Skip to main content
Log in

Choosing a catheter system for measuring intra-oesophageal pressure

  • Published:
Medical and Biological Engineering and Computing Aims and scope Submit manuscript

Abstract

The choice of catheter systems for the measurement of oesophageal pressure in respiratory physiology is discussed and the merits and drawbacks of liquid- and air-filled systems described. Tests to determine the dynamic characteristics of various systems were performed and it was found that air-filled balloon catheters have a better performance in both phase and amplitude characteristics than most liquid-filled catheters. Tests were also made to find whether the balloons deteriorated significantly in controlled conditions and it was found that they had a useful working life.

Sommaire

Cet article discute du choix d'un système cathétérique pour la mesure de la pression oesophagienne pour la physiologie respiratoire et compare les mérites et inconvénients des systèmes à liquide et à air. Des expériences de détermination des caractéristiques dynamiques des différents matériels ont été effectuées. Elles ont permis de découvrir que les cathéters à ballon d'air ont de meilleurs performances en ce qui concerne les caractéristiques d'amplitude comme de phase que la plupart des cathéters à liquide. D'autres expériences ont été également menées pour déterminer si les ballons se dérioraient de manière importante dans des conditions contrôlées; on s'est aperçu qu'en fait ils avaient une longévité suffisamment grande.

Zusammenfassung

Die Auswahl von Kathethersystemen zur Messung des Speiseröhrendrucks in der Atmungsphysiologie wird besprochen, und die Vor- und Nachteile flüssigkeits- und luftgefüllter Systeme werden beschrieben. Zur Bestimmung der dynamischen Eigenschaften verschiedener Systeme wurden Tests durchgeführt, und es wurde festgestellt, daß luftgefüllte Ballonkatheter sowohl hinsichtlich Phasen- als auch Amplitudencharakteristika leistungsfähiger sind als die meisten flüssigkeitsgefüllten Katheter. Es wurden auch Tests durchgeführt um herauszufinden, ob sich die Ballons unter gesteuerten Bedingungen erheblich verschlechterten, und es wurde festgestellt, daß sie eine nützliche Lebensdauer haben.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  • Ahlstrom, H. (1974) Studies on pulmonary mechanics in infants.Acta. Paediat. Scand.244, (Suppl.), 1–21.

    Google Scholar 

  • Harrison, V. C., Heese, M. De V. andKlein, M. (1968) The significance of grunting in hyaline membrane disease.Pediatrics41, 549–559.

    Google Scholar 

  • Hjalmarson, O. andOlsson, T. (1974) Mechanical and ventilatory parameters in healthy and diseased newborn infants.Acta. Paediat. Scand.247, (Suppl.), 26–48.

    Google Scholar 

  • Karlberg, P., Cherry, R. B., Escardo, F. andKoch, G. (1960) Respiratory studies in newborn infants.49, 345–357.

    Google Scholar 

  • Krieger, I. (1963) Studies on mechanics of respiration in infancy.Amer. J. Dis. Child.105, 439–448.

    Google Scholar 

  • McDonald, D. A. (1974) Blood flow in arteries (2nd edn.). Edward Arnold, London.

    Google Scholar 

  • Miller, H. D. andBaker, L. E. (1973) A stable ultraminiature catheter-tip pressure transducer.Med. & Biol. Eng.1, 86–89.

    Google Scholar 

  • Pride, M. B. (1971) The assessment of airflow obstruction. Role of measurements of airways resistance and of tests of forced expiration.Brit. J. Dis. Chest.65, 135–169.

    Article  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

Reprints and permissions

About this article

Cite this article

Maxted, K.J., Shaw, A. & Macdonald, T.H. Choosing a catheter system for measuring intra-oesophageal pressure. Med. Biol. Eng. Comput. 15, 398–401 (1977). https://doi.org/10.1007/BF02457992