Metallic implants produce a great deal of artifacts in standard CT images. These artifacts cover tissue structures and impede visualization of the anchorage of the implant within the bone. Thus a modified CT procedure has been developed to permit reconstructions free from artifacts. It is an iterative procedure using a maximum of measured projection data. Using this modified CT technique, the anchorage of a titanium-alloy femoral prosthesis implantet without cement was examined. Good contact of the implant with the cortical bone has documented. This promises good initial stability. The CT examinations were performed by using a special-purpose CT scanner. It is hoped that the modified CT technique will allow a quantification of bony tissue in the close vicinity of implants and thus provide a non-invasive procedure that will allow the long-term stability of implants to be evaluated.
Zusammenfassung
Metallische Implantate erzeugen im computertomographischen Querschnittsbild Artefakte. Diese überdecken Gewebestrukturen und verunmöglichen eine Beurteilung der Verankerung eines Implantates im Knochen. Es wurde daher ein neues Verfahren entwickelt, bei dem schrittweise der verfälschende Einfluß des Implantates auf die computertomographische Rekonstruktion reduziert wird. Dieses Verfahren wurde auf einen exzidierten Femur mit zementlos implantierter Schaftprothese aus einer Titanlegierung angewandt. Von der Implantatspitze bis zum Gelenkbeginn konnten damit Querschnitte von guter Qualität hergestellt werden. Mit dem neuen Verfahren sollte ein nicht-invasives Prozedere zur Verfügung stehen, das erlaubt, eine Quantifizierung des Knochengewebes in unmittelbarer Nähe des Implantates durchzuführen und die Langzeitstabilität einer Implantatsverankerung zu objektivieren.
Similar content being viewed by others
References
Probst KJ (1980) Stereo-Röntgen-Analyse (SRA) von Lockerungsvorgängen alloplastischer Hüftgelenkimplantate. Enke, Stuttgart
Lewitt RM, Bates RHT (1978) Image reconstructions from projections. Optik 50:189–204
Image processing for 2-D and 3-D reconstruction from projections. A digest of technical papers presented at the Topical Meeting on Image Processing from Projections: Theory and Practice in Medicine and the Physical Sciences. August 4–7, 1975. Stanford University, Stanford, California
Oppenheim BE (1977) Reconstruction tomography from incomplete projections. In: Reconstruction tomography in diagnostic radiology and nuclear medicine. University Park Press, pp 155–185
Hinderling T, Rüegsegger P, Anliker M, Dietschi C (1979) Computed tomography from hollow projections: an application to in vivo evaluation of artificial hip joints. J Comput Tomogr 3:52–57
Hinderling T (1978) Computertomographisches Verfahren zur Untersuchung der Gewebe- und Zementverteilung in der Umgebung metallischer Implantate. Dissertation Universität Zürich
Semlitsch M, Willert H-G (1980) Properties of implant alloys for artificial hip joints. Med Biol Engin Comput 18:511–520
Seitz P, Rüegsegger P (1982) Bone densitometry in the vicinity of metallic implants. J Comput Assist Tomogr 6:200
Semlitsch M, Willert H-G (1981) Biomaterialien für Implantate in der orthopädischen Chirurgie. Medizintechnik 3:66–72
Stebler B, Steiger P, Rüegsegger P (1982) A low dose X-ray CT-system for bone densitometry. J Comput Assist Tomogr 6:201
Genant HK, Boyd D (1977) Quantitative bone mineral analysis using dual energy computed tomography. Invest Radiol 6:545–551
Joseph PM, Spital RD (1981) The exponential Edge-Gradient effect in X-ray computed tomography. Phys Med Biol 3:473–487
Zweymüller K, Semlitsch M (1982) Concept and material properties of a cementless hip prosthesis system with Al2 O3 ceramic ball heads and wrought Ti-6A1-4V stems. Arch Orthop Trauma Surg 100:
Author information
Authors and Affiliations
Additional information
This study was supported partly by the Swiss National Science Foundation (3.893.-0.79) and the Medical Engineering Division of Sulzer Brothers Limited, Winterthur
Rights and permissions
About this article
Cite this article
Seitz, P., Rüegsegger, P. Anchorage of femoral implants visualized by modified computed tomography. Arch. Orth. Traum. Surg. 100, 261–266 (1982). https://doi.org/10.1007/BF00381666
Received:
Issue Date:
DOI: https://doi.org/10.1007/BF00381666