An impedance cardiography system: A new design

  • Published:
Annals of Biomedical Engineering Aims and scope Submit manuscript

Abstract

An IBM compatible impedance cardiac output monitoring prototype system has been developed for use at the bedside on patients in the ICU, CCU, ER, Cath. Lab, and OR, etc. This impedance cardiography (ICG) system, whose operation is completely technician-free, provides a continuous display with digital results and four channel color waveforms on an Enhanced Graphics Display screen. The software is written in C language with several special segments in assembly code where speed is essential. In this prototype system, a real-time algorithm was introduced to modify the ensemble averaging technique so that it averages nonperiodic signals such as: ECG, dZ/dT, ΔZ, etc. Also, a real-time algorithm was developed to adaptively detect R spikes from conventional ECG signals. A signal preprocessor was developed to process signals digitally before any further work is done. This procedure reduces muscle noise, 60 Hz interference, and ventilatory movement. A special digital filter was designed to cope with the cases in which pacemakers are used. A special algorithm was also developed to further reduce the ventilation artifacts so that a period of apnea is unnecessary during the performance of the measurements. An anatomically specified electrode configuration has been defined enabling precise and reproducible positioning of the electrodes—hopefully leading to electrode standardization. At the present time, this prototype system has been compared with standard hand calculation and correlated with the clinical “gold standard,” the Swan-Ganz thermodilution cardiac output. Using 144 sets of data from 10 healthy volunteers, 4 critically ill patients, and 8 healthy exercising volunteers, calculations of cardiac output were made using our system and the standard hand calculation of stroke volume, based upon Kubicek's equation; there was a relatively high and stable correlation: r=0.93, p<0.005 (healthy); r=0.94, p<0.002 (ill); r=0.95, p<0.002 (exercise). From 20 patients at two different hospitals all with Swan-Ganz catheters in their hearts 65 correlation studies between our system and the standard thermodilution technique were performed; the results were encouraging in terms of accuracy and consistency (r1=0.84, p<0.01, n=10 CCU patients), and (r2=0.93, p<0.01, n=ICU, patients). These results along with a growing body of data from other investigators indicate that this noninvasive and technician-free system for measuring cardiac output could have a significant role in patient care.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  1. Berhe, R.M.; Levy, M.N. Physiology, Missouri: The C. V. Mosby Company; 1983.

    Google Scholar 

  2. Bogner, R.E.; Constanitinies, A.G. (eds). Introduction to digital filtering. London: Wiley; 1975.

    Google Scholar 

  3. Bonjer, F.H.; Van Den Berg, J.W.; Dirken, M.N.J. The origin of the variations of body impedance occurring during the cardiac cycle. Circulation. 6:415–420; 1952.

    CAS  PubMed  Google Scholar 

  4. Demeter, R.J.; Toth, P.D.; Hawk, T.; Judy, W.V.; Tavel, M.E. The use of noninvasive impedance to determine cardiac output: factors affecting its accuracy. Am. J. Noninvas. Cardiol. 2:112–118; 1988.

    Google Scholar 

  5. Enghoff, E.; Lovheim, O. A comparison between the transthoracic electrical impedance method and the direct Fick and the dye dilution methods for cardiac output measurements in man. Scand. J. Clin. Lab. Invest. 39:585; 1979.

    CAS  PubMed  Google Scholar 

  6. Gessert, W.L.; Reid, K.A.; Byboer, J. Reliability of tetrapolar electrical impedance plethysmography. biomed. Sci. Instrum. 5:143; 1969.

    Google Scholar 

  7. Gore, J.M.; Goldberg, R.J.; Spodick, D.H.; Alpert J.S.; Dalen, J.E. A community-wide assessment of the use of pulmonary artery catheters in patients with acute myocardial infarction. CHEST /92/4/ Oct. 721–727; 1987.

  8. Goovaerts, H.G.; Ros, H.H.; Vanden Akker, T.J.; Schneider, H. A digital QRS detector based on the principle of contour limiting. IEEE Trans. Biomed. Eng.,BME-23:154; 1976.

    CAS  PubMed  Google Scholar 

  9. Hill, R. V.; Jansen, J.C.; Fling, J.L. Electrical impedance plethysmography: a critical analysis. J. Appl. Physiol. 22:161; 1967.

