Abstract
Using the electrical impedance measurement technique to investigate stroke volume estimation, three models of the ventricle were simulated. A four-electrode impedance catheter was used; two electrodes to set up an electric field in the model and the other two to measure the potential difference. A new approach, itself an application of the quasi-static case of a method used to solve electromagnetic field problems, was used to solve the electric field in the model. The behaviour of the estimation is examined with respect to the electrode configuration on the catheter and to catheter location with respect to the ventricle walls. Cardiac stroke volume estimation was found to be robust to catheter location generating a 10 per cent error for an offset of 40 per cent of the catheter from the chamber axis and rotation of 20° with respect to the axis. The electrode configuration has a dominant effect on the sensitivity and accuracy of the estimation. Certain configurations gave high accuracy, whereas in others high sensitivity was found with lower accuracy. This led to the conclusion that the electrode configuration should be carefully chosen according to the desired criteria.
Similar content being viewed by others
References
Baan, J., Jong, T. A., Kerkhof, P. L. M., Moene, R. J., van Dijk, A. D., van der Velde, E. T. andKoops, J. (1981). Continuous stroke volume and cardiac output from intra-ventricular dimensions obtained with impedance catheter.Cardiovasc. Res.,15, 328–334.
Baan, J., van der Velde, E. T., Bruin, H. G., Smeenk, G. J., Koops, J., van Dijk, A. D., Temmerman, D., Senden, J. andBuis, B. (1984) Continuous measurement of left ventricular volume in animals and humans by conductance catheter.Circ.,70, 812–823.
Barak, C., Leviatan, Y. andInbar, G. F. (1989) Estimation of cardiac stroke volume by intracardiac electrical impedance measurement. 16th IEEE Conf. of Electrical & Electronics Engs. in Israel, p. 2.4.2.
Geddes, L. A., Hoff, H. E., Mello, A. andPalmer, C. (1966) Continuous measurement of ventricular stroke volume by electrical impedance.Cardiac Res. Cen. Bull.,4, 118–130.
Inbar, G. F., Heinze, R., Hoekstein, K. N. andLiess, H. D. (1988) Development of a closed-loop pacemaker controller regulating mixed venous oxygen saturation level.IEEE Trans.,BME-35, 679–690.
Leviatan, Y. andBoag, A. (1987) Analysis of scattering from dielectric cylinders using a multifilament magnetic current mode. —Ibid.,AP-35, 1119–1127.
Mur, G. andBaan, J. (1984) Computation of the input impedances of a catheter for cardiac volumetry. —Ibid.BME-31, 448–453.
Rushmer, R. F., Crystal, D. K., Wagner, C. andEllis, R. M. (1953) Intracardiac impedance plethsmography.Circ. Res.,174, 171–174.
Salo, R. W., Wallner, T. G. andPederson, B. D. (1986) Measurement of ventricular volume by intracardiac impedance: theoretical and empirical approaches.IEEE Trans.,BME-33, 189–195.
Spinelli, J. C. andValentinuzzi, M. E. (1986) Conductivity and geometrical factors affecting volume measurements with an impedancimetric catheter.Med. & Biol. Eng. & Comput. 24, 460–464.
Woodard, J. C., Bertram, C. D. andGow, B. S. (1989) Effect of radial position on volume measurements using the conductance catheter. —Ibid.,27, 25–32.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Barak, C., Leviatan, Y., Inbar, G.F. et al. Simulation method for cardiac stroke volume estimation by intracardiac electrical impedance measurement. Med. Biol. Eng. Comput. 30, 473–480 (1992). https://doi.org/10.1007/BF02457824
Received:
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/BF02457824