Abstract
Introduction
Crossing Y-stent technique is a valid option for coiling wide-necked bifurcation aneurysms. Two main designs of stents, the closed- (CCS) and open-cell (OCS), are used in combination for a crossing Y-construct. This in vitro study was conducted to assess the mechanical characteristics of each stent-combination and to suggest an optimal combination for clinical practice.
Methods
The Enterprise and the Neuroform3 stents were used as closed-cell and open-cell stents, respectively. Four different Y-stent combinations; double CCSs (CCS-CCS; Enterprise-Enterprise), OCS followed by CCS (OCS-CCS; Neuroform-Enterprise), CCS followed by OCS (CCS-OCS; Enterprise-Neuroform), and double OCSs (OCS-OCS; Neuroform-Neuroform) were tested in a bifurcation aneurysm model. Images of the model were obtained by digital subtraction angiography (DSA), and morphological changes of Y-constructs caused by interaction between stents were compared.
Results
Double OCSs showed the best stent apposition to model branches and no collapse of the second stent at the intersection, whereas Y-stent combinations using a CCS as the second stent showed tubular collapse of the second stent at the crossing point. These combinations revealed unsatisfactory apposition to the model branch in which the second stent was deployed. Most narrowing of the second stent was noted in the double CCSs construct, which resulted in poorest stent apposition.
Conclusion
Based on a simple in vitro experiment, we suggest that double OCSs Y-construct is optimal for achieving best stent-wall apposition. Furthermore, our findings suggest that Y-stent combinations utilizing a CCS as the second stent may increase the risk of thromboembolic complications due to poor stent-wall apposition.
Similar content being viewed by others
References
Chow MM, Woo HH, Masaryk TJ, Rasmussen PA (2004) A novel endovascular treatment of a wide-necked basilar apex aneurysm by using a Y-configuration, double-stent technique. AJNR Am J Neuroradiol 25:509–512
Rohde S, Bendszus M, Hartmann M, Hähnel S (2010) Treatment of a wide-necked aneurysm of the anterior cerebral artery using two Enterprise stents in “Y”-configuration stenting technique and coil embolization: a technical note. Neuroradiology 52:231–235
Sani S, Lopes DK (2005) Treatment of a middle cerebral artery bifurcation aneurysm using a double neuroform stent “Y” configuration and coil embolization: technical case report. Neurosurgery 57(1 Suppl):E209
Spiotta AM, Gupta R, Fiorella D et al (2011) Mid-term results of endovascular coiling of wide-necked aneurysms using double stents in a Y configuration. Neurosurgery 69:421–429
Kanaan H, Jankowitz B, Aleu A et al (2010) In-stent thrombosis and stenosis after neck-remodeling device-assisted coil embolization of intracranial aneurysms. Neurosurgery 67:1523–1532
Mocco J, Fargen KM, Albuquerque FC et al (2011) Delayed thrombosis or stenosis following enterprise-assisted stent-coiling: is it safe? Midterm results of the interstate collaboration of enterprise stent coiling. Neurosurgery 69:908–913
Lee DH, Arat A, Morsi H, Shaltoni H, Harris JR, Mawad ME (2008) Dual antiplatelet therapy monitoring for neurointerventional procedures using a point-of-care platelet function test: a single-center experience. AJNR Am J Neuroradiol 29:1389–1394
Yamada NK, Cross DT 3rd, Pilgram TK, Moran CJ, Derdeyn CP, Dacey RG Jr (2007) Effect of antiplatelet therapy on thromboembolic complications of elective coil embolization of cerebral aneurysms. AJNR Am J Neuroradiol 28:1778–1782
Heller RS, Miele WR, Do-Dai DD, Malek AM (2011) Crescent sign on magnetic resonance angiography revealing incomplete stent apposition: correlation with diffusion-weighted changes in stent-mediated coil embolization of aneurysms. Neurosurgery 115:624–632
Buchanan GL, Basavarajaiah S, Chieffo A (2012) Stent thrombosis: incidence, predictors and new technologies. Thrombosis 2012:956962
Capranzano P, Dangas G (2012) Late stent thrombosis: the last remaining obstacle in coronary interventional therapy. Curr Cardiol Rep 14:408–417
