FormalPara Key Summary Points

Flash glucose monitoring can improve blood glucose in the short term of 14 days without the need to adjust hypoglycemic treatment.

The estimated A1c used by FreeStyle Libre™ is lower than actual glycated hemoglobin (HbA1c).

The gap between estimated A1c and actual measured HbA1c is positively correlated to the actual HbA1c.

Based on our results, we present an improved model for estimating A1c in flash glucose monitoring.

Introduction

The FreeStyle Libre™ (Abbott Diabetes Care, Witney, UK) flash glucose monitoring (FGM) system is one of a new generation of glucose-testing devices. The FGM system has made glucose checks easier with its ability to scan a sensor for a glucose reading at any time, no need for calibration, and a long sensor lifetime of 14 days. Many studies have confirmed the benefit of using the FGM in patients with type 1 or type 2 diabetes [14], including reductions in hyperglycemia, hypoglycemia, and associated medical costs.

Glycated hemoglobin (HbA1c) is considered to be the gold standard for the assessment of glycemic control [5] and has strong predictive value for diabetes complications [6] and mortality in patients with diabetes [7]. The relationship between HbA1c and average glucose levels has been explored in many studies, most of which used intermittent capillary blood glucose measurements [8, 9] or continuous glucose monitoring (CGM) [10, 11], resulting in a formula that allowed the calculation of an estimated A1c value using average glucose. The calculation, namely, estimated glycated hemoglobin (EA1c), calculated as A1c% = (mean glucose [MG] mmol/L + 2.59)/1.59 using the glucose data recorded by an FGM sensor for 14 days, was installed by Abbott Diabetes Care as a parameter in the FGM system. In some studies [1, 2] and clinical work, EA1c has been used to assess glycemic control, rather than the actual HbA1c. Dunn and colleagues analyzed the effects of FGM in more than 50,000 users worldwide and found that both EA1c and time in hyperglycemia were gradually reduced with increasing scan rate from the lowest to highest scan groups [2]. EA1c has also been used by clinicians to guide adjustments in the hypoglycemic regimen in a large number of patients. The use of EA1c instead of actual HbA1c measurements can be partially explained by the desire to reduce blood collections and hospital visits for HbA1c testing. However, an analysis of the relationship between EA1c and actual measured HbA1c has never been performed. Thus, the aim of the present study was to investigate the relationship between EA1c as measured in a FGM system and actual measured HbA1c in Chinese patients with type 2 diabetes.

Methods

Patients

This study was conducted in Nanjing First Hospital from December 2018 to July 2019. Patients with type 2 diabetes considered eligible to be enrolled in the study were (1) aged ≥ 18 years, (2) had HbA1c < 11.0%; and (3) were drug naïve or in stable hypoglycemic therapy for at least 3 months prior to the study. Exclusion criteria were: (1) pregnancy or currently attempting to conceive; (2) presence of infection or acute metabolic complications of diabetes, such as ketoacidosis or hyperosmolar state (coma); (3) allergy to medical grade adhesive or isopropyl alcohol used to prepare the skin; (4) presence of skin lesions, scarring, redness, infection, or edema at sensor application sites; (5) presence of anemia or other diseases that may affect HbA1c testing; (6) presence of any other condition considered to warrant exclusion according to the researchers involved in the study.

The study protocol was approved by the Institutional Ethics Committee of Nanjing First Hospital, Nanjing Medical University (number KY20170904-04), and was in accordance with the Helsinki Declaration of 1964, as revised in 2013. All patient provided written informed consent forms to participate in the study.

Study Design

Glycated hemoglobin was measured using a high-performance liquid chromatography assay (Bio-Rad Laboratories, Inc., Hercules, CA, USA) at baseline, and FGM was used for 14 days in patients after the baseline blood collection (period 1). The sensor was worn on the back of the upper arm to record subcutaneous interstitial glucose concentration at 15-min intervals without the requirement for self-monitoring of blood glucose (SMBG) for calibration. Patients could scan the sensor to read the glucose level at any time. The frequency of scanning by each patient for each sensor was calculated by counting the number of scans divided by the duration of sensor use (14 days), according to recorded start and end times. Patients were required to keep track of their food intake and exercise while wearing the FGM sensor and could alter their diet and exercise according to the glucose levels provided by FGM. Patients were also instructed to maintain their therapy with glucose-lowering agents during FGM in order to prevent the effects of a large glycemic variation caused by drug changes on the EA1c in FGM. The dosage of glucose-lowering agents could be adjusted by clinicians according to the results of the FGM and HbA1c when the FGM sensors were removed. To diminish the influence of any changes in diet and exercise on the EA1c, readings from FGM were also taken from days 76 to 90 (± 2) (period 2) and the HbA1c was measured again when at the end of this second FGM period. The height, weight, course of disease, concomitant diseases, and medication of all patients were recorded.

