Zur Blutdruckmessung im Säuglingsalter

The determination of blood pressure in infancy

II. Der Einfluß verschiedener Meßverfahren auf den erhaltenen Blutdruckwert

  • Published:
Zeitschrift für Kinderheilkunde Aims and scope Submit manuscript

Summary

In 100 children without cardio-vascular abnormalities (aged between 4 hours and 9 months) simultaneous determinations of blood pressure were carried out using auscultatory and photoelectric methods as well as an electric method which is similar to electric impedance plethysmography. The results for, the systolic pressure using the electric method were almost identical to the intraarterial pressure recordings obtained by anglo-american authors by catheterisation of the umbilical artery. The values for blood pressure taken simultaneously with the auscultatory and photoelectric method were 3–4 mm Hg lower compared to the electric method. This is probably due to the high sensitivity of the electric method which records the exact first pulse wave that gets through as soon as the cuff pressure is lowered. According to our results all three methods give clinically reliable systolic pressure recordings. We prefer the photoelectric and electric method for use in newborns and infants for practical reasons since the Korotkoff-sounds are extremely difficult to hear. In comparison to the above mentioned methods the flush-method does not give exact results.

Zusammenfassung

Bei 100 kreislaufgesunden Kindern im Alter von 4 Std bis zu 9 Monaten wurden simultan Blutdruckmessungen nach dem auskultatorischen, photoelektrischen und elektrischen Leitfähigkeitsprinzip durchgeführt. Die mit dem elektrischen Verfahren erzielten systolischen Blutdruckwerte entsprechen weitgehend den intraarteriell durch Katheterisierung der A. umbilicalis gewonnenen Werten englischer und amerikanischer Autoren. Die bei den Simultanmessungen erhaltenen auskultatorischen und photoelektrischen Meßwerte lagen im Vergleich zu den elektrischen Blutdruckwerten durchschnittlich, 3–4 mm Hg tiefer. Diese Differenz dürfte darauf zurückzuführen sein, daß unser elektrisches Meßverfahren von allen die höchste Empfindlichkeit aufweist und damit am sichersten die erste durchschlagende Pulswelle bei sinkendem Manschettendruck erfaßt.

Nach unseren Ergebnissen liefern alle drei untersuchten Meßmethoden praktisch übereinstimmende systolische Blutdruckwerte. Aus erörterten Gründen wird vor allem die photoelektrische und elektrische Leitfähigkeits-meßmethode zur Blutdruckmessung im Säuglingsalter empfohlen.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

Literatur

  1. Ashworth, A. M., G. A. Nelligan, and J. R. Rogers: Sphygmomanometer for the newborn. Lancet 1959 I, 801.

  2. Boucke, A., u. K. Brecht: Ein neuer elektrischer Pulsschreiber und seine einfache Anwendung in der ärtzlichen Praxis. Dtsch. med. Wschr. 77, 562 (1952).

    Google Scholar 

  3. Brecht, K., u. A. Boucke: Neues elektrostatisches Tieftonmikrophon und seine Anwendung in der Sphygmographie. Pflügers Arch. ges. Physiol. 256, 43 (1952).

    Google Scholar 

  4. Celander, O., and G. Thunell: A plethysmographic method for measuring the systolic blood pressure in newborn infants. Acta paediat. (Uppsala) 49, 488 (1960).

    Google Scholar 

  5. ——: The “mercury-in-rubber” strain gauge for measurement of blood pressure and peripheral circulation in newborn infants. Acta paediat. (Uppsala) 50, 505 (1961).

    Google Scholar 

  6. Forfar, H. O., and M. A. Kibel: Blood pressure in the newborn estimated by the flush method. Arch. Dis. Childh. 31, 126 (1956).

    Google Scholar 

  7. Goodman, H. G., G. R. Cumming, and M. B. Raber: Photocell oscillometer for measuring systolic pressure in newborn. Amer. J. Dis. Child. 103, 152 (1962).

    Google Scholar 

  8. Holland, W. W., and I. M. Young: Neonatal blood pressure in relation to maturity, mode of delivery and condition at birth. Brit. med. J. 1956 II, 1331.

  9. Kaindl, F., K. Polzer u. F. Schuhfried: Rheographie. Eine Methode zur Beurteilung peripherer Gefäße. Darmstadt: Steinkopff 1959.

    Google Scholar 

  10. Keuth, U.: Sphygmographische Blutdruckmessung bei Säuglingen mit dem Infratonsystem. Z. Kinderheilk. 17, 107 (1955a).

    Google Scholar 

  11. —: Sphygmographische Blutdruckmessung bei Säuglingen mit dem Infratonsystem. Z. Kinderheilk. 77, 443 (1955b).

    Google Scholar 

  12. Lang, V. O.: Ein neues Gerät zur automatischen Registrierung von Blutdruck, peripherem Puls und EKG in der Pädiatrie (mit besonderer Berücksichtigung des Säuglingsalters). Mschr. Kinderheilk. 115, 105 (1967).

    Google Scholar 

  13. —: Untersuchungen über die Wirkung gebräuchlicher Depot-Kreislaufmittel im Säuglingsalter, Teil I. Mschr. Kinderheilk. 115, 371 (1967).

    Google Scholar 

  14. Levison, H., B. S. L. Kidd, P. A. Gemell, and P. R. Swyer: Blood pressure in normal full-term and premature infants. Amer. J. Dis. Child. 11, 374 (1966).

    Google Scholar 

  15. Morse, R. L., G. L. Brownell, and J. H., Currens: The blood pressure of newborn infants: Indirect determination by an automatic blood pressure recorder in 20 infants. Pediatrics 25, 50 (1960).

    Google Scholar 

  16. Moss, A. J., W. Liebling, and F. H. Adams: Flush method or determining blood pressure in infants: II. Normal values during first year of life. Pediatrics 21, 950 (1958).

    Google Scholar 

  17. Nyboer, J.: Electrical impedance plethysmography: Physical and physiological approach to peripheral vascular study. Circulation 2, 811 (1950).

    Google Scholar 

  18. Rice, H. V., and L. J. Posener.: A practical method for the measurement of systolic blood pressures of infants. Pediatrics 23, 854 (1959).

    Google Scholar 

  19. Recklinghausen, H. v.: Blutdruckmessung und Kreislauf in den Arterien des Menschen. Dresden u. Leipzig: Steinkopff 1940.

    Google Scholar 

  20. Robinow, M., W. F. Hamilton, R. A. Woodbury, and P. P. Volpitto: Accuracy of clinical determinations of blood pressure in children, with values under normal and abnormal conditions. Amer. J. Dis. Child. 58, 102 (1939).

    Google Scholar 

  21. Schaffer, A. I.: Neonatal blood pressure studies. Amer. J. Dis. Child. 89, 204 (1955).

    Google Scholar 

  22. Woodbury, R. A., M. Robinow, and W. R. Hamilton: Blood pressure studies on infants. Amer. J. Physiol. 122, 472 (1938).

    Google Scholar 

  23. Young, M., and D. Cottom: Arterial and venous blood pressure during a reducting in blood volume and hypoxia and hypercapnia in infants during the first two days of life. Pediatrics 37, 733 (1966).

    Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

Reprints and permissions

About this article

Cite this article

Lang, V.O., Fischer, T. Zur Blutdruckmessung im Säuglingsalter. Z. Kinder-Heilk. 105, 165–171 (1969). https://doi.org/10.1007/BF00438938

Download citation

  • Received:

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF00438938

Schlüsselwörter

Navigation