Nonlinear Geometric Warping of the Mask Image: A New Method for Reducing Misregistration Artifacts in Digital Subtraction Angiography

  • Published:
CardioVascular and Interventional Radiology Aims and scope Submit manuscript

Abstract

Purpose: Misregistration artifact is the major cause of image degradation in digital subtraction angiography (DSA). The purpose of this study was to evaluate the efficacy of a newly developed nonlinear geometric warping method to reduce misregistration artifact in DSA.

Methods: The processing of the images was carried out on a workstation with a fully automatic computerized program. After making differential images with a lapracian filter, 49 regions of interest (ROIs) were set in the image to be processed. Each ROI of the live image scanned the corresponding ROI of the mask image searching for the best position to match itself. Each pixel of the mask image was shifted individually following the data calculated from the shifts of the ROIs. Five radiologists compared the images produced by the conventional parallel shift technique and those processed with this new method in 16 series of cerebral DSA.

Results: In 14 of 16 series (88%), more radiologists judged the images processed with the new method to be better in quality. Small arteries near the skull base and veins of low density were clearly visualized in the images processed by the new method.

Conclusion: This newly proposed method could be a simple and practical way to automatically reduce misregistration artifacts in DSA.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

Author information

Authors and Affiliations

Authors

Rights and permissions

Reprints and permissions

About this article

Cite this article

Hayashi, N., Sakai, T., Kitagawa, M. et al. Nonlinear Geometric Warping of the Mask Image: A New Method for Reducing Misregistration Artifacts in Digital Subtraction Angiography. Cardiovasc Intervent Radiol 21, 138–141 (1998). https://doi.org/10.1007/s002709900230

Download citation

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s002709900230

Navigation