Statistical Analysis

All statistical analyses were performed using SPSS version 16.0 software (IBM Corp., Armonk, NY, USA). Data are presented as the mean ± standard deviation (SD) or as percentages. Differences between the EA1c and HbA1c were examined using Student’s paired t test and univariate tests. Differences in daily MG level during the 14-day study period were analyzed by repeated measures analysis of variance (ANOVA) followed with a Bonferroni test. The correlations of EA1c and HbA1c in the two FGM periods were compared using the cocor package (https://comparingcorrelations.org/). The models for the relationship between EA1c and HbA1c were analyzed using stepwise linear regression and exponential and quadratic equations. Total MG and the MG for each day, scanning frequency, and the SD of glucose levels in the FGM system were used to construct the models. The best fits, based on assessment of correlation coefficient values closest to 1.0 using Pearson correlations, were determined. Stepwise linear regression was also performed to test potential risk factors possibly influencing the gap between EA1c and HbA1c (HbA1c − EA1c); these variables were included as independent variables. p values < 0.05 were considered to indicate statistical significance.

Results

Patient Characteristics

A total of 80 patients were enrolled in the study, of whom six were excluded from the analyses due to interruptions in FGM. The final study population thus comprised 74 patients, of whom 35 (47.3%) were men; the mean age (± SD) of the patients was 67.6 ± 5.2 years. At baseline, the mean duration of type 2 diabetes was 11.9 ± 7.8 months, mean body mass index (BMI) was 25.1 ± 3.8 kg/m2, and mean HbA1c was 7.7 ± 1.3%. Fifty-two (70.3%) patients were on insulin therapy, and two patients were drug naïve. The baseline characteristics of the patients are given in Table 1.

Table 1 Baseline characteristics of patients

Differences between EA1c and Actual Measured HbA1c

The HbA1c at baseline was significantly higher than the EA1c (7.7 ± 1.3 vs. 7.1 ± 1.3%, respectively; p < 0.001). After 3 months (period 2), the HbA1c remained higher than the EA1c (7.5 ± 1.1 vs. 6.9 ± 1.1%, respectively; p < 0.001), although it should be noted that the actual HbA1c measurement was made after the second period of FGM.

Mean Daily Glucose During FGM

The repeated ANOVA revealed that there was no difference in MG between the two periods of FGM (8.8 ± 2.2 vs. 8.6 ± 1.9; p = 0.663). However, the daily MG reading changed with time (F = 7.2, p < 0.001 in the multivariate tests). To assess the influence of these changes in blood glucose on the EA1c, we pooled the FGM data from the two FGM periods and calculated the MG of each day (see Fig. 1). The MG was highest on the third day and then gradually decreased; this reduction in MG compared to the third day became statistically significant on the tenth day (p = 0.015), and MG on the 14th day was the lowest readings. MG on the first day was slightly lower than that on the second and third days (p = 0.070 and 0.054, respectively), and there was no difference between MG on the second and third days (p = 0.722).

Fig. 1
figure 1

Daily mean glucose of the patients during 14 days of flash glucose monitoring (FGM). Values are pooled data from two 14-day FGM monitoring periods. Asterisk(s) indicates a signficance difference in mean glucose (MG) as compared to MG on third day of 14-day FGM (*p < 0.05; **p < 0.01)

Factors Influencing Difference between EA1c and Actual Measured HbA1c

To explore the factors that have the potential to influence the difference between EA1c and HbA1c (ΔA1c = HbA1c − EAlc), we performed linear regression, including educational level, age, gender, duration of diabetes, BMI, HbA1c at baseline and times of daily scan as independent variables in the model. The results showed that only HbA1c at baseline affected the ΔA1c (β = 0.319, p = 0.010). We then combined the baseline and endpoint data and observed that ΔA1c increased with increasing HbA1c (r = 0.316, p < 0.001), as shown in Fig. 2. The frequency of scanning [number of scans daily = 11.9 (interquartile range 7.7–15.9)] was correlated to EA1c and HbA1c (r = − 0.164 and − 0.225, p = 0.016 and 0.006, respectively). However, there was no correlation between scan rates and ΔA1c (p > 0.05).

Fig. 2
figure 2

Relationship between glycated hemoglobin (HbA1c) and the gap between actual measured HbA1c (A1c) and estimated A1c (EA1c) during FGM. A total of 148 HbA1c measurements and the corresponding EA1c in the FGM system were included in linear regression

