Flow Redirection Endoluminal Device (FRED) with or without Adjunctive Coiling in Treatment of Very Large and Giant Cerebral Aneurysms

  • Original Article
  • Published:
Clinical Neuroradiology Aims and scope Submit manuscript

Abstract

Purpose

Flow diverter stents are gaining wider use in the treatment of complex intracranial aneurysms; however, there are limited data on occlusion and complication rates of flow diverters in very large and giant aneurysms. This study assessed the safety and efficacy of flow redirection endoluminal device (FRED) and FRED Jr. stents in aneurysms ≥ 20 mm.

Methods

We retrospectively analyzed all aneurysms ≥ 20 mm treated with FRED/FRED Jr. between January 2010 and June 2020 from three centers. Endpoints for clinical safety were absence of major stroke, aneurysmal rupture, or death and complete or near-complete occlusion for efficacy.

Results

A total of 45 patients with very large (28 cases) and giant aneurysms (17 cases) were treated with FRED (41 cases), and FRED Jr. (4 cases) stents. The majority of the aneurysms (40/45, 88.9%) were in the anterior circulation. Adjunctive aneurysm coiling was performed in 21 aneurysms (46.7%).

Technical complications were encountered in 4 procedures (8.9%). Ischemic and hemorrhagic complication rates were 6.7% and 8.9%, respectively. There was no case with major stroke. The mortality rate was 8.9%, and all cases were due to hemorrhagic complications from aneurysmal rupture.

Median angiographic follow-up was 21.5 months (range 3–60 months). Complete occlusion was achieved in 32 aneurysms (71.1%), and near-complete occlusion in 5 cases (11.1%).

Conclusion

The use of FRED/FRED Jr. for the treatment of aneurysms ≥ 20 mm achieved a long-term aneurysm occlusion rate and a safety profile comparable to the large series reported with other flow diverter stents.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3

Similar content being viewed by others

References

  1. Wiebers DO, Whisnant JP, Huston J 3rd, Meissner I, Brown RD Jr, Piepgras DG, Forbes GS, Thielen K, Nichols D, O’Fallon WM, Peacock J, Jaeger L, Kassell NF, Kongable-Beckman GL, Torner JC; International Study of Unruptured Intracranial Aneurysms Investigators. Unruptured intracranial aneurysms: natural history, clinical outcome, and risks of surgical and endovascular treatment. Lancet. 2003;362:103–10.

    Article  PubMed  Google Scholar 

  2. Cagnazzo F, Mantilla D, Rouchaud A, Brinjikji W, Lefevre PH, Dargazanli C, Gascou G, Riquelme C, Perrini P, di Carlo D, Bonafe A, Costalat V. Endovascular Treatment of Very Large and Giant Intracranial Aneurysms: Comparison between Reconstructive and Deconstructive Techniques-A Meta-Analysis. AJNR Am J Neuroradiol. 2018;39:852–8.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  3. Biondi A, Jean B, Vivas E, Le Jean L, Boch AL, Chiras J, Van Effenterre R. Giant and large peripheral cerebral aneurysms: etiopathologic considerations, endovascular treatment, and long-term follow-up. AJNR Am J Neuroradiol. 2006;27:1685–92.

    CAS  PubMed  PubMed Central  Google Scholar 

  4. Kocer N. Giant intracranial aneurysms: a case-based atlas of imaging and treatment. New York: Springer; 2016.

    Book  Google Scholar 

  5. Becske T, Kallmes DF, Saatci I, McDougall CG, Szikora I, Lanzino G, Moran CJ, Woo HH, Lopes DK, Berez AL, Cher DJ, Siddiqui AH, Levy EI, Albuquerque FC, Fiorella DJ, Berentei Z, Marosfoi M, Cekirge SH, Nelson PK. Pipeline for uncoilable or failed aneurysms: results from a multicenter clinical trial. Radiology. 2013;267:858–68.

    Article  PubMed  Google Scholar 

  6. Adeeb N, Griessenauer CJ, Shallwani H, Shakir H, Foreman PM, Moore JM, Dmytriw AA, Gupta R, Siddiqui AH, Levy EI, Snyder K, Harrigan MR, Ogilvy CS, Thomas AJ. Pipeline Embolization Device in Treatment of 50 Unruptured Large and Giant Aneurysms. World Neurosurg. 2017;105:232–7.

