Abstract
An IBM compatible impedance cardiac output monitoring prototype system has been developed for use at the bedside on patients in the ICU, CCU, ER, Cath. Lab, and OR, etc. This impedance cardiography (ICG) system, whose operation is completely technician-free, provides a continuous display with digital results and four channel color waveforms on an Enhanced Graphics Display screen. The software is written in C language with several special segments in assembly code where speed is essential. In this prototype system, a real-time algorithm was introduced to modify the ensemble averaging technique so that it averages nonperiodic signals such as: ECG, dZ/dT, ΔZ, etc. Also, a real-time algorithm was developed to adaptively detect R spikes from conventional ECG signals. A signal preprocessor was developed to process signals digitally before any further work is done. This procedure reduces muscle noise, 60 Hz interference, and ventilatory movement. A special digital filter was designed to cope with the cases in which pacemakers are used. A special algorithm was also developed to further reduce the ventilation artifacts so that a period of apnea is unnecessary during the performance of the measurements. An anatomically specified electrode configuration has been defined enabling precise and reproducible positioning of the electrodes—hopefully leading to electrode standardization. At the present time, this prototype system has been compared with standard hand calculation and correlated with the clinical “gold standard,” the Swan-Ganz thermodilution cardiac output. Using 144 sets of data from 10 healthy volunteers, 4 critically ill patients, and 8 healthy exercising volunteers, calculations of cardiac output were made using our system and the standard hand calculation of stroke volume, based upon Kubicek's equation; there was a relatively high and stable correlation: r=0.93, p<0.005 (healthy); r=0.94, p<0.002 (ill); r=0.95, p<0.002 (exercise). From 20 patients at two different hospitals all with Swan-Ganz catheters in their hearts 65 correlation studies between our system and the standard thermodilution technique were performed; the results were encouraging in terms of accuracy and consistency (r1=0.84, p<0.01, n=10 CCU patients), and (r2=0.93, p<0.01, n=ICU, patients). These results along with a growing body of data from other investigators indicate that this noninvasive and technician-free system for measuring cardiac output could have a significant role in patient care.
Similar content being viewed by others
References
Berhe, R.M.; Levy, M.N. Physiology, Missouri: The C. V. Mosby Company; 1983.
Bogner, R.E.; Constanitinies, A.G. (eds). Introduction to digital filtering. London: Wiley; 1975.
Bonjer, F.H.; Van Den Berg, J.W.; Dirken, M.N.J. The origin of the variations of body impedance occurring during the cardiac cycle. Circulation. 6:415–420; 1952.
Demeter, R.J.; Toth, P.D.; Hawk, T.; Judy, W.V.; Tavel, M.E. The use of noninvasive impedance to determine cardiac output: factors affecting its accuracy. Am. J. Noninvas. Cardiol. 2:112–118; 1988.
Enghoff, E.; Lovheim, O. A comparison between the transthoracic electrical impedance method and the direct Fick and the dye dilution methods for cardiac output measurements in man. Scand. J. Clin. Lab. Invest. 39:585; 1979.
Gessert, W.L.; Reid, K.A.; Byboer, J. Reliability of tetrapolar electrical impedance plethysmography. biomed. Sci. Instrum. 5:143; 1969.
Gore, J.M.; Goldberg, R.J.; Spodick, D.H.; Alpert J.S.; Dalen, J.E. A community-wide assessment of the use of pulmonary artery catheters in patients with acute myocardial infarction. CHEST /92/4/ Oct. 721–727; 1987.
Goovaerts, H.G.; Ros, H.H.; Vanden Akker, T.J.; Schneider, H. A digital QRS detector based on the principle of contour limiting. IEEE Trans. Biomed. Eng.,BME-23:154; 1976.
Hill, R. V.; Jansen, J.C.; Fling, J.L. Electrical impedance plethysmography: a critical analysis. J. Appl. Physiol. 22:161; 1967.
Hill, R.V.; Jansen, J.C. Rebuttal. Med. Res. Eng. 8:7; 1969.
