Abstract
An implantable thermal flow sen treatment of hydrocephalus has been developed.The sensor uses passive telemetry at 13.56 MHz for power supply and read out of the measured flow rate. The in vitro performance of the sensor has been characterized using artificial Cerebrospinal Fluid (CSF) with increased protein concentration and artificial CSF with 10% fresh blood. No drift could be observed in the flow sensor measurement which could be associated to a deposition of proteins at the sensitive surface walls of the packaged flow sensor.
Access this chapter
Tax calculation will be finalised at checkout
Purchases are for personal use only
Preview
Unable to display preview. Download preview PDF.
Similar content being viewed by others
Author information
Authors and Affiliations
Editor information
Editors and Affiliations
Rights and permissions
Copyright information
© 2009 International Federation for Medical and Biological Engineering
About this paper
Cite this paper
Burger, J. et al. (2009). In-vitro characterization of an implantable thermal flow sensor for hydrocephalus. In: Dössel, O., Schlegel, W.C. (eds) World Congress on Medical Physics and Biomedical Engineering, September 7 - 12, 2009, Munich, Germany. IFMBE Proceedings, vol 25/8. Springer, Berlin, Heidelberg. https://doi.org/10.1007/978-3-642-03887-7_77
Download citation
DOI: https://doi.org/10.1007/978-3-642-03887-7_77
Publisher Name: Springer, Berlin, Heidelberg
Print ISBN: 978-3-642-03886-0
Online ISBN: 978-3-642-03887-7
eBook Packages: EngineeringEngineering (R0)