Abstract
A perfusion system was constructed which allows the fast application of different solutes underneath a water immersion objective. The perfusion system is mounted into the immersion objective by milling a slot into the frontal metal plate of the lens holder. It consists of a five-channel pipette fixed to the objective and solution reservoirs gated by computer controlled magnetic valves. Up to five different solutions can be applied to the specimen under study. The solution between objective and specimen is completely exchanged after 1–2 s as determined from fluorescence measurements. This arrangement is optimized for [Ca2+] measurements with a fluorescence measurement system in tissue slices, where upright microscopes are required. It offers the advantage of saving a micromanipulator for the perfusion pipette and facilitates a fast, reproducible and precise positioning of the perfusion system.
Similar content being viewed by others
References
Davies NW, Lux HD, Mrad M (1988) Site and mechanism of activation of proton-induced sodium current in chick dorsal root ganglion neurones. J Physiol (Lond) 400:159–187
Grynkiewicz G, Poenie M, Tsien RY (1985) A new generation of Ca2+ indicators with greatly improved fluorescence properties. J Biol Chem 260:3440
Kettenmann H, Backus KH, Schachner M (1984) Aspartate, glutamate and gamma-aminobutyric acid depolarize cultured astrocytes. Neurosci Lett 52:25–29
McCarthy KD, de Vellis J (1980) Preparation of separate astroglial and oligodendroglial cell cultures from rat cerebral tissue. J Cell Biol 85:890–902
Neher E (1989) Combined fura-2 and patch clamp measurements in rat peritoneal mast cells. In: Sellin LC, Libelius R, Thesleff S (eds) Neuromuscular junction. Elsevier (Biomedical Division), Amsterdam, pp 65–76
Schnitzer J, Schachner M (1981) Expression of Thy-1, H-2 and NS-4 cell surface antigens and tetanus toxin receptors in early postnatal and adult mouse cerebellum. J Immunol 1:429–456
Sommer I, Schachner M (1981) Monoclonal antibodies (O1 to O4) to oligodendrocyte surface — an immunological study in central nervous system. Dev Biol 83:311–327
Sontheimer GM, Sontheimer H, Ohlemeyer C, Blumenstiel S, Kettenmann H (1988) A universal data aquisition system for electrophysiology. In: Eslner N, Barth FG (eds) Sense organs, interfaces between environment and behaviour. Proceedings of the 16th Göttingen Neurobiology Conference. Thieme, Stuttgart New York
Stewart W (1981) Lucifer dyes — highly fluorescent dyes for biological tracing. Nature 292:17–21
Trotter J, Bitter-Suermann D, Schachner M (1989) Differentiation-regulated loss of polysialylated embryonic form and expression of the different polypeptides of the neural cell adhesion molecule (N-CAM) by cultured oligodendrocytes and myelin. J Neurosci Res 22:369–383
Tsien RY (1989) Fluorescent probes of cell signalling. Annu Rev Neurosci 12:227–253
Wienrich M, Kettenmann H (1989) Activation of Substance P receptors leads to membrane potential responses in cultured astrocytes. Glia 2:155–160
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Kirchhoff, F., Ohlemeyer, C. & Kettenmann, H. A fast perfusion system for single cell physiology optimized for microscopes with water immersion objectives. Pflugers Arch. 420, 573–577 (1992). https://doi.org/10.1007/BF00374635
Received:
Revised:
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/BF00374635