    CAS  PubMed  Google Scholar 

  10. Hill, R.V.; Jansen, J.C. Rebuttal. Med. Res. Eng. 8:7; 1969.

    Google Scholar 

  11. Hinnen, E. Further defense of impedance plethymography. Med. Res. Eng. 8:31; 1969.

    Google Scholar 

  12. Horton, J.W.; VanRavenswaay, A.C. Electrical impedance of the human body. J. Franklin Inst. 220–557; 1935.

  13. Huelsman, L.P. Basic circuit theory with digital computations. Englewood Cliffs, NJ: Prentice-Hall; 1972.

    Google Scholar 

  14. Ito, H.; Yamakoshi, K.I.; Yamada, A. Physiological and fluid-dynamic investigations of the transthoracic impedance plethysmography method for measuring cardiac output: Part II-Analysis of the transthoracic impedance wave by perfusing dogs. Med. Biological Eng. 373–377, July; 1976.

  15. Judy, W. V.; Hall, J.H.; Toth, P. D. Evaluation of impedance cardiography in critical care medicine. Cirt. Care Med. 12:270; 1984.

    Google Scholar 

  16. Jury, E.I. Theory and application of the Z-transform method. New York: Wiley 1964.

    Google Scholar 

  17. Kamakoshi, K.I.; Ito, H.; Yamada, A.; Miura, S.; Tomino, T. Physiological and fluid-dynamic investigations of the transthoracic impedance plethysmography method for measuring cardiac output: Part I—A fluid-dynamic approach to the theory using an expansible tube model. Med. Biological Eng. 365–372, July; 1976.

  18. Keim, H.J.; Wallace, J.M.; Thurston, H.; Case, D.B.; Drayer, J.I.M.; Laragh, J.H. Impedance cardiography for determination of stroke index. J. Appl. Physiol. 41:797; 1976.

    CAS  PubMed  Google Scholar 

  19. Kim, D.W.; Baker, L.E.; Pearce, J.A.; Kim, W.K. Origins of the impedance change in impedance cardiography by a three-dimsional finite element model. IEEE Trans. on Biomed. Eng. 35:Dec.; 1988.

  20. Kinnen, E. A defense of electrical impedance plethysmography. Med. Res. Eng. 8:6; 1969.

    CAS  PubMed  Google Scholar 

  21. Kinnen, E.; Kubicek, W.G.; Hill, P.; Turton, G. Thoracic cage impedance measurement; Tissue resistivity in vivo and transthoracic impedance at 10kHz. USAF Tech. Doc. Report, No. SAM-TDR-64-5. USAF School of Aerospace Medicine, Brook Air Force Base, Texas, March 1964.

  22. Kizakevich, P.N.; Jochem, W.J.; Jones, A.W. A personal cardiac bioimpedance monitoring system. Proc. IEEE Eng. in Med. and Bio. Society, 10:48–50; 1988.

    Google Scholar 

  23. Kubicek, W.G.; Karnegis, J.N.; Patterson, R.P.; Witsoe, D.A.; Mattson, R.H.. Development and evaluation of an impedance cardiac output system. Aerosp. Med. 37:1208; 1966.

    CAS  PubMed  Google Scholar 

  24. Kubicek, W.G Applications of the Minnesota Impedance Cardiograph. Development and evaluation of an impedance cardiographic system to measure cardiac output and other cardiac parameters. University of Minnesota College of Medical Sciences Lib. 1–15; 1970.

  25. Lamberts, R.; Visser, K.R.; Zijlstra, W.G. Impedance cardiography. Van Gorcum, Assen, The Netherlands; 1984.

    Google Scholar 

  26. Lee, F.K.; Sun, H.H.; Van De Water, J.M. Impedance Cardiography-Coming of age. Proc. of 13th NE Bioeng. Conf. 418–421; 1987.

  27. Lynn, P.A. Recursive digital filters for biological signals. Med. Biol. Eng. 9:37; 1971.

    CAS  PubMed  Google Scholar 

  28. Lynn, P.A. Recursive digital filters with linear-phase characteristics. Comput. J. 15:337; 1972.

    Article  Google Scholar 

  29. MacAdam, D.P. Development of a non-invasive means for monitoring stroke volume via impedance cardiography. M.S. Thesis, Worcester Polytechnic Institute; 1986.

  30. Mohapatra, S.N. Non-invasive cardiovascular monitoring by electrical impedance technique. Pitman Medical Ltd. 226–244; 1981.

  31. Muzi, M.; Jeutter, D.C. Computer-automated impedance-derived cardiac indexes. IEEE Trans. BME-33. No. 1.Jan.; 1986.

  32. Myboer, J.; Bagno, S.; Barnett A.; Halsey, R.H. Radiocardiograms: electrical impedance changes of the heart in relation to electrocardiograms and heart sounds. J. Clin. Invest. 19:963; 1940.