Hinnen, E. Further defense of impedance plethymography. Med. Res. Eng. 8:31; 1969.
Horton, J.W.; VanRavenswaay, A.C. Electrical impedance of the human body. J. Franklin Inst. 220–557; 1935.
Huelsman, L.P. Basic circuit theory with digital computations. Englewood Cliffs, NJ: Prentice-Hall; 1972.
Ito, H.; Yamakoshi, K.I.; Yamada, A. Physiological and fluid-dynamic investigations of the transthoracic impedance plethysmography method for measuring cardiac output: Part II-Analysis of the transthoracic impedance wave by perfusing dogs. Med. Biological Eng. 373–377, July; 1976.
Judy, W. V.; Hall, J.H.; Toth, P. D. Evaluation of impedance cardiography in critical care medicine. Cirt. Care Med. 12:270; 1984.
Jury, E.I. Theory and application of the Z-transform method. New York: Wiley 1964.
Kamakoshi, K.I.; Ito, H.; Yamada, A.; Miura, S.; Tomino, T. Physiological and fluid-dynamic investigations of the transthoracic impedance plethysmography method for measuring cardiac output: Part I—A fluid-dynamic approach to the theory using an expansible tube model. Med. Biological Eng. 365–372, July; 1976.
Keim, H.J.; Wallace, J.M.; Thurston, H.; Case, D.B.; Drayer, J.I.M.; Laragh, J.H. Impedance cardiography for determination of stroke index. J. Appl. Physiol. 41:797; 1976.
Kim, D.W.; Baker, L.E.; Pearce, J.A.; Kim, W.K. Origins of the impedance change in impedance cardiography by a three-dimsional finite element model. IEEE Trans. on Biomed. Eng. 35:Dec.; 1988.
Kinnen, E. A defense of electrical impedance plethysmography. Med. Res. Eng. 8:6; 1969.
Kinnen, E.; Kubicek, W.G.; Hill, P.; Turton, G. Thoracic cage impedance measurement; Tissue resistivity in vivo and transthoracic impedance at 10kHz. USAF Tech. Doc. Report, No. SAM-TDR-64-5. USAF School of Aerospace Medicine, Brook Air Force Base, Texas, March 1964.
Kizakevich, P.N.; Jochem, W.J.; Jones, A.W. A personal cardiac bioimpedance monitoring system. Proc. IEEE Eng. in Med. and Bio. Society, 10:48–50; 1988.
Kubicek, W.G.; Karnegis, J.N.; Patterson, R.P.; Witsoe, D.A.; Mattson, R.H.. Development and evaluation of an impedance cardiac output system. Aerosp. Med. 37:1208; 1966.
Kubicek, W.G Applications of the Minnesota Impedance Cardiograph. Development and evaluation of an impedance cardiographic system to measure cardiac output and other cardiac parameters. University of Minnesota College of Medical Sciences Lib. 1–15; 1970.
Lamberts, R.; Visser, K.R.; Zijlstra, W.G. Impedance cardiography. Van Gorcum, Assen, The Netherlands; 1984.
Lee, F.K.; Sun, H.H.; Van De Water, J.M. Impedance Cardiography-Coming of age. Proc. of 13th NE Bioeng. Conf. 418–421; 1987.
Lynn, P.A. Recursive digital filters for biological signals. Med. Biol. Eng. 9:37; 1971.
Lynn, P.A. Recursive digital filters with linear-phase characteristics. Comput. J. 15:337; 1972.
MacAdam, D.P. Development of a non-invasive means for monitoring stroke volume via impedance cardiography. M.S. Thesis, Worcester Polytechnic Institute; 1986.
Mohapatra, S.N. Non-invasive cardiovascular monitoring by electrical impedance technique. Pitman Medical Ltd. 226–244; 1981.
Muzi, M.; Jeutter, D.C. Computer-automated impedance-derived cardiac indexes. IEEE Trans. BME-33. No. 1.Jan.; 1986.
