Summary
The development of a computer-aided rapid-scanning detector for HPLC based on the linear photodiode array UV-visible spectrophotometer is described. Chomatograms monitored at up to three wavelengths, with automatic capture of spectra for eluted peaks and post-run processing of spectral data to generate log10 (A) spectra, second derivative and fourth derivative spectra, are described. Methods are reported for the analysis of forensic samples of diacetylmorphine (heroin) in the presence of the degradation products and potential contaminants caffeine, papaverine, 6-acetylcodeine, thebaine, 6-acetyl-morphine, procaine and morphine separated by HPLC. The novel use of second and higher derivative spectra for the qualitative characterisation of drugs and contaminants separated by HPLC is described.
Similar content being viewed by others
References
Y. Talmi, Ed., “Multichannel Image Detectors”, ACS Symposium Series, No. 102, American Chemical Society, Wash., D.C., 1979.
M. J. Milano, S. Lam, E. Grushka, J. Chromatogr.125, 315 (1976).
H. L. Pardue, Top. Autom. Chem. Anal.1, 163 (1979).
A. E. McDowell, H. L. Pardue, Anal. Chem.49, 1171 (1977).
L. N. Klatt, J. Chromatogr. Sci.17, 225 (1979).
A. F. Fell, Anal. Proc.17, 512 (1980).
K. L. Ratzlaff, S. L. Paul, Appl. Spectrosc.33, 240 (1979).
G. Horlick, Appl. Spectrosc.30, 113 (1976).
R. M. Hoffman, H. L. Pardue, Anal. Chem.51, 1267 (1979).
I. M. Warner, J. B. Callis, et al., Anal. Lett.8, 665 (1975).
L. W. Hershberger, J. B. Callis, G. D. Christian, Anal. Chem.53, 971 (1981).
M. P. Fogarty, D. C. Shelly, I. M. Warner, J. High Resolut. Chromatogr. Chromatogr. Commun.4, 561 (1981).
D. C. Shelly, M. P. Fogarty, I. M. Warner, ibid4, 616 (1981).
B. G. Willis, Application Note No. 8450-5, Hewlett-Packard Inc. Palo Alto, 1979.
M. J. Milano, S. Lam, et al. J. Chromatogr.149, 599 (1977).
D. T. Zelt, J. A. Owen, G. S. Marks, J. Chromatogr.189, 209 (1980).
A. F. Fell, UV Spectrom. Group Bull.8, 5 (1980).
R. N. Smith, M. Zetlein, J. Chromatogr.130, 314 (1977).
S. K. Soni, S. M. Dugar, J. Forens. Sci.24, 437 (1979).
G. E. James, Can. Res.13, (8), 39 (1980).
R. E. Dessy, W. D. Reynolds, et al., J. Chromatogr.126, 347 (1976).
D. W. Johnson, J. W. Gunn, J. Forensic Sci.17, 629 (1972).
W. L. Butler, D. W. Hopkins, Photochem. Photobiol.12, 439 (1970).
G. Talsky, L. Mayring, H. Kreuzer, Angew. Chem. Int. Ed. Engl.17, 785 (1978).
P. B. Baker, T. A. Gough, J. Chromatogr. Sci.19, 483 (1981).
A. F. Fell, Proc. Anal. Div. Chem. Soc.15, 260 (1978).
T. C. O'Haver, Clin. Chem. (Winston-Salem, N.C.)22, 1548 (1979).
R. Gill, T. S. Bal, A. C. Moffat, J. Forensic Sci. Soc.22, 165 (1982).
L. Fishbein, “Chromatography of Environmental Hazards, Vol. 4: Drugs of Abuse”, Elsevier Scientific, Amsterdam and New York, 1982, p. 209.
A. F. Fell, H. P. Scott, R. Gill, A. C. Moffat, Anal. Proc. (in press).
L. Kryger, Talanta28, 871 (1981).