    Google Scholar 

  33. Patterson, R.P.; Cardiac output determinations using impedance plethysmography. M.S. thesis. University of Minnesota Lib.; 1965.

  34. Penny, B.C. Theory and cardiac applications of electrical impedance measurements. CRC reviews in Biomed. Eng. 13:227–281; 1986.

    Google Scholar 

  35. Penny, B.C. Patwardhan, N.A.; Wheeler, H.B. Simplified electrode array for impedance cardiography. Med. Biol. Eng. Comput. 23:1–7, Jan.; 1985.

    Google Scholar 

  36. Qu, M.H.; Zhang, Y.J.; Webster, J.G.; Tompkins, W.J. Motion artifact from spot and band electrodes during impedance cardiography. IEEE Trans. on BioMed. Eng.BME-33, No. 11, Nov.; 1986.

  37. Reddy S.P.; Shyu, L.Y.; Hurwitz, B.E.; Nagel, J.H.; Schneiderman, N. Improved reliabilty of impedance cardiography by new signal processing techniques. Proc. IEEE Eng. Med. Bio. Society, 10:43–44; 1988.

    Google Scholar 

  38. Robin, E.D. Death by pulmonary artery flow-directed catheter (editorial)—Time for a moratorium? CHEST/92/4/Oct. 727–731; 1987.

  39. Sakamoto, K.; Muto, K.; Kanai, H.; Lizuka M. Problems of impedance cardiography. Med. Biol. Eng. Comput. 17:697–709; 1979.

    CAS  PubMed  Google Scholar 

  40. Secher, N.J.; Arnsbo, P.; Anderson, H.L.; Thomsen, A. Measurements of cardiac stroke volume in various body positions in pregnancy and during caesarean section: a comparison between thermodilution and impedance cardiography. Scand. J. Clin. Lab. Invest. 39:569; 1979.

    CAS  PubMed  Google Scholar 

  41. Shyu, L.Y.; Reddy, S.P.; Nagel, J.H.; Schneiderman, N. New signal processing techniques for improved reliability of impedance cardiography. Proc. IEEE Eng. in Med. and Biol. Society. 10:41–42; 1988.

    Google Scholar 

  42. Seigel, L.C.; Shafer, S.L.; Martinez, G.M.; Ream, A.K.; Scott, J.C. Simultaneous measurements of cardiac output by thermodilution, Esophageal Doppler, and electrical impedance in anesthetized patients. J. Cardiothoracic Anesthesia 2:590–595; 1988.

    Google Scholar 

  43. Sramek, B.B. Noninvasive technique for measurement of cardiac output by means of electrical impedance. Proc. 5th Int. Conf. Electr. Bio-impedance, Tokyo, 39; 1981.

  44. Thakor, N.V.; Webster, J.G.; Tompkins W.J. Optimal QRS detector. Med. Biol. Eng. Comput. 21: 343–350; 1983.

    CAS  PubMed  Google Scholar 

  45. Van De Water, J.M.; Miller, I.T.; Milne, E.N.C.; Hanson, E.L.; Sheldon, G.F.; Kagey, K.S. Impedance plethysmography: a noninvasive means of monitoring the thoracic surgery patients. J. Thorac. Cardiovasc. Surg. 60:641; 1970.

    PubMed  Google Scholar 

  46. Visser K.R. Origin of the impedance cardiogram. Proc. IEEE Eng. in Med. and Bio. Society, 10: 763–765; 1988.

    Google Scholar 

  47. Wang, X., Adamson, D.; Sun H.H.; Van De Water, J.M. Impedance cardiac profile monitoring by a modified ensemble averaging technique. Proc. IEEE Eng. in Med. and Bio. Society, 10:39–40; 1988.

    CAS  Google Scholar 

  48. Zhang, Y.J.; Qu, M.H.; Webster, J.G.; Tompkins, W.J.; Ward, B.A.; Bassett Jr. D.R. Cardiac output monitoring by impedance cardiography during treadmill exercise. IEEE Trans. on BioMed. Eng.,BME-33 No. 11, Nov.; 1986.

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

Reprints and permissions

About this article

Cite this article

Wang, X., Sun, H.H., Adamson, D. et al. An impedance cardiography system: A new design. Ann Biomed Eng 17, 535–556 (1989). https://doi.org/10.1007/BF02368072

Download citation

  • Received:

  • Revised:

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF02368072

Keywords

Navigation