Abstract
We present a case in which a dip in the capnogram of an anesthetized patient, which may indicate clinical complications such as spontaneous respiratory effort, was caused by a malfunction of the anesthesia delivery system. The rubber diaphragm of the ventilator relief valve was found to be coated with a sticky substance which may have caused adhesion at the valve seat. This adhesion blocked the flow ot excess gas to the scavenging system during exhalation. It was demonstrated that a pressure of 5 cm H2O was needed to overcome this adhesion.
Similar content being viewed by others
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Hensler, T., Dhamee, M.S. Anesthesia machine malfunction simulating spontaneous respiratory effort. J Clin Monitor Comput 6, 128–131 (1990). https://doi.org/10.1007/BF02828289