Myboer, J.; Bagno, S.; Barnett A.; Halsey, R.H. Radiocardiograms: electrical impedance changes of the heart in relation to electrocardiograms and heart sounds. J. Clin. Invest. 19:963; 1940.
Patterson, R.P.; Cardiac output determinations using impedance plethysmography. M.S. thesis. University of Minnesota Lib.; 1965.
Penny, B.C. Theory and cardiac applications of electrical impedance measurements. CRC reviews in Biomed. Eng. 13:227–281; 1986.
Penny, B.C. Patwardhan, N.A.; Wheeler, H.B. Simplified electrode array for impedance cardiography. Med. Biol. Eng. Comput. 23:1–7, Jan.; 1985.
Qu, M.H.; Zhang, Y.J.; Webster, J.G.; Tompkins, W.J. Motion artifact from spot and band electrodes during impedance cardiography. IEEE Trans. on BioMed. Eng.BME-33, No. 11, Nov.; 1986.
Reddy S.P.; Shyu, L.Y.; Hurwitz, B.E.; Nagel, J.H.; Schneiderman, N. Improved reliabilty of impedance cardiography by new signal processing techniques. Proc. IEEE Eng. Med. Bio. Society, 10:43–44; 1988.
Robin, E.D. Death by pulmonary artery flow-directed catheter (editorial)—Time for a moratorium? CHEST/92/4/Oct. 727–731; 1987.
Sakamoto, K.; Muto, K.; Kanai, H.; Lizuka M. Problems of impedance cardiography. Med. Biol. Eng. Comput. 17:697–709; 1979.
Secher, N.J.; Arnsbo, P.; Anderson, H.L.; Thomsen, A. Measurements of cardiac stroke volume in various body positions in pregnancy and during caesarean section: a comparison between thermodilution and impedance cardiography. Scand. J. Clin. Lab. Invest. 39:569; 1979.
Shyu, L.Y.; Reddy, S.P.; Nagel, J.H.; Schneiderman, N. New signal processing techniques for improved reliability of impedance cardiography. Proc. IEEE Eng. in Med. and Biol. Society. 10:41–42; 1988.
Seigel, L.C.; Shafer, S.L.; Martinez, G.M.; Ream, A.K.; Scott, J.C. Simultaneous measurements of cardiac output by thermodilution, Esophageal Doppler, and electrical impedance in anesthetized patients. J. Cardiothoracic Anesthesia 2:590–595; 1988.
Sramek, B.B. Noninvasive technique for measurement of cardiac output by means of electrical impedance. Proc. 5th Int. Conf. Electr. Bio-impedance, Tokyo, 39; 1981.
Thakor, N.V.; Webster, J.G.; Tompkins W.J. Optimal QRS detector. Med. Biol. Eng. Comput. 21: 343–350; 1983.
Van De Water, J.M.; Miller, I.T.; Milne, E.N.C.; Hanson, E.L.; Sheldon, G.F.; Kagey, K.S. Impedance plethysmography: a noninvasive means of monitoring the thoracic surgery patients. J. Thorac. Cardiovasc. Surg. 60:641; 1970.
Visser K.R. Origin of the impedance cardiogram. Proc. IEEE Eng. in Med. and Bio. Society, 10: 763–765; 1988.
Wang, X., Adamson, D.; Sun H.H.; Van De Water, J.M. Impedance cardiac profile monitoring by a modified ensemble averaging technique. Proc. IEEE Eng. in Med. and Bio. Society, 10:39–40; 1988.
Zhang, Y.J.; Qu, M.H.; Webster, J.G.; Tompkins, W.J.; Ward, B.A.; Bassett Jr. D.R. Cardiac output monitoring by impedance cardiography during treadmill exercise. IEEE Trans. on BioMed. Eng.,BME-33 No. 11, Nov.; 1986.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Wang, X., Sun, H.H., Adamson, D. et al. An impedance cardiography system: A new design. Ann Biomed Eng 17, 535–556 (1989). https://doi.org/10.1007/BF02368072
Received:
Revised:
Issue Date:
DOI: https://doi.org/10.1007/BF02368072