Skip to main content

Development of a Submillimeter Portable Gamma-Ray Imaging Detector, Based on a GAGG:Ce—Silicon Photomultiplier Array

  • Conference paper
  • First Online:
Engineering of Scintillation Materials and Radiation Technologies (ISMART 2018)

Part of the book series: Springer Proceedings in Physics ((SPPHY,volume 227))

  • 681 Accesses

Abstract

In this study we present the development of a gamma-ray detector based on 1 mm pixel Gd3Al2Ga3O12:Ce (GAGG:Ce) scintillators and a silicon photomultiplier array (ArraySL-4) for possible applications in medical imaging detectors (with focus in PET applications) as well as in personal gamma radiation monitoring applications (with focus on 137Cs radioisotope recognition). A 12 × 12 matrix of pixellated scintillators is coupled to a 4 × 4 pixel elements of a SIPM array covering an active area of 13.4 mm2. Experimental evaluation was carried out with 22Na and 137Cs radioactive sources and the parameters studied were intrinsic spatial resolution, energy resolution and peak-to-valley ratio. ArraySL-4 is a commercially available, 4 × 4 array detector covering an active area of 13.4 mm2. The GAGG:Ce scintillator array used in this study has 1 × 1 × 10 mm3 pixel size elements with 0.1 mm thickness white BaSO4 reflector material between the crystals. A symmetric resistive charge division matrix was applied reducing array’s 16 outputs to 4 position signals. A Field Programmable Gate Array (FPGA) Spartan 6 LX150T was used for triggering and signal processing of the signal pulses digitized using free running Analog to Digital Converters. Raw images and horizontal profiles of the 12 × 12 GAGG:Ce scintillator array produced under 511 keV and 662 keV excitation are illustrated. Moreover, the energy spectra obtained with 22Na and 137Cs radioactive sources are shown. All GAGG:Ce central pixels were separated in the 2-dimensional position histograms providing submillimeter intrinsic spatial resolution with an average peak-to-valley (P/V) ratio of 1.75 for 511 keV and P/V ratio 1.85 for 662 keV. The energy resolution was 16.9% for 511 keV and 14% for 662 keV. Those encouraging results, prove that this system could build up to a compact miniature 137Cs spectrometer (with 2D isotope mapping capability) for homeland security applications as well as can be used in small animal imaging PET detector systems.

This is a preview of subscription content, log in via an institution to check access.

Access this chapter

Chapter
USD 29.95
Price excludes VAT (USA)
  • Available as PDF
  • Read on any device
  • Instant download
  • Own it forever
eBook
USD 149.00
Price excludes VAT (USA)
  • Available as EPUB and PDF
  • Read on any device
  • Instant download
  • Own it forever
Softcover Book
USD 199.99
Price excludes VAT (USA)
  • Compact, lightweight edition
  • Dispatched in 3 to 5 business days
  • Free shipping worldwide - see info
Hardcover Book
USD 199.99
Price excludes VAT (USA)
  • Durable hardcover edition
  • Dispatched in 3 to 5 business days
  • Free shipping worldwide - see info

Tax calculation will be finalised at checkout

Purchases are for personal use only

Institutional subscriptions

References

  1. W. Chaiphaksa, P. Limkitjaroenporn, H.J. Kim, J. Kaewkhao, The mass attenuation coefficients, effective atomic numbers and effective electron densities for GAGG:Ce and CaMoO4 scintillators. Prog. Nucl. Energy 92, 48–53 (2016)

    Article  Google Scholar 

  2. J. Iwanowska, L. Swiderski, T. Szczesniak, P. Sibczynski, M. Moszynski, M. Grodzicka et al., Performance of cerium-doped Gd3Al2G.A3O12 (GAGG:Ce) scintillator. Nucl. Instrum. Methods Phys. Res. A 712, 34–40 (2013)

    Article  ADS  Google Scholar 

  3. K. Kamada, T. Yanagida, T. Endo, K. Tsutumi, Y. Usuki, M. Nikl et al., 2 inch diameter single crystal growth and scintillation properties of Ce:Gd3Al2Ga3O12. J. Cryst. Growth 52, 88–90 (2012)

    Article  ADS  Google Scholar 

  4. M. Yoneyama, J. Kataoka, M. Arimoto, T. Masuda, M. Yoshino, K. Kamada, A. Yoshikawa, H. Sato, Y. Usuki, Evaluation of GAGG: Ce scintillators for future space applications. JINST 13, P02023 (2018)

    Article  ADS  Google Scholar 

  5. M. Lucchini, V. Babin, P. Bohacek, S. Gundacker, K. Kamada, M. Nikl, A. Petrosyan, A. Yoshikawa, E. Auffray, Effect of Mg2+ ions co-doping on timing performance and radiation tolerance of cerium doped Gd3Al2Ga3O12 crystals. Nucl. Instrum. Methods Phys. Res. Sect. A: Accel. Spectrometers Detect. Assoc. Equip. A 816, 176–183 (2016)

    Article  ADS  Google Scholar 

  6. M.T. Lucchinia, O. Buganov, E. Auffray, P. Bohacek, M. Korjik, D. Kozlov, S. Nargelas et al., Measurement of non-equilibrium carriers dynamics in Ce-doped YAG, LuAG and GAGG crystals with and without Mg-codoping. J. Lumin. 194, 1–7 (2018)

    Article  Google Scholar 

  7. S. Yamamoto, K. Kuroda, M. Senda, Scintillator selection for MR-compatible gamma detectors. IEEE Trans. Nucl. Sci. 50(5), 1683–1685 (2003)