Correlation Analysis between EA1c and Actual Measured HbA1c

Pearson analysis was performed to investigate the relationship between EA1c and the actual measured HbA1c, with both EA1c values found to be significantly and positively correlated to the real HbA1c values (r = 0.752 and 0.810, respectively; both p < 0.001). The correlations between EA1c and the actual measured HbA1c in the first and second FGM periods were similar (p = 0.255). Thus, we combined the EA1c from these two FGM periods and the values from the two HbA1c tests; the correlation coefficient was 0.822 [p < 0.001 (n = 148)]. We then performed linear regression under different conditions to search for a better model of EA1c (shown in Table 2). Models 1 to 4 in Table 2 were calculated using MG only. The correlation coefficient in models 2 to 4 was not higher than the previous EA1c and was lower in the exponential and quadratic equation than in the linear regression. Given the influence of scanning frequency and variation in daily glucose levels, in the linear regression for models 5 to 8 we included MG on the second to 14th day of FGM, the SD of glucose levels (which represented glycemic variability), and the number of scans daily. In addition, given the strong effect of baseline HbA1c, which indicated blood glucose control, on ΔA1c, in two models we stratified the patients by MG (median value in model 6; by quartile in model 7). The best model was found to be EA1c% = MG mmol/L × 0.669 − 0.213 × 8th MG + 3.351 when MG > 9.7 mmol/L and EA1c% = (MG mmol/L + 2.590)/1.590 when MG ≤ 9.7 mmol/L (model 7 in Table 2). The correlation coefficient in model 7 was significantly higher than that in model 1 (previous EA1c) (p < 0.001).

Table 2 Models for estimated HbA1c calculation

Discussion

The results of our study reveal that in our patient population there was a gap between the EA1c in the FreeStyle Libre FGM system and the actual measured HbA1c and that this gap between EA1c and actual HbA1c was positively correlated to the levels of baseline HbA1c. Therefore, we conclude that EA1c should be calculated with the MG and adjusted using MG on the eighth day of FGM when MG is at a high level because the FreeStyle Libre FGM system itself has the effect reducing blood glucose (Table 2).

The difference between EA1c and actual measured HbA1c is mainly caused by the reduction in MG. Several studies have reported the effects of FGM on blood glucose [12, 13]. However, these studies were conducted in patients who were taking insulin therapy, and the insulin doses were adjusted during FGM. Moreover, these studies had durations of > 10 weeks. We found that the MG decreased significantly in a short period of 14 days without any change in the use of glucose-lowering agents. One possibly explanation for this change in blood glucose is an improved diet and increased physical activity. However, this possiblity needs to be further examined via analyses of the changes in diet and exercise during FGM in the present study population (study remains ongoing). Previous studies showed that FGM improved blood glucose control better in patients with high HbA1c levels than those with relatively lower levels [14, 15]. Although these studies explained the positive correlation between baseline HbA1c and ΔA1c, they did not calculate the EA1c in patients with high MG separately. Thus, we performed linear regressions among patients stratified according to MG to investigate different models of EA1c calculation in patients with different blood glucose control. Moreover, all of the variables we included in the models were from FGM; patients therefore did not need to provide additional blood samples or do other tests.

Kumagai et al. found that glucose values tended to be lower when measured using FGM than with CGM [16]. Moreover, it has been reported that the accuracy of data on the first day is lower than that on the other days during the FGM period [17, 18]. As a result, total MG (MG over the 14-day FGM period) in FGM is lower than the actual measured blood glucose level, and the EA1c calculated with total MG only is also lower than the actual HbA1c. The accuracy is stable in the middle stage of FGM [18], and thus adjustment of the EA1c using MG on eighth day of FGM partially solves this problem. The correlation coefficient increased in the model that was adjusted with MG on the eighth day.

Dunn et al. found that EA1c was negatively correlated to scan rates [2]. Our result confirmed that finding. However, we found no correlation between scan rates and ΔA1c, although we cannot exclude the influence of a smaller sample size and/or shorter study duration compared to the previous study.

This study has a number of limitations that need to be discussed. First, the change in diet and physical activity during FGM needs to be further studied. Secondly, findings using a larger sample size may be more convincing. Thirdly, our study was limited to a single center, which should be expanded. Moreover, although we did not allow any change in the dosage of glucose-lowering agents during FGM, we did permit changes in glucose-lowering agents immediately after the FGMs, in view of medical ethics; thus, the reduction in glucose level during the second FGM period may have been affected. However, the interval between the two FGMs (62 days) is sufficiently long to diminish the influence of hypoglycemic agent adjustment, and the influence of glucose-lowering agent adjustment on EA1c and HbA1c may be similar, so that it may have little influence on the relationship between EA1c and HbA1c in the second FGM. The correlations between EA1c and actual HbA1c in the first and second FGM period were similar in this study, which confirms our hypothesis.

In clinical work, the underestimation of actual HbA1c may result in a neglect of the effects of glucotoxicity during the past 3 months, which in turn may lead to diabetic complications. In studies, the use of EA1c may lead to a misjudgment of glycemic control in patients. We suggest a better model for EA1c calculation to resolve these problems. The proper use of EA1c will reduce both the number of blood samples taken and hospital visits. The results of this study may be useful for the application of FGM.

Conclusions

The EA1c in FreeStyle Libre is lower than the actual measured HbA1c. The improvement in glucose levels during FGM in these patients may contribute to the lowering of EA1c.