Cook S, Wenaweser P, Togni M et al (2007) Incomplete stent apposition and very late stent thrombosis after drug-eluting stent implantation. Circulation 115:2426–2434
Tanaka N, Martin JB, Tokunaga K et al (2004) Conformity of carotid stents with vascular anatomy: evaluation in carotid models. AJNR Am J Neuroradiol 25:604–607
Cekirge HS, Yavuz K, Geyik S, Saatci I (2011) A novel “Y” stent flow diversion technique for the endovascular treatment of bifurcation aneurysms without endosaccular coiling. AJNR Am J Neuroradiol 32:1262–1268
Lozen A, Manjila S, Rhiew R, Fessler R (2009) Y-stent-assisted coil embolization for the management of unruptured cerebral aneurysms: report of six cases. Acta Neurochir (Wien) 151:1663–1672
Akgul E, Aksungur E, Balli T et al (2011) Y-stent-assisted coil embolization of wide-neck intracranial aneurysms. A single center experience. Interv Neuroradiol 17:36–48
Kono K, Shintani A, Yoshimura R et al (2013) Triple antiplatelet therapy with addition of cilostazol to aspirin and clopidogrel for Y-stent-assisted coil embolization of cerebral aneurysms. Acta Neurochir (Wien) 155:1549–1557
Thorell WE, Chow MM, Woo HH, Masaryk TJ, Rasmussen PA (2005) Y-configured dual intracranial stent-assisted coil embolization for the treatment of wide-necked basilar tip aneurysms. Neurosurgery 56:1035–1040
De Bock S, Iannaccone F, De Santis G et al (2012) Our capricious vessels: the influence of stent design and vessel geometry on the mechanics of intracranial aneurysm stent deployment. J Biomech 45:1353–1359
Ebrahimi N, Claus B, Lee CY, Biondi A, Benndorf G (2007) Stent conformity in curved vascular models with simulated aneurysm necks using flat-panel CT: an in vitro study. AJNR Am J Neuroradiol 28:823–829
Sedat J, Chau Y, Mondot L, Vargas J, Szapiro J, Lonjon M (2009) Endovascular occlusion of intracranial wide-necked aneurysms with stenting (Neuroform) and coiling: mid-term and long-term results. Neuroradiology 51:401–409
Peluso JP, van Rooij WJ, Sluzewski M, Beute GN (2008) A new self-expandable nitinol stent for the treatment of wide-neck aneurysms: initial clinical experience. AJNR Am J Neuroradiol 29:1405–1408
Klisch J, Eger C, Sychra V, Strasilla C, Basche S, Weber J (2009) Stent-assisted coil embolization of posterior circulation aneurysms using solitaire ab: preliminary experience. Neurosurgery 65:258–266
Valdivia y Alvarado M, Ebrahimi N, Benndorf G (2009) Study of conformability of the new leo plus stent to a curved vascular model using flat-panel detector computed tomography (DynaCT). Neurosurgery 64:130–134
Conrad MD, Brasiliense LB, Richie AN, Hanel RA (2013) Y stenting assisted coiling using a new low profile visible intraluminal support device for wide necked basilar tip aneurysms: a technical report. J Neurointerv Surg 0:1–5
Krischek O, Miloslavski E, Fischer S, Shrivastava S, Henkes H (2011) A comparison of functional and physical properties of self-expanding intracranial stents [Neuroform3, Wingspan, Solitaire, Leo+, Enterprise]. Minim Invasive Neurosurg 54:21–28
Kadkhodayan Y, Rhodes N, Blackburn S, Derdeyn CP, Cross DT 3rd, Moran CJ (2013) Comparison of Enterprise with Neuroform stent-assisted coiling of intracranial aneurysms. AJR Am J Roentgenol 200:872–878
Lijnen HR, Collen D (1997) Endothelium in hemostasis and thrombosis. Prog Cardiovasc Dis 39:343–350
Fargen KM, Mocco J, Neal D et al (2013) A multicenter study of stent-assisted coiling of cerebral aneurysms with a Y configuration. Neurosurgery 73:466–472
Ethical standards and patient consent
We declare that this manuscript does not contain clinical studies or patient data.
Conflict of interest
We declare that we have no conflict of interest.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Lee, CY., Park, SH., Kim, CH. et al. The impact of stent design on the structural mechanics of the crossing Y-stent: an in vitro study. Neuroradiology 56, 655–659 (2014). https://doi.org/10.1007/s00234-014-1378-3
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00234-014-1378-3