    Article  ADS  Google Scholar 

  8. S. Yamamoto, H. Watabe, Y. Kanai, K. Kato, J. Hatazawa et al., Development of an ultrahigh-resolution Si-PM-based dual-head GAGG coincidence imaging system. Nucl. Instrum. Methods Phys. Res. A. 703, 183–189 (2013)

    Article  ADS  Google Scholar 

  9. J.Y. Yeom, S. Yamamoto, S.E. Derenzo, V.C. Spanoudaki, K. Kamada, T. Endo et al., First performance results of Ce:GAGG scintillation crystals with silicon photomultipliers. IEEE Trans. Nucl. Sci. 60(2), 988–992 (2013)

    Article  ADS  Google Scholar 

  10. T. Kato, J. Kataoka, T. Nakamori, T. Miura, H. Matsuda, A. Kishimoto et al., A novel gamma-ray detector with submillimeter resolutions using a monolithic MPPC array with pixelized Ce:LYSO and Ce:GGAG crystals. Nucl. Instr. and Meth. Phys. Res. A 699, 235–241 (2013)

    Article  ADS  Google Scholar 

  11. S. Yamamoto, J.Y. Yeom, K. Kamada, T. Endo, C.S. Levin, Development of an ultrahigh resolution block detector based on 0.4 mm pixel Ce:GAGG scintillators and a silicon photomultiplier array. IEEE Trans. Nucl. Sci. 60(6), 4582–4587 (2013)

    Article  ADS  Google Scholar 

  12. M. Georgiou, S. David, E. Fysikopoulos, G. Loudos, Development of a SIPM based gamma-ray imager using a Gd3Al2Ga3O12:Ce (GAGG:Ce) scintillator array, in IEEE Nuclear Science Symposium and Medical Imaging Conference, pp. M21–M48 (2013)

    Google Scholar 

  13. S. David, M. Georgiou, E. Fysikopoulos, G. Loudos, Evaluation of a SIPM array coupled to a Gd3Al2Ga3O12: Ce (GAGG:Ce) discrete scintillator. Phys. Med. 31, 763–766 (2015)

    Article  Google Scholar 

  14. J.H. Kim, K.H. Park, K.S. Joo, Development of compact and real-time radiation detector based on SiPM for gamma-ray spectroscopy. JINST 13, P07018 (2018)

    Article  ADS  Google Scholar 

  15. J.H. Kim, K.H. Park, K.S. Joo, Development of low-cost, compact, real-time, and wireless radiation monitoring system in underwater environment. Nucl. Eng. Technol. 50, 801–805 (2018)

    Article  Google Scholar 

  16. Available online at: http://www.furukawa-denshi.co.jp/cgi-bin/pdfdata/20140428162950.pdf

  17. V. Popov, Matrix output device readout system. U.S. Patent 6,747,263 (2004)

    Google Scholar 

  18. V. Popov, S. Majewski, B. Welch, A novel readout concept for multianode photomultiplier tubes with pad matrix anode layout. Nucl. Instrum. Methods A. 567, 319–322 (2006)

    Article  ADS  Google Scholar 

  19. V. Popov, S. Majewski, A.G. Weisenberger, Readout electronics for multianode photomultiplier tubes with pad matrix anode layout. IEEE Nucl. Sci. Conf. R 3, 2156–2159 (2003)

    Google Scholar 

  20. S. David, M. Georgiou, E. Fysikopoulos, G. Loudos, Comparison of three resistor network division circuits for the readout of 4 × 4 pixel SiPM arrays. Nucl. Instrum. Methods Phys. Res. A 702, 121–125 (2013)

    Article  Google Scholar 

  21. 12-Bit Octal-Channel ADC Family Up to 65MSPS (Texas Instruments, Dallas, TX, USA, 2008)

    Google Scholar 

  22. E. Fysikopoulos, G. Loudos, M. Georgiou, S. David, G. Matsopoulos, A Spartan 6 FPGA-based data acquisition system for dedicated imagers in nuclear medicine. Meas. Sci. Technol. 23(12), 125403 (2012)

    Article  ADS  Google Scholar 

  23. E. Pratiwi, H.T. Leem, J.H. Park, Y. Choi, S. Yamamoto, K. Kamada, A. Yoshikawa, J.Y. Yeom, Performance of a 0.4 mm pixelated Ce:GAGG block detector with digital silicon photomultiplier. IEEE Trans. Radiat. Plasma Med. Sci. 1, 30–35 (2017)

    Article  Google Scholar 

Download references

Acknowledgements

This research is implemented through IKY scholarships programme and co-financed by the European Union (European Social Fund—ESF) and Greek national funds through the action entitled” Reinforcement of Postdoctoral Researchers”, in the framework of the Operational Programme” Human Resources Development Program, Education and Lifelong Learning” of the National Strategic Reference Framework (NSRF) 2014–2020.

We would like to thanks Prof. S. Majewski who has provided us the GAGG:Ce matrix as well as Dr. L. Fysikopoulos and Dr. M. Georgiou for their contribution in read out signal processing and front end electronics.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to S. David .

Editor information

Editors and Affiliations

Rights and permissions

Reprints and permissions

Copyright information

© 2019 Springer Nature Switzerland AG

About this paper

Check for updates. Verify currency and authenticity via CrossMark

Cite this paper

David, S., Kandarakis, I. (2019). Development of a Submillimeter Portable Gamma-Ray Imaging Detector, Based on a GAGG:Ce—Silicon Photomultiplier Array. In: Korzhik, M., Gektin, A. (eds) Engineering of Scintillation Materials and Radiation Technologies. ISMART 2018. Springer Proceedings in Physics, vol 227. Springer, Cham. https://doi.org/10.1007/978-3-030-21970-3_15

Download citation

Publish with us

Policies and ethics