Abstract
Internal fixation devices of less bending stiffness than conventional plates made of stainless steel or vitallium were compared with conventional plates in a study of fracture healing. The material for this investigation was a fine graphite fiber reinforced methyl methacrylate resin composite with a modulus of elasticity approximately ten times less than that of stainless steel. Osteotomies were performed on canine radii. Internal fixation was accomplished by means of a composite plate on the left side, and a stainless steel plate on the right. Clinical assessment, as well as biomechanical and quantitative histological techniques, were used to compare osteotomy healing of the two sides. At four months, all osteotomies had healed and the bioengineering tests showed radii from the two sides had equivalent strength. However, significantly less cortical porosity was found in the side with the composite plate (6.8%), as compared to that of the stainless steel plated side (14%). These results suggest that a less stiff fixation plate may have some advantage in the treatment of long bone fracture if there is no implant failure, and if union rates are equivalent.
Similar content being viewed by others
References
Bardos, D. I.: An evaluation of Ti-6Al-4V alloy for orthopedic application. Proceedings of the 20th Annual Orthopedic Research Soc., J. Bone Jt Surg. A56, 847 (1974)
Burstein, A. H., Currey, J., Frankel, V. H., Heiple, K. G., Lunseth, P., Vessely, J. C.: Bone strength. J. Bone Jt Surg. A54, 1143–1156 (1972)
Burstein, A. H., Frankel, V. H.: Standard tests for laboratory animals bone. J. Biomech.4, 155–158 (1971)
Dodge, H. S., Cady, G. W.: Treatment of fractures of the radius and ulna with compression plates. J. Bone Jt Surg. A54, 1167–1176 (1972)
Enneking, W. F., Burchardt, Pahl, J. J., Thornby, J.: Temporal and spatial activity in mirror segments of mature dog fibulae. Calcif. Tiss. Res.9, 283–295 (1972)
Harris, W. H., Haywood, E. A., Lavorgna, J., Hamblen, D. L.: Spatial and temporal variation in cortical bone formation in dogs. J. Bone Jt Surg. A50, 1118–1128 (1968)
Harris, W. H., Weinberg, H. E.: Microscopic method of measuring increases in cortical bone volume and mass. Calcif. Tiss. Res.8, 190–196 (1972)
Laurence, M., Freeman, M. A. R., Swanson, S. A. V.: Engineering considerations in the internal fixation of fractures of the tibial shaft. J. Bone Jt Surg. B51, 754–768 (1969)
Olerud, S., Danckwardt-Lilliestrom, G.: Fracture healing in compression osteosynthesis. Acta orthop. scand., Suppl. No. 137 (1971)
Olerud, S., Karlstrom, G.: Tibial fractures treated by AO Compression osteosynthesis. Acta orthop. scand., Suppl. No. 140 (1972)
Uhthoff, H. K., Dubuc, F. L.: Bone structure change in a dog under rigid internal fixation. Clin. Orthop. and Rel. Res.81, 165–170 (1971)
Woo, S. L-Y., Akeson, W. H., Levenetz, B., Coutts, R. D., Matthews, J. V., Amiel, D.: Potential application of graphite fiber and methyl methacrylate resin composites as internal fixation plates. J. Biomed. Mat'l Res.8, 321–338 (1974)
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Akeson, W.H., Woo, S.L.Y., Coutts, R.D. et al. Quantitative histological evaluation of early fracture healing of cortical bones immobilized by stainless steel and composite plates. Calc. Tis Res. 19, 27–37 (1975). https://doi.org/10.1007/BF02563988
Received:
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/BF02563988