    Article  PubMed  Google Scholar 

  7. Peschillo S, Caporlingua A, Resta MC, Peluso JPP, Burdi N, Sourour N, Diana F, Guidetti G, Clarençon F, Bloemsma GC, Di Maria F, Donatelli M, Resta M. Endovascular Treatment of Large and Giant Carotid Aneurysms with Flow-Diverter Stents Alone or in Combination with Coils: A Multicenter Experience and Long-Term Follow-up. Oper Neurosurg (Hagerstown). 2017;13:492–502.

    Article  Google Scholar 

  8. Killer-Oberpfalzer M, Kocer N, Griessenauer CJ, Janssen H, Engelhorn T, Holtmannspötter M, Buhk JH, Finkenzeller T, Fesl G, Trenkler J, Reith W, Berlis A, Hausegger K, Augustin M, Islak C, Minnich B, Möhlenbruch M. European Multicenter Study for the Evaluation of a Dual-Layer Flow-Diverting Stent for Treatment of Wide-Neck Intracranial Aneurysms: The European Flow-Redirection Intraluminal Device Study. AJNR Am J Neuroradiol. 2018;39:841–7.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  9. O’Kelly CJ, Krings T, Fiorella D, Marotta TR. A novel grading scale for the angiographic assessment of intracranial aneurysms treated using flow diverting stents. Interv Neuroradiol. 2010;16:133–7.

    Article  PubMed  PubMed Central  Google Scholar 

  10. Brinjikji W, Murad MH, Lanzino G, Cloft HJ, Kallmes DF. Endovascular treatment of intracranial aneurysms with flow diverters: a meta-analysis. Stroke. 2013;44:442–7.

    Article  PubMed  Google Scholar 

  11. Becske T, Brinjikji W, Potts MB, Kallmes DF, Shapiro M, Moran CJ, Levy EI, McDougall CG, Szikora I, Lanzino G, Woo HH, Lopes DK, Siddiqui AH, Albuquerque FC, Fiorella DJ, Saatci I, Cekirge SH, Berez AL, Cher DJ, Berentei Z, Marosfoi M, Nelson PK. Long-Term Clinical and Angiographic Outcomes Following Pipeline Embolization Device Treatment of Complex Internal Carotid Artery Aneurysms: Five-Year Results of the Pipeline for Uncoilable or Failed Aneurysms Trial. Neurosurgery. 2017;80:40–8.

    Article  PubMed  Google Scholar 

  12. Molyneux AJ, Ellison DW, Morris J, Byrne JV. Histological findings in giant aneurysms treated with Guglielmi detachable coils. Report of two cases with autopsy correlation. J Neurosurg. 1995;83:129–32.

    Article  CAS  PubMed  Google Scholar 

  13. Szikora I, Turányi E, Marosfoi M. Evolution of Flow-Diverter Endothelialization and Thrombus Organization in Giant Fusiform Aneurysms after Flow Diversion: A Histopathologic Study. AJNR Am J Neuroradiol. 2015;36:1716–20.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  14. Slater LA, Soufan C, Holt M, Chong W. Effect of flow diversion with silk on aneurysm size: A single center experience. Interv Neuroradiol. 2015;21:12–8. Erratum in: Interv Neuroradiol. 2015;21:282.

    Article  PubMed  PubMed Central  Google Scholar 

  15. Szikora I, Marosfoi M, Salomváry B, Berentei Z, Gubucz I. Resolution of mass effect and compression symptoms following endoluminal flow diversion for the treatment of intracranial aneurysms. AJNR Am J Neuroradiol. 2013;34:935–9.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  16. Becske T, Potts MB, Shapiro M, Kallmes DF, Brinjikji W, Saatci I, McDougall CG, Szikora I, Lanzino G, Moran CJ, Woo HH, Lopes DK, Berez AL, Cher DJ, Siddiqui AH, Levy EI, Albuquerque FC, Fiorella DJ, Berentei Z, Marosföi M, Cekirge SH, Nelson PK. Pipeline for uncoilable or failed aneurysms: 3-year follow-up results. J Neurosurg. 2017;127:81–8.

    Article  PubMed  Google Scholar 

  17. Briganti F, Leone G, Cirillo L, de Divitiis O, Solari D, Cappabianca P. Postprocedural, midterm, and long-term results of cerebral aneurysms treated with flow-diverter devices: 7-year experience at a single center. Neurosurg Focus. 2017;42:E3.

