study. AJNR Am J Neuroradiol 28:823–829

    CAS  PubMed  Google Scholar 

  21. Sedat J, Chau Y, Mondot L, Vargas J, Szapiro J, Lonjon M (2009) Endovascular occlusion of intracranial wide-necked aneurysms with stenting (Neuroform) and coiling: mid-term and long-term results. Neuroradiology 51:401–409

    Article  PubMed  Google Scholar 

  22. Peluso JP, van Rooij WJ, Sluzewski M, Beute GN (2008) A new self-expandable nitinol stent for the treatment of wide-neck aneurysms: initial clinical experience. AJNR Am J Neuroradiol 29:1405–1408

    Article  CAS  PubMed  Google Scholar 

  23. Klisch J, Eger C, Sychra V, Strasilla C, Basche S, Weber J (2009) Stent-assisted coil embolization of posterior circulation aneurysms using solitaire ab: preliminary experience. Neurosurgery 65:258–266

    Article  PubMed  Google Scholar 

  24. Valdivia y Alvarado M, Ebrahimi N, Benndorf G (2009) Study of conformability of the new leo plus stent to a curved vascular model using flat-panel detector computed tomography (DynaCT). Neurosurgery 64:130–134

    Google Scholar 

  25. Conrad MD, Brasiliense LB, Richie AN, Hanel RA (2013) Y stenting assisted coiling using a new low profile visible intraluminal support device for wide necked basilar tip aneurysms: a technical report. J Neurointerv Surg 0:1–5

    Google Scholar 

  26. Krischek O, Miloslavski E, Fischer S, Shrivastava S, Henkes H (2011) A comparison of functional and physical properties of self-expanding intracranial stents [Neuroform3, Wingspan, Solitaire, Leo+, Enterprise]. Minim Invasive Neurosurg 54:21–28

    Article  CAS  PubMed  Google Scholar 

  27. Kadkhodayan Y, Rhodes N, Blackburn S, Derdeyn CP, Cross DT 3rd, Moran CJ (2013) Comparison of Enterprise with Neuroform stent-assisted coiling of intracranial aneurysms. AJR Am J Roentgenol 200:872–878

    Article  PubMed  Google Scholar 

  28. Lijnen HR, Collen D (1997) Endothelium in hemostasis and thrombosis. Prog Cardiovasc Dis 39:343–350

    Article  CAS  PubMed  Google Scholar 

  29. Fargen KM, Mocco J, Neal D et al (2013) A multicenter study of stent-assisted coiling of cerebral aneurysms with a Y configuration. Neurosurgery 73:466–472

    Article  PubMed  Google Scholar 

Download references

Ethical standards and patient consent

We declare that this manuscript does not contain clinical studies or patient data.

Conflict of interest

We declare that we have no conflict of interest.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Chang-Young Lee.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Lee, CY., Park, SH., Kim, CH. et al. The impact of stent design on the structural mechanics of the crossing Y-stent: an in vitro study. Neuroradiology 56, 655–659 (2014). https://doi.org/10.1007/s00234-014-1378-3

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00234-014-1378-3

Keywords

Navigation