A comparison of measurements of blood pressure, heart-rate and oxygenation during inter-hospital transport of the critically ill

  • Originals
  • Published:
Intensive Care Medicine Aims and scope Submit manuscript

Abstract

Critically ill patients transferred between hospitals are often inadequately monitored in transit, with outcome adversely affected. In 22 such patients, we compared direct and palpated measurements of systolic pressure, oscilloscopic and aneroid manometric measurements of mean pressure, ECG and palpated measurements of heart-rate and clinical and oximetric assessments of oxygenation. On average, palpated readings of systolic pressure under-read direct readings by 29% and palpated readings of heart-rate under-read ECG readings by 2%. The mean difference between oscilloscopic and manometric readings of mean pressure was zero. Oxygen saturation readings did not reach a level which allowed valid comparison. If direct measurement of heart-rate and blood pressure by battery-powered monitors is not feasible, palpation of heart-rate and manometric measurement of mean arterial pressure are acceptable alternatives during secondary transport of the critically ill.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  1. Wright IH, McDonald JC, Rogers PN, Ledingham IMcA (1988) Provision of facilities for secondary transport of seriously ill patients in the United Kingdom. Br Med J 296:543–545

    Google Scholar 

  2. Gentleman D, Jennett B (1981) Hazards of inter-hospital transfer of comatose head-injured patients. Lancet II:853–855

    Google Scholar 

  3. Bion JF, Wilson IH, Taylor PA (1988) Transporting critically ill patients by ambulance: audit by sickness scoring. Br Med J 296:170

    Google Scholar 

  4. Zorab JSM (1969) Continous display of the arterial pressure: a simple manometric technique. Anaesthesia 24:431–437

    Google Scholar 

  5. Allan MWB, Gray WM, Asbury AJ (1988) Measurement of arterial pressure using catheter-transducer systems. Br J Anaesth 60:413–418

    Google Scholar 

  6. Bland JM, Altman DG (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. I:307–310

    Google Scholar 

  7. Taylor MB, Whitwam JG (1988) The accuracy of pulse oximeters. A comparative clinical evaluation of five pulse oximeters. Anaesthesia 43:229–232

    Google Scholar 

  8. Konarzewski WH, Banerjee A (1988) Transporting critically ill patients by ambulance. Br Med J 296:645 (letter)

    Google Scholar 

  9. Waddell G, Stuart B, Tehrani MA, McGarrity G, Reyes A, Smith HC, Ledingham IMcA (1975) Intra-arterial monitoring of critically ill patients in ambulances. Br Med J 4:206

    Google Scholar 

  10. Gardner RM (1981) Direct blood pressure measurement—dynamic response requirements. Anesthesiology 54:227–236

    Google Scholar 

  11. Goldhill DR (1986) Bilateral simultaneous indirect systolic blood pressure measurements. Cardiovasc Res 20:774–777

    Google Scholar 

  12. Snook R (1972) Medical aspects of ambulance design. Br Med J 3:574–578

    Google Scholar 

  13. Cohn JN (1967) Blood pressure measurement in shock: mechanism of inaccuracy in auscultatory and palpatory methods. JAMA 199:118–122

    Google Scholar 

  14. O'Rourke MF, Blazek JV, Moreels CL, Krovetz LJ (1955) Pressure wave transmission along the human aorta—changes with age and in arterial degenerative disease. Circ Res 23:567–579

    Google Scholar 

  15. Gravlee GP, Wong AB, Adkins TG, Case LD, Pauca AL (1989) A comparison of radial, brachial and aortic pressures after cardiopulmonary bypass. J Cardiothorac Anesth 3:20–26

    Google Scholar 

  16. Abou-Madi M, Lenis S, Archer D, Ravussin P, Trop D (1986) Comparison of direct blood pressure measurements at the radial and dorsalis pedis arteries during sodium nitroprusside- and isoflurane-induced hypotension. Anesthesiology 65:692–695

    Google Scholar 

  17. Slogoff S, Keats AS, Arlund C (1983) On the safety of radial artery cannulation. Anesthesiology 59:42–47

    Google Scholar 

  18. Severinghaus J (1957) Aneroid manometer for arterial blood pressure. Anesthesiology 24:872

    Google Scholar 

  19. Dawson ADG, Babington PCB (1987) an intensive care trolley—an economical and viable alternative to the MICU. Anaesth Intensive Care 15:229–233

    Google Scholar 

  20. Hickey DD (1986) A simple device for the direct measurement of mean arterial pressure and for calibration of arterial pressure monitors. J Med Eng Technol 10:188–192

    Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

Reprints and permissions

About this article

Cite this article

Runcie, C.J., Reeve, W., Reidy, J. et al. A comparison of measurements of blood pressure, heart-rate and oxygenation during inter-hospital transport of the critically ill. Intensive Care Med 16, 317–322 (1990). https://doi.org/10.1007/BF01706357

Download citation

  • Received:

  • Accepted:

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF01706357

Key words

Navigation