    Article  PubMed  Google Scholar 

  18. Kallmes DF, Hanel R, Lopes D, Boccardi E, Bonafé A, Cekirge S, Fiorella D, Jabbour P, Levy E, McDougall C, Siddiqui A, Szikora I, Woo H, Albuquerque F, Bozorgchami H, Dashti SR, Delgado Almandoz JE, Kelly ME, Turner R 4th, Woodward BK, Brinjikji W, Lanzino G, Lylyk P. International retrospective study of the pipeline embolization device: a multicenter aneurysm treatment study. AJNR Am J Neuroradiol. 2015;36:108–15. Erratum in: AJNR Am J Neuroradiol. 2015;36:E39–40.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  19. Griessenauer CJ, Möhlenbruch MA, Hendrix P, Ulfert C, Islak C, Sonnberger M, Engelhorn T, Müller-Thies-Broussalis E, Finkenzeller T, Holtmannspötter M, Buhk JH, Reith W, Simgen A, Janssen H, Kocer N, Killer-Oberpfalzer M. The FRED for Cerebral Aneurysms of the Posterior Circulation: A Subgroup Analysis of the EuFRED Registry. AJNR Am J Neuroradiol. 2020;41:658–62.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  20. Rouchaud A, Brinjikji W, Lanzino G, Cloft HJ, Kadirvel R, Kallmes DF. Delayed hemorrhagic complications after flow diversion for intracranial aneurysms: a literature overview. Neuroradiology. 2016;58:171–7.

    Article  PubMed  Google Scholar 

  21. Griessenauer CJ, Thomas AJ, Enriquez-Marulanda A, Deshmukh A, Jain A, Ogilvy CS, Kocer N, Engelhorn T, Möhlenbruch M, Holtmannspötter M, Janssen H, Finkenzeller T, Reith W, Sonnberger M, Buhk JH, Schirmer CM, Killer-Oberpfalzer M. Comparison of Pipeline Embolization Device and Flow Re-Direction Endoluminal Device Flow Diverters for Internal Carotid Artery Aneurysms: A Propensity Score-Matched Cohort Study. Neurosurgery. 2019;85:E249–55.

    Article  PubMed  Google Scholar 

  22. Griessenauer CJ, Enriquez-Marulanda A, Xiang S, Hong T, Zhang H, Taussky P, Grandhi R, Waqas M, Tutino VM, Siddiqui AH, Levy EI, Ogilvy CS, Thomas AJ, Ulfert C, Möhlenbruch MA, Renieri L, Limbucci N, Parra-Fariñas C, Burkhardt JK, Kan P, Rinaldo L, Lanzino G, Brinjikji W, Spears J, Müller-Thies-Broussalis E, Killer-Oberpfalzer M, Islak C, Kocer N, Sonnberger M, Engelhorn T, Ghuman M, Yang VX, Salehani A, Harrigan MR, Radovanovic I, Pereira VM, Krings T, Matouk CC, Chen K, Aziz-Sultan MA, Ghorbani M, Schirmer CM, Goren O, Dalal SS, Koch MJ, Stapleton CJ, Patel AB, Finkenzeller T, Holtmannspötter M, Buhk JH, Foreman PM, Cress M, Hirschl R, Reith W, Simgen A, Janssen H, Marotta TR, Dmytriw AA. Comparison of PED and FRED flow diverters for posterior circulation aneurysms: a propensity score matched cohort study. J Neurointerv Surg. 2021;13:153–8.

    Article  PubMed  PubMed Central  Google Scholar 

Download references

Funding

No funds, grants, or other support were received.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Naci Kocer.

Ethics declarations

Conflict of interest

S.G. Kandemirli, F. Baltacioglu, J. Jesser and O. Kizilkilic declare that they have no competing interests. N. Kocer: consulting and proctoring agreement with Microvention. C. Islak: proctoring and training agreement with Microvention, NDI cofounder and shareholder, Neuravention INC shareholder. M. Möhlenbruch has received grants from Balt, Medtronic, MicroVention and Stryker outside the submitted work.

Ethical standards

This retrospective chart review study involving human participants was in accordance with the ethical standards of the institutional and national research committee and with the 1964 Helsinki Declaration and its later amendments or comparable ethical standards. The Institutional Review Board (IRB) of each participating center approved this study.

Additional information

The authors Markus Möhlenbruch and Naci Kocer contributed equally to the manuscript.

Availability of data and material: anonymized data are available upon request.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Kandemirli, S.G., Baltacioglu, F., Jesser, J. et al. Flow Redirection Endoluminal Device (FRED) with or without Adjunctive Coiling in Treatment of Very Large and Giant Cerebral Aneurysms. Clin Neuroradiol 32, 471–480 (2022). https://doi.org/10.1007/s00062-021-01061-x

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00062-021-01061-x

Keywords

Navigation